Residual Solvents Testing per ASTM F1884-04 with the Versa and HT3 Automated Headspace Sampler
Applications | 2012 | Teledyne LABSInstrumentation
The determination of residual solvents in packaging materials is critical for ensuring consumer safety, regulatory compliance, and product quality. Solvent traces from printing inks can migrate into food or pharmaceutical products, making sensitive analytical methods essential for routine quality control in packaging production.
This application study evaluates the performance of the Teledyne Tekmar Versa and HT3 Automated Headspace Samplers in accordance with ASTM F1884-04. Using a Restek Rxi-624Sil MS column and a Shimadzu 2010 Plus GC-FID system, the work investigates calibration linearity, the impact of added water on solvent recovery, and the use of helium versus nitrogen as carrier gases.
Samples and standards were prepared following ASTM F1884-04 guidelines. A five-point calibration curve was generated using an internal standard (4-heptanone) in water. Printable polypropylene film was pre-treated in a vacuum oven at 90 °C for 24 h, then loaded into 22 mL headspace vials either dry or with added water or internal standard solution.
Calibration curves exhibited excellent linearity (r² ≥ 0.99) for the majority of residual solvents, with the lowest detection limits around 1 mg/ream. Both helium and nitrogen carrier gases delivered comparable performance. The addition of water to film samples enhanced early-eluting solvent signals but suppressed later-eluting compounds, highlighting matrix effects inherent in headspace analysis.
The automated headspace approach offers minimal sample preparation, flexible carrier gas choice, and robust sensitivity for residual solvent quantification. Packaging manufacturers can reliably monitor solvent residues to meet tightening regulatory thresholds and ensure product safety.
Further method optimization may tailor conditions for specific solvents or packaging matrices. Future work could explore alternative detectors, expanded solvent libraries, and real-time monitoring to enhance throughput and broaden applicability across diverse materials.
The Teledyne Tekmar Versa and HT3 samplers combined with GC-FID meet the rigorous demands of ASTM F1884-04, delivering high linearity, low detection limits, and operational flexibility. Matrix effects observed with added water suggest targeted optimization may be required for certain analytes.
GC, HeadSpace
IndustriesEnergy & Chemicals , Materials Testing
ManufacturerShimadzu, Teledyne LABS
Summary
Significance of the Topic
The determination of residual solvents in packaging materials is critical for ensuring consumer safety, regulatory compliance, and product quality. Solvent traces from printing inks can migrate into food or pharmaceutical products, making sensitive analytical methods essential for routine quality control in packaging production.
Study Objectives and Overview
This application study evaluates the performance of the Teledyne Tekmar Versa and HT3 Automated Headspace Samplers in accordance with ASTM F1884-04. Using a Restek Rxi-624Sil MS column and a Shimadzu 2010 Plus GC-FID system, the work investigates calibration linearity, the impact of added water on solvent recovery, and the use of helium versus nitrogen as carrier gases.
Methodology and Instrumentation
Samples and standards were prepared following ASTM F1884-04 guidelines. A five-point calibration curve was generated using an internal standard (4-heptanone) in water. Printable polypropylene film was pre-treated in a vacuum oven at 90 °C for 24 h, then loaded into 22 mL headspace vials either dry or with added water or internal standard solution.
- Headspace Sampler (Versa and HT3): Sample temperature 90 °C; equilibration 20 min; pressurization 13 psig; transfer line 110 °C; valve oven 105 °C.
- GC-FID (Shimadzu 2010 Plus): Restek Rxi-624Sil MS, 30 m × 0.32 mm ID, 1.8 µm; oven program from 40 °C (2 min) to 250 °C; inlet at 250 °C, split 10:1; FID at 250 °C with hydrogen (40 mL/min) and air (400 mL/min).
Results and Discussion
Calibration curves exhibited excellent linearity (r² ≥ 0.99) for the majority of residual solvents, with the lowest detection limits around 1 mg/ream. Both helium and nitrogen carrier gases delivered comparable performance. The addition of water to film samples enhanced early-eluting solvent signals but suppressed later-eluting compounds, highlighting matrix effects inherent in headspace analysis.
Benefits and Practical Applications
The automated headspace approach offers minimal sample preparation, flexible carrier gas choice, and robust sensitivity for residual solvent quantification. Packaging manufacturers can reliably monitor solvent residues to meet tightening regulatory thresholds and ensure product safety.
Future Trends and Applications
Further method optimization may tailor conditions for specific solvents or packaging matrices. Future work could explore alternative detectors, expanded solvent libraries, and real-time monitoring to enhance throughput and broaden applicability across diverse materials.
Conclusion
The Teledyne Tekmar Versa and HT3 samplers combined with GC-FID meet the rigorous demands of ASTM F1884-04, delivering high linearity, low detection limits, and operational flexibility. Matrix effects observed with added water suggest targeted optimization may be required for certain analytes.
References
- ASTM Method F1884-04, Standard Test Methods for Determining Residual Solvents in Packaging Materials, ASTM International, West Conshohocken, PA.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Two Headspace Solutions for the Analysis of Dissolved Gases in Water by Method RSKSOP-175
2012|Agilent Technologies|Applications
Two Headspace Solutions for the Analysis of Dissolved Gases in Water by Method RSKSOP-175 Application Note Abstract Dissolved gas analysis has been performed using a simplified headspace method following US EPA method RSKSOP-175. This method describes the preparation and analysis…
Key words
versa, versaethene, etheneequil, equilpressurize, pressurizedissolved, dissolvedethane, ethanepropane, propanetemp, tempheadspace, headspacetime, timemethane, methaneplaten, platenstock, stockmixer, mixerfill
Blood Alcohol Determination with the New Teledyne Tekmar Versa Headspace Instrument
2011|Shimadzu|Applications
Blood Alcohol Determination with the New Teledyne Tekmar Versa Headspace Instrument Application Note Introduction Recently, Teledyne Tekmar released the Versa, their latest innovative headspace instrument. Initially developed for the pharmaceutical industry to provide an economical headspace system for residual solvent…
Key words
versa, versaarea, areaequil, equilblood, bloodalcohol, alcoholheadspace, headspacepressurize, pressurizeacetaldehyde, acetaldehydetemp, tempisopropanol, isopropanolplaten, platenacetone, acetoneethanol, ethanolpropanol, propanoleconomical
Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers
2013|Agilent Technologies|Applications
Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers Application Note Versa HT3 Abstract With the rising prices of fossil fuels, more emphasis is being put on renewable resources and green technologies. Alternative fuel sources, such…
Key words
versa, versahelium, heliumnitrogen, nitrogenequil, equilminimum, minimumpressurize, pressurizemethanol, methanolcoefficient, coefficienttemp, tempplaten, platennumber, numbersample, sampletime, timemaximum, maximumheadspace
Alternative Methods to RSK 175 Using Purge and Trap Concentration and Automated Headspace for the Analysis of Dissolved Gases in Drinking Water
2013|Teledyne LABS|Applications
Alternative Methods to RSK 175 Using Purge and Trap Concentration and Automated Headspace for the Analysis of Dissolved Gases in Drinking Water Application Note Abstract With the growing price of oil, many alternative energy sources are being explored. Natural gas…
Key words
temp, temppurge, purgebake, baketime, timedesorb, desorbversa, versapressurize, pressurizeequil, equilcondenser, condenserrinse, rinseethene, etheneflow, flowplaten, platencalibration, calibrationvariable