PBDE Analysis Using an Agilent J&W DB-5ms Ultra Inert GC Column
Applications | 2011 | Agilent TechnologiesInstrumentation
Polybrominated diphenyl ethers (PBDEs) are widely used flame retardants that persist in the environment and bioaccumulate in food chains. Accurate trace-level analysis of these compounds is critical for monitoring environmental contamination, ensuring food safety, and assessing human exposure risk. The analytical challenge of high retention and thermal instability, particularly of highly brominated congeners like BDE-209, demands highly inert separation media and sensitive detection.
This study demonstrates a robust method for quantifying eight PBDE congeners at sub-nanogram levels using electron impact gas chromatography–mass spectrometry (GC–MS). Key goals include minimizing analyte degradation, achieving sharp peak shapes on a shortened column to reduce residence time, and verifying linearity and sensitivity down to 0.005 ng on-column for most congeners and 0.025 ng for BDE-209.
The method achieved excellent separation of eight PBDE congeners with symmetrical peaks and baseline resolution in a 15-min run. Calibration curves for all congeners displayed linearity (R2 ≥ 0.997) down to 0.005 ng on-column (0.025 ng for BDE-209). A signal-to-noise ratio greater than 3:1 was observed for BDE-209 at the lowest level, underscoring the sensitivity imparted by the Ultra Inert column and optimized SIM detection.
Upcoming developments include coupling Ultra Inert columns with next-generation triple-axis detectors for even lower detection limits, expanding the methodology to emerging flame retardants, and integrating automated sample preparation for high-throughput environmental monitoring and compliance testing.
The presented GC–MS approach employing an Agilent J&W DB-5ms Ultra Inert column offers a reliable, sensitive, and rapid solution for trace-level PBDE analysis. Superior column inertness and optimized MS detection ensure accurate quantification of challenging congeners, supporting environmental assessment and food safety initiatives.
GC/MSD, GC/SQ, GC columns, Consumables
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Importance of the topic
Polybrominated diphenyl ethers (PBDEs) are widely used flame retardants that persist in the environment and bioaccumulate in food chains. Accurate trace-level analysis of these compounds is critical for monitoring environmental contamination, ensuring food safety, and assessing human exposure risk. The analytical challenge of high retention and thermal instability, particularly of highly brominated congeners like BDE-209, demands highly inert separation media and sensitive detection.
Objectives and Study Overview
This study demonstrates a robust method for quantifying eight PBDE congeners at sub-nanogram levels using electron impact gas chromatography–mass spectrometry (GC–MS). Key goals include minimizing analyte degradation, achieving sharp peak shapes on a shortened column to reduce residence time, and verifying linearity and sensitivity down to 0.005 ng on-column for most congeners and 0.025 ng for BDE-209.
Methodology and Instrumentation
- GC–MS System: Agilent 6890N gas chromatograph coupled to a 5973B mass selective detector in SIM mode with an inert electron impact source.
- Column: Agilent J&W DB-5ms Ultra Inert, 15 m × 0.25 mm i.d., 0.25 µm film thickness (p/n 122-5512UI) to limit high-temperature exposure time.
- Injection: Pulsed splitless (325 °C, 20 psi for 1.5 min), with a 5.0 µL syringe in splitless mode delivering 5 ng of each PBDE standard per injection.
- Carrier Gas: Helium at 72 cm/s constant flow.
- Oven Program: Ramp from 150 °C to 325 °C at 17 °C/min, hold for 5 min.
- Mass Spectrometry: Source at 300 °C, quadrupole at 150 °C, transfer line at 300 °C, scan range 200–1000 amu.
Key Results and Discussion
The method achieved excellent separation of eight PBDE congeners with symmetrical peaks and baseline resolution in a 15-min run. Calibration curves for all congeners displayed linearity (R2 ≥ 0.997) down to 0.005 ng on-column (0.025 ng for BDE-209). A signal-to-noise ratio greater than 3:1 was observed for BDE-209 at the lowest level, underscoring the sensitivity imparted by the Ultra Inert column and optimized SIM detection.
Benefits and Practical Applications
- Enhanced Inertness: The DB-5ms Ultra Inert phase virtually eliminates active sites, preventing analyte adsorption and degradation.
- Reduced Analysis Time: A 15 m column shortens residence time for thermally labile PBDEs without sacrificing resolution.
- Trace-Level Sensitivity: Sub-nanogram detection limits enable monitoring of PBDEs in environmental and food matrices at regulatory thresholds.
- Reproducibility: High linearity and stable baselines support routine quality control and comparability across laboratories.
Future Trends and Applications
Upcoming developments include coupling Ultra Inert columns with next-generation triple-axis detectors for even lower detection limits, expanding the methodology to emerging flame retardants, and integrating automated sample preparation for high-throughput environmental monitoring and compliance testing.
Conclusion
The presented GC–MS approach employing an Agilent J&W DB-5ms Ultra Inert column offers a reliable, sensitive, and rapid solution for trace-level PBDE analysis. Superior column inertness and optimized MS detection ensure accurate quantification of challenging congeners, supporting environmental assessment and food safety initiatives.
References
- Lynam K, Smith D (2011). PBDE Analysis Using an Agilent J&W DB-5ms Ultra Inert GC Column. Agilent Technologies Application Brief 5990-5651EN.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Polybrominated Diphenyl Ether (PBDE) Analysis Using an Agilent J&W DB-5ms Ultra Inert Capillary GC Column
2008|Agilent Technologies|Applications
Polybrominated Diphenyl Ether (PBDE) Analysis Using an Agilent J&W DB-5ms Ultra Inert Capillary GC Column Application Environmental Authors Ken Lynam and Doris Smith Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808 USA Abstract Trace and ultra trace-level polybrominated diphenyl…
Key words
bde, bdepbdes, pbdesinertness, inertnesspbde, pbdepolybrominated, polybrominateddiphenyl, diphenylultra, ultrainert, inertabundance, abundancesuccessful, successfultendency, tendencysupplies, suppliesagilent, agilentvials, vialscolumn
Lower Detection Limits and Quantitate with Confidence with Breakthrough Ultra Inert Technology
2011|Agilent Technologies|Presentations
Lower Detection Limits and Quantitate with Confidence with Breakthrough Ultra Inert Technology Simon Jones Application Engineer September 22, 2011 Goals of the Presentation 1. Understand what it means to be INERT, and why it is important to have an inert…
Key words
liner, linerinert, inertdeactivation, deactivationprobes, probesinertness, inertnessultra, ultrapoor, poortest, testdeact, deactliners, linersagilent, agilenttrimethyl, trimethylactive, activecolumn, columnactivity
Trace Level Analysis Made Routine - New Agilent J&W Ultra Inert Capillary GC Columns
2011|Agilent Technologies|Presentations
New Agilent J&W Ultra Inert Capillary GC Columns Raising the Bar for CONSISTENT Column Inertness Performance Trace Level Analysis Made Routine Page 1 Agilent Restricted 6 April 2011 For Research Use Only. Not for use in diagnostic procedures. Outline •…
Key words
restricted, restrictedinertness, inertnessagilent, agilentinert, inertcolumn, columntest, testprobes, probesultra, ultradeactivation, deactivationtrimethyl, trimethyldemanding, demandingactive, activetrace, tracepeak, peakmix
Solutions that meet your demands - Environmental Compendium - Emerging Contaminants
2009|Agilent Technologies|Applications
Solutions that meet your demands for: speed accuracy productivity Excellent choices for environmental applications Emerging Contaminants PPCPs Applications > Return to Table of Contents > Search entire document Determination of Pharmaceuticals in Water by SPE and LC/MS/MS in Both Positive…
Key words
pbde, pbdebde, bdepfoa, pfoapbdes, pbdesabundance, abundancepfos, pfospolybrominated, polybrominatedusa, usaion, ionmass, massdiphenyl, diphenylagilent, agilentfragmentor, fragmentorcompounds, compoundscollision