Comparison of BTEXS in Olive Oils by Static and Dynamic HT3 Headspace
Applications | 2011 | Teledyne LABSInstrumentation
The characterization of volatile organic compounds such as benzene, toluene, ethylbenzene, xylenes and styrene (BTEXS) in olive oil has become critical due to potential health implications and processing contaminants. Comprehensive analysis ensures product quality and consumer safety.
This study compares static and dynamic headspace techniques using the Teledyne Tekmar HT3 system coupled with GC/MS to quantify BTEXS residues in seven commercially available olive oils.
The Teledyne Tekmar HT3 system effectively performs both static and dynamic headspace analyses to detect BTEXS in olive oil. Dynamic mode significantly enhances sensitivity and reduces sample requirements, making it an ideal tool for quality control and contaminant screening in the olive oil industry.
GC/MSD, HeadSpace, GC/SQ
IndustriesFood & Agriculture
ManufacturerThermo Fisher Scientific, Teledyne LABS
Summary
Significance of the Topic
The characterization of volatile organic compounds such as benzene, toluene, ethylbenzene, xylenes and styrene (BTEXS) in olive oil has become critical due to potential health implications and processing contaminants. Comprehensive analysis ensures product quality and consumer safety.
Objectives and Overview
This study compares static and dynamic headspace techniques using the Teledyne Tekmar HT3 system coupled with GC/MS to quantify BTEXS residues in seven commercially available olive oils.
Methodology and Instrumentation
- Sample Preparation
Static mode used 10 g of oil per vial; dynamic mode used 2 g per vial with Teflon septa and aluminum caps. - Instrumentation
Teledyne Tekmar HT3 static/dynamic headspace sampler
Thermo Scientific Focus GC with DSQ II mass spectrometer
Phenomenex Zebron ZB-624 capillary column
Supelco Vocarb 3000 trap for dynamic sampling - Analytical Parameters
Static headspace: 90 °C equilibrium, 20 min equilibration; dynamic headspace: trap and desorption at 250 °C, sweep flow 75 mL/min, with optimized timing. GC oven ramp from 35 °C to 240 °C over 19 min; MS scan range 25–350 m/z.
Main Results and Discussion
- Calibration and Detection
Primary and confirmation ions were defined for each BTEXS compound; retention times shifted slightly between modes. - Sensitivity Comparison
Dynamic headspace delivered at least 100-fold higher sensitivity than static mode, with a fivefold improvement in signal-to-noise using only one-fifth of the sample mass. - Sample Variability
Normalized peak areas revealed distinct BTEXS profiles. Benzene was most abundant in the light-flavor oil. Toluene and xylenes varied among samples, with the organic extra virgin oil showing generally lower BTEXS levels.
Benefits and Practical Applications
- Dynamic headspace enables low-level detection of trace contaminants, strengthening food safety testing.
- Combined static and dynamic methods allow simultaneous profiling of flavor volatiles and contaminant quantification in a single analytical sequence.
- Reduced sample consumption and trap-based matrix reduction improve laboratory throughput and data reliability.
Future Trends and Potential Applications
- Integration of automated headspace sampling with high-resolution mass spectrometry for comprehensive volatile fingerprinting.
- Application of dynamic headspace methods to other food products and environmental matrices requiring ultra-trace analysis.
- Development of real-time headspace monitoring for inline quality control during production.
Conclusion
The Teledyne Tekmar HT3 system effectively performs both static and dynamic headspace analyses to detect BTEXS in olive oil. Dynamic mode significantly enhances sensitivity and reduces sample requirements, making it an ideal tool for quality control and contaminant screening in the olive oil industry.
References
- B. Gilbert-López et al., Talanta 83 (2010) 391-399
- F. Peña et al., Journal of Chromatography A 1052 (2004) 137-143
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Gasoline Range Organics (GRO) by Headspace/GC/FID
2010|Agilent Technologies|Applications
Gasoline Range Organics (GRO) by Headspace/GC/FID Application Note Abstract Recent oil spills in the Gulf of Mexico and the Great Barrier Reef have renewed interest in the rapid analysis of crude oil byproducts in aqueous and solid samples including sand…
Key words
temp, tempgro, grotime, timepressurize, pressurizesum, sumexternal, externalstatic, statictrap, trapdynamic, dynamiccalibration, calibrationbake, bakemtbe, mtbedesorb, desorbwisconsin, wisconsinequil
Meeting the Korean Method ES 04603.2b for 12 VOC (휘발성유기화합물) with Static and Dynamic Headspace GC/MS
2016|Agilent Technologies|Applications
Meeting the Korean Method ES 04603.2b for 12 VOC (휘발성유기화합물) with Static and Dynamic Headspace GC/MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Introduction The Korean Standard Method for Drinking Water ES 04603.2b requires…
Key words
mql, mqltemp, tempxylenes, xylenesfluorobenzene, fluorobenzenetetrachloroethylene, tetrachloroethylenepreheat, preheattrichloroethylene, trichloroethylenetetrachloride, tetrachloridedichloromethane, dichloromethanesum, sumppb, ppbethylbenzene, ethylbenzenetrap, trapchloroform, chloroformxylene
Using the Teledyne Tekmar HT3TM Headspace Analyzer to Meet the Requirements of the New Korean Standard Method for Water Pollution, ES04602.1, for Bromoform, Vinyl Chloride and Acrylonitrile
2011|Agilent Technologies|Applications
Using the Teledyne Tekmar HT3TM Headspace Analyzer to Meet the Requirements of the New Korean Standard Method for Water Pollution, ES04602.1, for Bromoform, Vinyl Chloride and Acrylonitrile Application Note Abstract The Ministry of Environment of the Republic of Korea recently…
Key words
temp, temppreheat, preheattrap, trapacrylonitrile, acrylonitriletime, timepressurize, pressurizemixing, mixingbake, bakedesorb, desorbequil, equilkorean, koreanmixer, mixerloop, loopsweep, sweepplaten
A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3TM Automatic Headspace Analyzer
2009|Agilent Technologies|Applications
A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3TM Automatic Headspace Analyzer Application Note By: Anne Jurek Abstract This application note presents method development using classical headspace sample preparative techniques for the TM…
Key words
temp, tempppb, ppbpreheat, preheatconcentration, concentrationplaten, platentba, tbamixer, mixertime, timeetbe, etbetrap, trapmixing, mixingvariable, variablebake, bakelevel, levelstandby