Meeting the Korean Method ES 04605.3 for Formaldehyde (폼알데하이드) with Static Headspace GC/MS
Applications | 2016 | Teledyne LABSInstrumentation
Formaldehyde is a common disinfection by-product in drinking water with known health risks even at trace levels. Regulatory bodies set stringent limits, requiring reliable detection methods that ensure consumer safety and compliance.
The study aimed to validate the static headspace GC/MS method described in Korean Standard Method ES 04605.3 for formaldehyde monitoring in drinking water. Specific goals included achieving a Minimum Quantitation Limit below 10 ppb and demonstrating calibration linearity (r² > 0.98) with acceptable precision (RSD < 25 %).
• Derivatization of formaldehyde with O-(2,3,4,5,6-pentafluorobenzyl)hydroxylamine (PFBHA) to form a pentafluorobenzyl oximine adduct.
• Preparation of calibration and MQL standards by spiking deionized water (10 mL) with formaldehyde stock, internal standard (p-bromofluorobenzene), sodium chloride, and PFBHA reagent.
• Automated sample mixing and equilibration using the Teledyne Tekmar HT3 headspace sampler, followed by static headspace extraction and GC/MS analysis.
• Teledyne Tekmar HT3 Automated Static and Dynamic Headspace Vial Sampler fitted with mixer assembly.
• Agilent 7890B Gas Chromatograph coupled to 5977A Mass Spectrometer.
• Agilent DB-624UI capillary column (20 m × 0.18 mm ID, 1 µm film thickness).
• Calibration exhibited excellent linearity (r² = 0.9968) across 5–100 ppb, surpassing the r² > 0.98 criterion.
• The calculated MQL by linear regression was 9.3 ppb (118% recovery) and 9.8 ppb after background subtraction, meeting the < 10 ppb target.
• Relative standard deviation of response factors improved below 25% only after subtracting the PFBHA background formaldehyde.
• Linear calibration without background correction delivered reliable quantification, simplifying data analysis.
• Provides robust, automated monitoring of trace formaldehyde in drinking water.
• Ensures compliance with regulatory standards through low detection limits and high precision.
• Reduces manual handling and potential variability via automated headspace sampling.
• Adoption of real-time headspace sampling integrated with online GC/MS for continuous monitoring.
• Development of miniaturized or portable headspace systems for field analysis.
• Application of advanced data processing and machine learning for background correction and peak identification.
The HT3 sampler combined with static headspace GC/MS and PFBHA derivatization consistently met Korean Method ES 04605.3 criteria. Linear calibration proved superior, offering accurate low-level formaldehyde quantification without complex background corrections.
GC/MSD, HeadSpace, GC/SQ
IndustriesEnvironmental
ManufacturerAgilent Technologies, Teledyne LABS
Summary
Importance of the Topic
Formaldehyde is a common disinfection by-product in drinking water with known health risks even at trace levels. Regulatory bodies set stringent limits, requiring reliable detection methods that ensure consumer safety and compliance.
Study Objectives and Overview
The study aimed to validate the static headspace GC/MS method described in Korean Standard Method ES 04605.3 for formaldehyde monitoring in drinking water. Specific goals included achieving a Minimum Quantitation Limit below 10 ppb and demonstrating calibration linearity (r² > 0.98) with acceptable precision (RSD < 25 %).
Methodology
• Derivatization of formaldehyde with O-(2,3,4,5,6-pentafluorobenzyl)hydroxylamine (PFBHA) to form a pentafluorobenzyl oximine adduct.
• Preparation of calibration and MQL standards by spiking deionized water (10 mL) with formaldehyde stock, internal standard (p-bromofluorobenzene), sodium chloride, and PFBHA reagent.
• Automated sample mixing and equilibration using the Teledyne Tekmar HT3 headspace sampler, followed by static headspace extraction and GC/MS analysis.
Used Instrumentation
• Teledyne Tekmar HT3 Automated Static and Dynamic Headspace Vial Sampler fitted with mixer assembly.
• Agilent 7890B Gas Chromatograph coupled to 5977A Mass Spectrometer.
• Agilent DB-624UI capillary column (20 m × 0.18 mm ID, 1 µm film thickness).
Key Results and Discussion
• Calibration exhibited excellent linearity (r² = 0.9968) across 5–100 ppb, surpassing the r² > 0.98 criterion.
