Achieving Low-Level Detection of Benzene in Beverages with an Automated Headspace Vial Sampler
Applications | 2013 | Teledyne LABSInstrumentation
Accurate monitoring of benzene in beverages is critical due to its carcinogenic nature and potential formation from benzoate preservatives in the presence of ascorbic acid. Regulatory limits for benzene in drinking water range from 1 to 5 ppb, but beverage matrices and routine quality control demand detection capabilities at lower concentration levels to ensure consumer safety.
This study evaluates the performance of the Teledyne Tekmar HT3 Automated Headspace Vial Sampler combined with GC/MS for detecting benzene at parts-per-trillion levels in eight commercially available beverages. Key objectives include establishing low-level quantitation, assessing linearity over a broad range, and demonstrating method applicability for routine beverage analysis.
The static headspace approach employed the HT3 sampler set at 70 °C platen and 120 °C transfer line, with 10 min sample equilibration and a GC cycle of 32 min. GC separation used a Restek Rtx-502.2 column (30 m×0.25 mm×1.4 µm) with helium carrier gas. The oven program began at 60 °C, ramped to 150 °C at 8 °C/min, then to 260 °C at 20 °C/min. The mass spectrometer operated in full scan and selected ion monitoring (SIM) modes, targeting benzene ions at m/z 78, 77, 56 and internal standard benzene-d6 at m/z 84. A benzene stock standard (0.5 ppm) and benzene-d6 internal standard (4.2 ppb) were used to prepare calibration vials from 0.05 to 20 ppb. Eight beverage types—including sodas, energy drinks and cocktail mixes—were spiked with internal standard and analyzed under identical conditions.
Linearity over 0.05–20 ppb benzene was excellent (r2=0.9987). The method quantitated benzene at 0.05 ppb, surpassing European and US drinking water requirements by factors of 20 and 100. Beverage results revealed all samples below 1 ppb, with the highest concentration of 0.782 ppb in a margarita mix. An energy drink showed 0.065 ppb, well below existing detection limits. Overlay of total ion chromatograms highlighted additional volatile compounds, demonstrating the method’s capability for broader flavor and quality profiling.
The HT3 headspace sampler coupled with GC/MS provides a robust, sensitive and linear method for benzene determination in beverages at ppt levels. Its performance exceeds current drinking water standards and offers valuable qualitative insights into other volatile constituents, making it a powerful tool for beverage quality and safety laboratories.
GC/MSD, HeadSpace, GC/SQ
IndustriesFood & Agriculture
ManufacturerTeledyne LABS
Summary
Importance of the Topic
Accurate monitoring of benzene in beverages is critical due to its carcinogenic nature and potential formation from benzoate preservatives in the presence of ascorbic acid. Regulatory limits for benzene in drinking water range from 1 to 5 ppb, but beverage matrices and routine quality control demand detection capabilities at lower concentration levels to ensure consumer safety.
Aims and Overview of the Study
This study evaluates the performance of the Teledyne Tekmar HT3 Automated Headspace Vial Sampler combined with GC/MS for detecting benzene at parts-per-trillion levels in eight commercially available beverages. Key objectives include establishing low-level quantitation, assessing linearity over a broad range, and demonstrating method applicability for routine beverage analysis.
Methodology and Instrumentation
The static headspace approach employed the HT3 sampler set at 70 °C platen and 120 °C transfer line, with 10 min sample equilibration and a GC cycle of 32 min. GC separation used a Restek Rtx-502.2 column (30 m×0.25 mm×1.4 µm) with helium carrier gas. The oven program began at 60 °C, ramped to 150 °C at 8 °C/min, then to 260 °C at 20 °C/min. The mass spectrometer operated in full scan and selected ion monitoring (SIM) modes, targeting benzene ions at m/z 78, 77, 56 and internal standard benzene-d6 at m/z 84. A benzene stock standard (0.5 ppm) and benzene-d6 internal standard (4.2 ppb) were used to prepare calibration vials from 0.05 to 20 ppb. Eight beverage types—including sodas, energy drinks and cocktail mixes—were spiked with internal standard and analyzed under identical conditions.
Main Results and Discussion
Linearity over 0.05–20 ppb benzene was excellent (r2=0.9987). The method quantitated benzene at 0.05 ppb, surpassing European and US drinking water requirements by factors of 20 and 100. Beverage results revealed all samples below 1 ppb, with the highest concentration of 0.782 ppb in a margarita mix. An energy drink showed 0.065 ppb, well below existing detection limits. Overlay of total ion chromatograms highlighted additional volatile compounds, demonstrating the method’s capability for broader flavor and quality profiling.
Benefits and Practical Applications
- Ultra-low detection limits enable compliance with stringent regulations and early identification of contamination.
- Static headspace automation increases sample throughput and reproducibility in QC laboratories.
- SIM and full scan modes allow simultaneous quantitation and qualitative screening for multiple volatiles.
- Minimal sample preparation reduces solvent use and operational complexity.
Future Trends and Potential Applications
- Extension to other drink matrices such as juices, teas and flavored waters.
- Integration with tandem mass spectrometry for enhanced selectivity and lower detection limits.
- Development of rapid screening workflows using chemometric analysis of headspace profiles.
- Automation advances for unattended, high-throughput beverage safety testing.
Conclusion
The HT3 headspace sampler coupled with GC/MS provides a robust, sensitive and linear method for benzene determination in beverages at ppt levels. Its performance exceeds current drinking water standards and offers valuable qualitative insights into other volatile constituents, making it a powerful tool for beverage quality and safety laboratories.
References
- ICBA Guidance Document to Mitigate the Potential for Benzene Formation in Beverages, First Update 22 June 2006, The British Soft Drinks Association Ltd
- US FDA Determination of Benzene in Soft Drinks and Other Beverages, May 19 2006
- Teledyne Tekmar Application Note, Achieving ppt Levels of Environmental Volatiles with the HT3 Autosampler, July 2013
- NIST MS Search 2.0, National Institute of Standards and Technology
Similar PDF
Meeting the Korean Method ES 04603.2b for 12 VOC (휘발성유기화합물) with Static and Dynamic Headspace GC/MS
2016|Agilent Technologies|Applications
Meeting the Korean Method ES 04603.2b for 12 VOC (휘발성유기화합물) with Static and Dynamic Headspace GC/MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Introduction The Korean Standard Method for Drinking Water ES 04603.2b requires…
Key words
mql, mqltemp, tempxylenes, xylenesfluorobenzene, fluorobenzenetetrachloroethylene, tetrachloroethylenepreheat, preheattrichloroethylene, trichloroethylenedichloromethane, dichloromethanetetrachloride, tetrachloridesum, sumppb, ppbtrap, trapethylbenzene, ethylbenzenechloroform, chloroformxylene
Meeting the Korean Method ES 04607.1 for Naphthalene (나프탈렌) with Static and Dynamic Headspace GC/MS
2016|Agilent Technologies|Applications
Meeting the Korean Method ES 04607.1 for Naphthalene (나프탈렌) with Static and Dynamic Headspace GC/MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Introduction The Korean Standard Method for Drinking Water ES 04607.1 requires static…
Key words
mql, mqltemp, temppreheat, preheatppb, ppbtrap, trapheadspace, headspacecalculation, calculationtime, timeequil, equilnaphthalene, naphthaleneplaten, platensim, simvariable, variablemixer, mixerpressurize
Comparison of BTEXS in Olive Oils by Static and Dynamic HT3 Headspace
2011|Thermo Fisher Scientific|Applications
Comparison of BTEXS in Olive Oils by Static and Dynamic HT3 Headspace Application Note Abstract The health benefits of consumption of olive oils as part of a healthy diet reaches back to the mid 1950’s. In the late 1990’s, interest…
Key words
static, statictemp, tempdynamic, dynamicbtexs, btexspreheat, preheatolive, olivevirgin, virginitaly, italytime, timeequil, equilextra, extraglass, glasstrap, trapplaten, platenclear
Meeting the Korean Method ES 04602.1b for Vinyl Chloride, Acrylonitrile, and Bromoform (염화비닐, 아크릴로니트릴, 브로모포름) with Static and Dynamic Headspace GC/MS
2016|Agilent Technologies|Applications
Meeting the Korean Method ES 04602.1b for Vinyl Chloride, Acrylonitrile, and Bromoform (염화비닐, 아크릴로니트릴, 브로모포름) with Static and Dynamic Headspace GC/MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Introduction The Korean Method ES 04602.1b…
Key words
mql, mqltemp, tempacrylonitrile, acrylonitrilepreheat, preheatvinyl, vinylppb, ppbsim, simtrap, traptime, timechloride, chlorideequil, equilplaten, platenvariable, variablemixer, mixerheadspace