GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Investigation of Key Parameters for a Smooth Method Transfer to a New Optimized Valve and Loop Headspace Autosampler

Posters | 2019 | Thermo Fisher Scientific | PittconInstrumentation
GC, HeadSpace
Industries
Pharma & Biopharma
Manufacturer
Thermo Fisher Scientific

Summary

Significance of Topic


Ensuring reproducible transfer of headspace GC methods between instruments is crucial in regulated environments. Reliable method migration supports consistent quality control, minimizes re-validation and accelerates lab workflows.

Objectives and Overview


This study evaluates the migration of USP 467 Class 2A residual solvent headspace methods from legacy valve-and-loop autosamplers (TriPlus 300 HS, Agilent 7697A) to the new Thermo Scientific TriPlus 500 HS. Four key parameters—system temperatures, vial shaking, vial pressurization and loop pressurization—were assessed for their impact on sensitivity, repeatability and method equivalence.

Methodology and Instrumentation


The existing method conditions were applied to both old and new autosamplers under identical GC settings. Parameter adjustments were limited to those unique to each system:
  • System temperatures: incubation, loop and transfer line
  • Vial shaking: Quick Spin Shaking vs conventional agitation
  • Vial and loop pressurization modes and times
  • Sample set: USP 467 residual solvent standard mixture

Instrumentation Used


  • Thermo Scientific TriPlus 500 HS autosampler with direct GC column interface
  • Thermo Scientific TriPlus 300 HS autosampler (Standard and Pressure modes)
  • Thermo Scientific TRACE 1310 GC-FID system


Main Results and Discussion


  • Quick Spin Shaking accelerated equilibrium, reducing incubation time by up to 25% while maintaining or improving recovery for polar and nonpolar analytes.
  • Proprietary pneumatic control in the TriPlus 500 HS delivered precise vial and loop pressures, yielding lower RSDs (≈1.6%) compared to legacy systems (>5%).
  • Method parameters ported directly (same setpoints) produced equivalent or higher peak areas, confirming smooth transfer without extensive re-optimization.
  • Direct column interface minimized dead volumes, reducing carryover and enhancing sensitivity for high-boiling residual solvents.


Benefits and Practical Applications


  • Simplified method development: fewer parameters to adjust, streamlined user interface.
  • Higher throughput: faster incubation and reliable automation reduce cycle times.
  • Regulatory compliance: consistent results facilitate audits and limit re-validation efforts.
  • Improved data quality: enhanced repeatability and sensitivity across a broad compound range.


Future Trends and Opportunities


Advanced headspace designs will likely incorporate real-time pressure and temperature feedback loops, AI-driven parameter optimization and expanded direct interfaces for LC and GC–MS platforms. Such developments can further reduce method transfer hurdles and broaden applications to environmental and food testing.

Conclusion


The TriPlus 500 HS autosampler demonstrates seamless transfer of USP 467 headspace methods from older valve-loop systems, offering superior repeatability, faster cycle times and simplified setup. The provided guidelines support efficient migration with minimal re-validation.

References


  • United States Pharmacopeia General Chapter USP 467 Organic Volatile Impurities, 2021.
  • Firor RL. Analysis of USP 467 Residual Solvents with Improved Repeatability Using the Agilent 7697A Headspace Sampler. Agilent Appl Note 5990-7625EN, 2012.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Investigation of key parameters for a smooth method transfer to the new Thermo Scientific TriPlus 500 Headspace Autosampler
WHITE PAPER 10705 Investigation of key parameters for a smooth method transfer to the new Thermo Scientific TriPlus 500 Headspace Autosampler Author Manuela Bergna, Davide Bressanello, Thermo Fisher Scientific, Milan, IT Goal Demonstrate how to efficiently transfer an analytical method…
Key words
loop, looppressure, pressuresame, sameshaking, shakingvial, vialpressurization, pressurizationautosampler, autosamplerheadspace, headspaceincubation, incubationporting, portingtransfer, transferxylene, xylenetemperature, temperaturefilling, fillingtime
Boost Productivity with Reliable GC Sampling Solutions
Boost Productivity with Reliable GC Sampling Solutions
2019|Thermo Fisher Scientific|Presentations
Boost Productivity with Reliable GC Sampling Solutions Daniela Cavagnino Product Marketing Manager GC and Autosamplers The world leader in serving science Thermo Scientific GC Sampling Systems Thermo Scientific™ TriPlus™ 500 Headspace Autosampler Thermo Scientific™ TriPlus™ RSH™ Autosampler Sample Throughput Thermo…
Key words
vial, vialsample, sampleincubation, incubationloop, looprsh, rshheadspace, headspacebest, besttriplus, tripluscarryover, carryovervortex, vortexinjection, injectionderivatization, derivatizationspme, spmeitex, itexunattended
Simplified, cost-effective headspace GC method for residual solvents analysis in pharmaceutical products
TECHNICAL NOTE 10681 Simplified, cost-effective headspace GC method for residual solvents analysis in pharmaceutical products Authors Goal Giulia Riccardino and Cristian Cojocariu, Thermo Fisher Scientific, Runcorn, UK The aim of this work was to develop a rapid, cost-effective, modified USP…
Key words
residual, residualvial, vialmethycyclohexane, methycyclohexanexylene, xyleneincubation, incubationloop, loopdichloromethane, dichloromethanechromatographic, chromatographiccumene, cumenesolvent, solventrelief, reliefpain, painheadspace, headspacefid, fidshaking
System Parameter and Performance Comparison Between Agilent 7697A and Agilent G1888A Headspace Samplers for USP <467>
System Parameter and Performance Comparison Between Agilent 7697A and Agilent G1888A Headspace Samplers for USP <467> Technical Overview Introduction The Agilent 7697A Headspace Sampler introduced onboard electronic pneumatics control and several instrument control enhancements to improve performance over earlier models…
Key words
vial, vialloop, looppressure, pressurefill, fillpressurization, pressurizationprocedure, procedurecontrol, controlsettings, settingsflow, flowshaking, shakingmin, minamount, amountresidual, residualdefault, defaultarea
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike