Practical Steps in GC Troubleshooting

Presentations | 2017 | Agilent TechnologiesInstrumentation
GC
Industries
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Gas chromatography troubleshooting is essential for maintaining reliable analyses in research, quality control, industrial process monitoring, and environmental testing. Rapid diagnosis of injector, flow, column, detector, and electronic faults minimizes downtime, prevents sample loss, and ensures data integrity.

Goals and Study Overview


This application note presents a practical workflow for identifying and resolving common gas chromatography issues. It outlines systematic steps to isolate problem sources, explains diagnostic tests, and illustrates solutions with real-world examples.

Methodology and Instrumentation


The troubleshooting strategy centers on isolating five key system components:
  • Injector
  • Carrier gas flow
  • Column
  • Detector
  • Electronics

Diagnostic tools include blank runs, bleed profiles, non-retained peaks, test mixtures, condensation tests, and jumper tube tests. Used instrumentation examples:
  • Columns: DB-5 (30 m × 0.53 mm I.D., 1.5 µm), DB-624 (30 m × 0.53 mm I.D., 3.0 µm)
  • Oven temperature programs: isothermal and gradient ramps
  • Carrier gas: helium (36 cm/s), hydrogen (37.3 cm/s)
  • Injector: split (1 100) at 250–260 °C
  • Detector: FID at 250–320 °C

Main Results and Discussion


Key symptoms and root causes:
  • Peak tailing: active sites in injector or column, flow path dead volume
  • Bonus/ghost peaks: carryover, septum bleed, contaminants in gas or hardware
  • Split or fronting peaks: injector misconfiguration, solvent mismatch, column overload
  • No peaks or changed response: blocked syringes, leaks, detector malfunction
  • Retention shifts and resolution loss: leaks, column aging, contamination, temperature drift, flow variation
  • Baseline disturbances and noise: column bleed, electronic faults, gas impurities
  • Quantitation errors: detector linearity limits, adsorption, co-elution, matrix effects

Illustrated case studies demonstrate diagnosing a clogged syringe and removing column contamination via section trimming and backflush with solvents.

Benefits and Practical Applications


This structured approach streamlines fault identification and corrective action. Laboratories can apply these procedures to reduce instrument downtime, uphold method performance, and maintain consistent analytical results across diverse applications.

Future Trends and Opportunities


Emerging directions include automated diagnostic algorithms, real-time system monitoring, AI-driven maintenance scheduling, advanced inert column coatings, and enhanced detector designs to further improve uptime and data quality.

Conclusion


Systematic troubleshooting based on component isolation, targeted tests, and clear corrective steps empowers chromatographers to resolve issues efficiently. Adopting these best practices maximizes instrument performance and ensures robust analytical outcomes.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Troubleshooting GC Capillary Columns
Troubleshooting GC Capillary Columns
2011|Agilent Technologies|Presentations
Troubleshooting GC Capillary Columns Techniques, Tips, and Tricks Series 8 Simon Jones Application Engineer June 22, 2010 Page 1 What went wrong and how to fix it… Page 2  Common problems  Troubleshooting tools  Troubleshooting examples “Everything was…
Key words
injector, injectorcolumn, columnjumper, jumperdetector, detectortest, testtroubleshooting, troubleshootingcontamination, contaminationisolate, isolateelectronics, electronicstube, tubeblank, blankproblem, problemchange, changecontaminated, contaminatedpeaks
Practical Steps in GC Troubleshooting Techniques, Tips, and Tricks
Practical Steps in GC Troubleshooting Techniques, Tips, and Tricks Mark Sinnott Application Engineer GC Columns & Supplies Page 1 “Everything was just fine and then this happened! How do I go about TROUBLESHOOTING?” Page 3 “Everything was just fine and…
Key words
injector, injectorcolumn, columnjumper, jumperdetector, detectortest, testcontamination, contaminationisolate, isolatepeaks, peaksproblem, problemtroubleshooting, troubleshootingelectronics, electronicstube, tubeblank, blankchange, changecontaminated
A Tale of Two Peaks: Troubleshooting Poor Peak Shape
A Tale of Two Peaks: Troubleshooting Poor Peak Shape
2019|Agilent Technologies|Presentations
A Tale of Two Peaks: Troubleshooting Poor Peak Shape Alexander Ucci January 17, 2019 1 Agilent Restricted “Everything was Just Fine and then this Happened!” “How do I go about Troubleshooting?” Track events- log book • Changed column, liner, septum,…
Key words
restricted, restrictedagilent, agilentproblem, probleminjector, injectorwhat, whatelectronics, electronicsdetector, detectorcolumn, columnsyringe, syringedoes, doeslook, lookjumper, jumperoccurrence, occurrenceput, puttroubleshooting
General GC Column Troubleshooting
General GC Column Troubleshooting
2021|Agilent Technologies|Presentations
Get Your Peaks into Shape General GC Column Troubleshooting Alexander Ucci October 5, 2021 1 October 5, 2021 Get Your Peaks Into Shape DE44470.3469212963 “Everything Was Just Fine... and Then This Happened!” “How do I troubleshoot?” Track your actions/keep a…
Key words
get, getshape, shapepeaks, peaksyour, yourinjector, injectorcolumn, columnjumper, jumperdetector, detectorbaseline, baselinepeak, peakcontamination, contaminationtest, testtroubleshooting, troubleshootingtridecane, tridecaneelectronics
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike