Meeting the Korean Method ES 04602.1b for Vinyl Chloride, Acrylonitrile, and Bromoform (염화비닐, 아크릴로니트릴, 브로모포름) with Static and Dynamic Headspace GC/MS
Applications | 2016 | Teledyne LABSInstrumentation
Accurate monitoring of vinyl chloride, acrylonitrile and bromoform at sub-ppb levels in drinking water is critical for regulatory compliance and safeguarding public health. The Korean Method ES 04602.1b establishes stringent limits and performance criteria that drive innovation in automated headspace GC/MS analysis.
The study aimed to implement and validate both static and dynamic headspace GC/MS approaches using the Teledyne Tekmar HT3 sampler to meet ES 04602.1b requirements: a minimum quantitation limit (MQL) below 5 ppb, calibration linearity (r² > 0.98) or RSD of response factors under 25%.
Samples and calibration standards were prepared in 20 mL headspace vials containing 3 g NaCl, 10 mL deionized water and 10 µL of a 2 ppm internal standard mix (fluorobenzene, chlorobenzene-d5, 1,4-dichlorobenzene-d4). A 5 ppm stock solution provided analyte levels for calibration (0, 5, 10, 25 ppb). Seven replicate MQL samples at 5 ppb were analyzed to calculate method precision and MQL by multiplying the standard deviation by 10.
Static headspace SIM analysis yielded MQLs of ~2.0 ppb for vinyl chloride, 4.9 ppb for acrylonitrile and 1.2 ppb for bromoform with RSDs below 12% and linearity r² > 0.991. Dynamic headspace full-scan delivered MQLs of 1.8 ppb (vinyl chloride), 5.7 ppb (acrylonitrile), 3.2 ppb (bromoform) with excellent precision. Dynamic SIM further improved acrylonitrile sensitivity, achieving ~1.7 ppb MQL and RSDs under 5% for all targets. All results exceed ES 04602.1b criteria.
The HT3 static and dynamic headspace GC/MS workflows reliably met or exceeded Korean Method ES 04602.1b performance targets. Dynamic SIM analysis in particular delivered superior sensitivity for acrylonitrile, underscoring the value of flexible headspace strategies in regulatory and research laboratories.
GC/MSD, HeadSpace, GC/SQ
IndustriesEnvironmental
ManufacturerAgilent Technologies, Teledyne LABS
Summary
Importance of the Topic
Accurate monitoring of vinyl chloride, acrylonitrile and bromoform at sub-ppb levels in drinking water is critical for regulatory compliance and safeguarding public health. The Korean Method ES 04602.1b establishes stringent limits and performance criteria that drive innovation in automated headspace GC/MS analysis.
Objectives and Study Overview
The study aimed to implement and validate both static and dynamic headspace GC/MS approaches using the Teledyne Tekmar HT3 sampler to meet ES 04602.1b requirements: a minimum quantitation limit (MQL) below 5 ppb, calibration linearity (r² > 0.98) or RSD of response factors under 25%.
Methodology
Samples and calibration standards were prepared in 20 mL headspace vials containing 3 g NaCl, 10 mL deionized water and 10 µL of a 2 ppm internal standard mix (fluorobenzene, chlorobenzene-d5, 1,4-dichlorobenzene-d4). A 5 ppm stock solution provided analyte levels for calibration (0, 5, 10, 25 ppb). Seven replicate MQL samples at 5 ppb were analyzed to calculate method precision and MQL by multiplying the standard deviation by 10.
Used Instrumentation
- Headspace sampler: Teledyne Tekmar HT3 Automated Static and Dynamic Headspace Vial Sampler
- Gas chromatograph: Agilent 7890B with DB-624UI column (20 m × 0.18 mm ID, 1 µm)
- Mass spectrometer: Agilent 5977A MSD operated in Scan/SIM modes
- Data system: Agilent Environmental ChemStation
Main Results and Discussion
Static headspace SIM analysis yielded MQLs of ~2.0 ppb for vinyl chloride, 4.9 ppb for acrylonitrile and 1.2 ppb for bromoform with RSDs below 12% and linearity r² > 0.991. Dynamic headspace full-scan delivered MQLs of 1.8 ppb (vinyl chloride), 5.7 ppb (acrylonitrile), 3.2 ppb (bromoform) with excellent precision. Dynamic SIM further improved acrylonitrile sensitivity, achieving ~1.7 ppb MQL and RSDs under 5% for all targets. All results exceed ES 04602.1b criteria.
Benefits and Practical Applications
- Automated headspace sampling increases throughput and reproducibility for routine water quality monitoring.
- Dynamic modes enhance sensitivity for challenging analytes such as acrylonitrile.
- Robust internal standard approach corrects for matrix and instrument variability.
Future Trends and Potential Applications
- Integration with high-resolution MS for broader contaminant screening.
- Miniaturized, field-deployable headspace modules for near-real-time monitoring.
- Advanced data analytics and machine learning for automated peak identification and quantitation.
- Expansion to multi-residue protocols in complex matrices beyond drinking water.
Conclusion
The HT3 static and dynamic headspace GC/MS workflows reliably met or exceeded Korean Method ES 04602.1b performance targets. Dynamic SIM analysis in particular delivered superior sensitivity for acrylonitrile, underscoring the value of flexible headspace strategies in regulatory and research laboratories.
References
- Korean Method ES 04602.1b, Vinyl Chloride, Acrylonitrile, Bromoform – Headspace/GC/MS
- ES 04602.1b: 염화비닐, 아크릴로니트릴, 브로모포름-드스페이스/기체크로마토그래피-질량분석법
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Using the Teledyne Tekmar HT3TM Headspace Analyzer to Meet the Requirements of the New Korean Standard Method for Water Pollution, ES04602.1, for Bromoform, Vinyl Chloride and Acrylonitrile
2011|Agilent Technologies|Applications
Using the Teledyne Tekmar HT3TM Headspace Analyzer to Meet the Requirements of the New Korean Standard Method for Water Pollution, ES04602.1, for Bromoform, Vinyl Chloride and Acrylonitrile Application Note Abstract The Ministry of Environment of the Republic of Korea recently…
Key words
temp, temppreheat, preheattrap, trapacrylonitrile, acrylonitriletime, timepressurize, pressurizemixing, mixingbake, bakedesorb, desorbkorean, koreanequil, equilmixer, mixerloop, loopsweep, sweepplaten
Meeting the Korean Method ES 04603.2b for 12 VOC (휘발성유기화합물) with Static and Dynamic Headspace GC/MS
2016|Agilent Technologies|Applications
Meeting the Korean Method ES 04603.2b for 12 VOC (휘발성유기화합물) with Static and Dynamic Headspace GC/MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Introduction The Korean Standard Method for Drinking Water ES 04603.2b requires…
Key words
mql, mqltemp, tempxylenes, xylenesfluorobenzene, fluorobenzenetetrachloroethylene, tetrachloroethylenepreheat, preheattrichloroethylene, trichloroethylenetetrachloride, tetrachloridedichloromethane, dichloromethanesum, sumppb, ppbethylbenzene, ethylbenzenetrap, trapchloroform, chloroformxylene
Meeting the Korean Method ES 04607.1 for Naphthalene (나프탈렌) with Static and Dynamic Headspace GC/MS
2016|Agilent Technologies|Applications
Meeting the Korean Method ES 04607.1 for Naphthalene (나프탈렌) with Static and Dynamic Headspace GC/MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Introduction The Korean Standard Method for Drinking Water ES 04607.1 requires static…
Key words
mql, mqltemp, temppreheat, preheatppb, ppbtrap, trapheadspace, headspacecalculation, calculationtime, timeequil, equilnaphthalene, naphthaleneplaten, platenvariable, variablemixer, mixersim, simpressurize
Meeting the Korean Method ES 04601.2 for 1,4- Dioxane (1,4-다이옥산-헤드스페이스/기체크로마토그래피- 질량분석법) with Static Headspace GC/MS
2016|Agilent Technologies|Applications
Meeting the Korean Method ES 04601.2 for 1,4Dioxane (1,4-다이옥산-헤드스페이스/기체크로마토그래피질량분석법) with Static Headspace GC/MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Introduction The Korean Method ES 04601.2 requires static headspace with GC/MS to monitor 1,4-dioxane.…
Key words
mql, mqltemp, tempequil, equilplaten, platenpressurize, pressurizeppb, ppbtime, timevariable, variablemixer, mixertable, tablecalibration, calibrationfill, fillmixing, mixingvalue, valuesim