System Parameter and Performance Comparison Between Agilent 7697A and Agilent G1888A Headspace Samplers for USP <467>

Technical notes | 2014 | Agilent TechnologiesInstrumentation
GC, HeadSpace
Industries
Pharma & Biopharma
Manufacturer
Agilent Technologies

Summary

Significance of the topic


The analysis of residual solvents in pharmaceutical products is critical for ensuring patient safety, compliance with regulatory guidelines, and maintaining product quality. United States Pharmacopeia Method <467> is widely adopted for quantifying volatile organic impurities, following ICH Q3C recommendations. Headspace gas chromatography remains the gold standard for this application due to its precision and simplicity.

Study objectives and overview


This technical overview compares the performance and method parameters of two Agilent headspace samplers—the electronic pneumatic-controlled 7697A and the earlier G1888A model—when applied to USP <467> residual solvent analysis. The goal is to demonstrate that the 7697A meets or exceeds system suitability requirements for signal-to-noise, resolution, precision, and accuracy, and to facilitate method transfer and SOP documentation for laboratories upgrading instrumentation.

Methodology and instrumentation


Samples of Class 1 and Class 2A solvents were prepared at concentrations spanning five-fold below to six-fold above USP limits using 20 mL vials. Both samplers were connected to an Agilent 7890B GC with FID detection. Procedure A utilized a DB-624UI column for identification, while Procedure B employed an HP-INNOWax column for confirmation. Key headspace parameters included vial equilibration (30 min), oven temperature (80 °C), loop fill using active backpressure control, and a transfer line at 100 °C. Default loop fill mode, vial fill pressure (15 psi), and shaking level (7 for high sensitivity, 5 for moderate) were evaluated. Instrumentation options like replaceable fused-silica transfer lines (0.25–0.53 mm id) provided method optimization flexibility.

Key results and discussion


• Signal-to-noise ratios for 1,1,1-trichloroethane and benzene exceeded USP thresholds on both samplers, with the 7697A showing a 35.7 S/N versus 26.2 for G1888A in Procedure A and comparable gains in Procedure B.
• Resolution between critical pairs (acetonitrile/methylene chloride, acetonitrile/cis-dichloroethene) met or surpassed USP <467> standards on both systems; adjusting final loop pressure on the 7697A allowed tuning of sensitivity versus resolution.
• Precision improved from RSDs below 5% on the G1888A to generally under 2.5% RSD on the 7697A for Class 1 and Class 2A solvents, attributed to stable thermal zones and electronic pneumatic control.
• Accuracy testing across low and high concentration levels yielded deviations within 10% for low and 6% for high samples on both instruments.

Benefits and practical applications


The 7697A’s onboard pneumatic control eliminates the need for a GC AUX module, compensates for laboratory pressure fluctuations, and enhances sensitivity and repeatability. Laboratories upgrading from the G1888A or 7694 can reuse existing methods with minimal adjustments—typically setting vial fill mode, loop fill mode, and shaking speed—while adhering to USP <1224> method transfer guidelines.

Future trends and potential applications


• Further automation of headspace workflows and integration with laboratory information management systems (LIMS).
• Miniaturization of sample loops and lower dead-volume transfer lines for trace-level analyses.
• Coupling with mass spectrometry for enhanced identification of unknown volatiles.
• Advanced software algorithms for real-time method optimization and remote diagnostics.

Conclusion


The Agilent 7697A Headspace Sampler delivers equal or superior performance to the G1888A model for USP <467> residual solvent analysis, offering improved precision, sensitivity, and operational flexibility. Method revalidation is typically unnecessary, though formal method transfer documentation should reference the updated pneumatic and loop fill parameters to ensure compliance and reproducibility.

Reference


  • United States Pharmacopeia. General Chapter <467> Organic Volatile Impurities. USP 32-NF 27.
  • International Conference on Harmonisation. ICH Q3C Guidelines for Residual Solvents.
  • Firor RL. Analysis of USP <467> Residual Solvents with Improved Repeatability Using the Agilent 7697A Headspace Sampler. Agilent Technologies 5990-7625EN.
  • Agilent Technologies. Agilent 7697A vs G1888A Headspace Samplers. Publication 5991-1542EN.
  • Firor RL. Optimizing Vial Pressurization Parameters for USP <467> Using the 7697A Headspace Sampler. Application Note 5990-9106EN.
  • Bushey J. Thermal Zone Considerations for the Agilent 7697A Headspace Sampler. White Paper 5990-9892EN.
  • Jacq K, et al. A Generic Method for Residual Solvent Analysis Using Static Headspace-GC-FID/MS. Agilent Technologies 5989-9726EN.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Agilent 7697A Headspace Sampler Versus an Agilent G1888A Headspace Sampler
Agilent 7697A Headspace Sampler Versus an Agilent G1888A Headspace Sampler Technical Overview The Agilent 7697A Headspace Sampler has onboard electronics pneumatics control which was added to improve the overall instrument performance. With this addition, several significant differences can be noted…
Key words
loop, loopvent, ventvial, vialpassive, passivepressurization, pressurizationvalve, valvebackpressure, backpressuresettings, settingsactive, activecontrol, controlpressurize, pressurizepressure, pressurecarrier, carrieractions, actionsarea
Investigation of key parameters for a smooth method transfer to the new Thermo Scientific TriPlus 500 Headspace Autosampler
WHITE PAPER 10705 Investigation of key parameters for a smooth method transfer to the new Thermo Scientific TriPlus 500 Headspace Autosampler Author Manuela Bergna, Davide Bressanello, Thermo Fisher Scientific, Milan, IT Goal Demonstrate how to efficiently transfer an analytical method…
Key words
loop, looppressure, pressuresame, sameshaking, shakingvial, vialpressurization, pressurizationautosampler, autosamplerheadspace, headspaceincubation, incubationporting, portingtransfer, transferxylene, xylenetemperature, temperaturefilling, fillingtime
Residual Solvents Analysis Using an Agilent Intuvo 9000 GC with 8697 Headspace Sampler
Application Note Pharmaceutical Small Molecules Residual Solvents Analysis Using an Agilent Intuvo 9000 GC with 8697 Headspace Sampler Author Ian Eisele Agilent Technologies, Inc. Abstract In accordance with United States Pharmacopeia (USP) <467>, residual solvent analysis was performed on an…
Key words
loop, loopwax, waxresidual, residualsolvents, solventscpme, cpmemethylisobutylketone, methylisobutylketonevial, vialcyclopentyl, cyclopentylprocedure, procedureheadspace, headspacetime, timeinlet, inletchip, chipacetonitrile, acetonitrilefill
Investigation of Key Parameters for a Smooth Method Transfer to a New Optimized Valve and Loop Headspace Autosampler
Investigation of Key Parameters for a Smooth Method Transfer to a New Optimized Valve and Loop Headspace Autosampler Davide Bressanello1, Manuela Bergna1, Daniela Cavagnino1, Carlos Garcia2 1.Thermo Fisher Scientific, Milan, Italy 2.Thermo Fisher Scientific, Austin, TX ABSTRACT Figure 1. Comparison…
Key words
loop, looppressure, pressurevial, vialsame, sametransfer, transferincubation, incubationpressurization, pressurizationtemperature, temperatureshaking, shakingheadspace, headspacemode, modeequilibration, equilibrationline, linevalve, valvexylene
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike