EPA Method 524 for Determination of VOCs in Drinking Water Using Agilent 5975T LTM GC/MSD with Static Headspace
Applications | 2010 | Agilent TechnologiesInstrumentation
Volatile organic compounds in drinking water pose serious health risks and frequently arise from improper storage or spills of industrial solvents and fuels. Rapid, onsite analysis is crucial for early detection, informed decision making in pollution incidents and ensuring compliance with regulatory standards.
This study presents an ultra fast analytical workflow based on static headspace sampling coupled with a transportable GC/MS system. The aim was to separate and quantify 54 priority VOCs in drinking water within a 9-minute run time while meeting EPA Method 524.2 performance criteria for detection limits, calibration linearity and recovery.
The protocol employs a static headspace sampler with 10-minute equilibration at 80 °C and 1 mL loop injection. Sample preparation involves spiking defined volumes of a multi-analyte VOC standard and internal standards into 10 mL of organic-free water. Calibration curves span 1 to 200 µg/L and method detection limits are established by seven replicate measures near the expected detection threshold. Precision and recovery are evaluated using five replicate 25 µg/L spiked samples.
The method achieved baseline separation of 54 VOCs in under 9 minutes. Calibration linearity across seven levels produced relative standard deviations below 15% in all cases. Method detection limits ranged from 0.2 to 1.0 µg/L. Recovery tests at 25 µg/L yielded values between 90 and 110% with repeatability RSDs below 6%. A custom deconvolution library processed by Agilent DRS identified targets in under one minute, reducing data review time and improving confidence in compound identification.
The streamlined headspace-LTM GC/MS workflow delivers fast, accurate VOC analysis suitable for mobile environmental labs. Reduced run times and rapid data processing enable high sample throughput during field campaigns. Compliance with EPA Method 524.2 ensures results are accepted for regulatory monitoring and emergency response.
Further miniaturization of GC instruments and enhanced deconvolution algorithms will improve sensitivity and selectivity for trace VOCs in complex matrices. Integration with wireless networks and AI-driven interpretation could enable real-time water quality surveillance and remote decision support in environmental monitoring.
The described static headspace-LTM GC/MS method meets stringent EPA 524.2 requirements while cutting analysis and data review times by roughly 40%. Its portability and rapid turnaround make it an effective solution for onsite VOC monitoring in drinking water.
GC/MSD, HeadSpace, GC/SQ
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Significance of the topic
Volatile organic compounds in drinking water pose serious health risks and frequently arise from improper storage or spills of industrial solvents and fuels. Rapid, onsite analysis is crucial for early detection, informed decision making in pollution incidents and ensuring compliance with regulatory standards.
Objectives and study overview
This study presents an ultra fast analytical workflow based on static headspace sampling coupled with a transportable GC/MS system. The aim was to separate and quantify 54 priority VOCs in drinking water within a 9-minute run time while meeting EPA Method 524.2 performance criteria for detection limits, calibration linearity and recovery.
Methodology
The protocol employs a static headspace sampler with 10-minute equilibration at 80 °C and 1 mL loop injection. Sample preparation involves spiking defined volumes of a multi-analyte VOC standard and internal standards into 10 mL of organic-free water. Calibration curves span 1 to 200 µg/L and method detection limits are established by seven replicate measures near the expected detection threshold. Precision and recovery are evaluated using five replicate 25 µg/L spiked samples.
Used Instrumentation
- Static headspace sampler Agilent 7694E with 13-minute total cycle time, 80 °C vial and 90 °C needle temperatures
- Transportable GC/MS system Agilent 5975T LTM MSD equipped with an LTM DB-624 column (20 m × 0.18 mm, 1.0 µm)
- Carrier gas helium at constant pressure, temperature program from 50 °C to 220 °C with rapid LTM heating and cooling
- Data acquisition in scan and selected ion monitoring (SIM) modes
Results and discussion
The method achieved baseline separation of 54 VOCs in under 9 minutes. Calibration linearity across seven levels produced relative standard deviations below 15% in all cases. Method detection limits ranged from 0.2 to 1.0 µg/L. Recovery tests at 25 µg/L yielded values between 90 and 110% with repeatability RSDs below 6%. A custom deconvolution library processed by Agilent DRS identified targets in under one minute, reducing data review time and improving confidence in compound identification.
Benefits and practical applications
The streamlined headspace-LTM GC/MS workflow delivers fast, accurate VOC analysis suitable for mobile environmental labs. Reduced run times and rapid data processing enable high sample throughput during field campaigns. Compliance with EPA Method 524.2 ensures results are accepted for regulatory monitoring and emergency response.
Future trends and opportunities
Further miniaturization of GC instruments and enhanced deconvolution algorithms will improve sensitivity and selectivity for trace VOCs in complex matrices. Integration with wireless networks and AI-driven interpretation could enable real-time water quality surveillance and remote decision support in environmental monitoring.
Conclusion
The described static headspace-LTM GC/MS method meets stringent EPA 524.2 requirements while cutting analysis and data review times by roughly 40%. Its portability and rapid turnaround make it an effective solution for onsite VOC monitoring in drinking water.
References
- US EPA Method 524.2 Methods for the Determination of Organic Compounds in Drinking Water
- US EPA Method 8260 Volatile Organic Compounds by Gas Chromatography Mass Spectrometry
- Philip L Wylie Developing an eMethod for the Analysis of Volatile Organic Compounds in Water Using Purge and Trap GC with Agilent 5975 inert Mass Spectrometer 5989 3347EN
- Xiaofei Ping Building Agilent GC MSD Deconvolution Reporting Libraries for Any Application 5989 2249EN
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Agilent 5975T LTM GC/MSD
2013|Agilent Technologies|Brochures and specifications
Agilent 5975T LTM GC/MSD PERFORM CONFIDENT, LAB-QUALITY ANALYSIS IN THE FIELD Introducing the Agilent 5975T LTM GC/MSD Fast, reliable results for all of your out-of-lab analysis You’re looking at Agilent innovation on the move. The new Agilent 5975T Low Thermal…
Key words
field, fieldvocs, vocscts, ctsagilent, agilentsim, simlab, labtsp, tspfaster, fastertransportable, transportableanalysis, analysissampling, samplingyour, yourdrs, drsbenzo, benzohyperbolic
Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD
2010|Agilent Technologies|Applications
Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD Application Note Environmental Author Abstract Suli Zhao, Andy Zhai This application note describes an onsite solution for detecting 13 volatile aromatic Agilent Technologies Co., Ltd.…
Key words
soil, soilaromatic, aromaticvolatile, volatilespme, spmebenzene, benzenecompounds, compoundsdrs, drsamdis, amdisfluorobenzene, fluorobenzenetoluene, tolueneheadspace, headspacebromobenzene, bromobenzenemode, modetargets, targetsfresh
Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS
2011|Agilent Technologies|Applications
Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application Environmental Author Philip L. Wylie Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA (E-mail: [email protected]) Abstract The analysis of volatile organic compounds in water…
Key words
purge, purgebake, bakesame, sametemperature, temperatureblank, blankwater, watermsd, msdcalibration, calibrationresponse, responsedesorb, desorbvoa, voavocs, vocsistd, istdrsd, rsdmust
Agilent water analysis solution - Volatile Organic Compounds Analysis in Water
2011|Agilent Technologies|Others
Agilent water analysis solution Volatile Organic Compounds Analysis in Water Environmental Due to an increase in the frequency of water pollution, requirements for onsite water quality monitoring has increased. With new high performance transportable instruments, such as the Agilent 5975T…
Key words
vocs, vocswater, waterfast, fastsite, sitefifty, fiftyagilent, agilentdrs, drscritically, criticallyltm, ltmonsite, onsitenotice, noticenine, ninecompounds, compoundssubject, subjectdeconvolution