• The calculated MQL by linear regression was 9.3 ppb (118% recovery) and 9.8 ppb after background subtraction, meeting the < 10 ppb target.
• Relative standard deviation of response factors improved below 25% only after subtracting the PFBHA background formaldehyde.
• Linear calibration without background correction delivered reliable quantification, simplifying data analysis.
Benefits and Practical Applications
• Provides robust, automated monitoring of trace formaldehyde in drinking water.
• Ensures compliance with regulatory standards through low detection limits and high precision.
• Reduces manual handling and potential variability via automated headspace sampling.
Future Trends and Potential Applications
• Adoption of real-time headspace sampling integrated with online GC/MS for continuous monitoring.
• Development of miniaturized or portable headspace systems for field analysis.
• Application of advanced data processing and machine learning for background correction and peak identification.
Conclusion
The HT3 sampler combined with static headspace GC/MS and PFBHA derivatization consistently met Korean Method ES 04605.3 criteria. Linear calibration proved superior, offering accurate low-level formaldehyde quantification without complex background corrections.
References
- Korean Ministry of Environment. Drinking Water Standard Method ES 04605.3: Formaldehyde-Headspace/Gas Chromatography-Mass Spectrometry.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Meeting the Korean Method ES 04607.1 for Naphthalene (나프탈렌) with Static and Dynamic Headspace GC/MS
2016|Agilent Technologies|Applications
Meeting the Korean Method ES 04607.1 for Naphthalene (나프탈렌) with Static and Dynamic Headspace GC/MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Introduction The Korean Standard Method for Drinking Water ES 04607.1 requires static…
Key words
mql, mqltemp, temppreheat, preheatppb, ppbtrap, trapheadspace, headspacecalculation, calculationtime, timeequil, equilnaphthalene, naphthaleneplaten, platenvariable, variablesim, simmixer, mixerpressurize
Meeting the Korean Method ES 04602.1b for Vinyl Chloride, Acrylonitrile, and Bromoform (염화비닐, 아크릴로니트릴, 브로모포름) with Static and Dynamic Headspace GC/MS
2016|Agilent Technologies|Applications
Meeting the Korean Method ES 04602.1b for Vinyl Chloride, Acrylonitrile, and Bromoform (염화비닐, 아크릴로니트릴, 브로모포름) with Static and Dynamic Headspace GC/MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Introduction The Korean Method ES 04602.1b…
Key words
mql, mqltemp, temppreheat, preheatacrylonitrile, acrylonitrilevinyl, vinylppb, ppbsim, simtrap, traptime, timeequil, equilchloride, chlorideplaten, platenvariable, variablemixer, mixerpressurize
Meeting the Korean Method ES 04603.2b for 12 VOC (휘발성유기화합물) with Static and Dynamic Headspace GC/MS
2016|Agilent Technologies|Applications
Meeting the Korean Method ES 04603.2b for 12 VOC (휘발성유기화합물) with Static and Dynamic Headspace GC/MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Introduction The Korean Standard Method for Drinking Water ES 04603.2b requires…
Key words
mql, mqltemp, tempxylenes, xylenesfluorobenzene, fluorobenzenetetrachloroethylene, tetrachloroethylenepreheat, preheattrichloroethylene, trichloroethylenetetrachloride, tetrachloridedichloromethane, dichloromethanesum, sumppb, ppbethylbenzene, ethylbenzenetrap, trapchloroform, chloroformxylene
Meeting the Korean Method ES 04601.2 for 1,4- Dioxane (1,4-다이옥산-헤드스페이스/기체크로마토그래피- 질량분석법) with Static Headspace GC/MS
2016|Agilent Technologies|Applications
Meeting the Korean Method ES 04601.2 for 1,4Dioxane (1,4-다이옥산-헤드스페이스/기체크로마토그래피질량분석법) with Static Headspace GC/MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Introduction The Korean Method ES 04601.2 requires static headspace with GC/MS to monitor 1,4-dioxane.…
Key words
mql, mqltemp, tempequil, equilplaten, platenpressurize, pressurizeppb, ppbtime, timevariable, variablemixer, mixertable, tablecalibration, calibrationfill, fillmixing, mixingvalue, valuesim