Femtogram GC/MSD Detection Limits for Environmental Semivolatiles Using a Triple-Axis Detector
Applications | 2008 | Agilent TechnologiesInstrumentation
The ability to detect semivolatile organic compounds at femtogram levels by GC/MS with a triple-axis detector addresses critical demands in environmental analysis. Ultra-low detection limits enable more reliable monitoring of complex mixtures of pesticides, acids, bases, and neutral compounds in drinking and source water. Enhanced sensitivity supports compliance with stringent regulatory requirements and improves confidence in trace-level quantitation.
This work examines how key GC/MSD parameters influence instrument detection limits (IDLs) for over 100 semivolatiles using:
The goal is to identify a workflow that consistently delivers femtogram-level detection and reproducible quantitation across a broad analyte panel.
Sample preparation was performed by liquid-solid extraction on C18 cartridges or disks, with dichloromethane liquid-liquid extraction as an alternative. Extracts were analyzed by GC/MS under the following conditions:
Signal-to-noise (S/N) ratios for selected analytes showed: SIM mode improved S/N by 3× to 15× over scan acquisition, enabling reliable integration at 0.1 ppb (0.2 pg on column) for many compounds. Scan mode at 2 pg produced S/N values above 2–7 for most targets, while SIM at 0.2 pg delivered S/N of 15–77. Linearity across 0.2–200 pg (SIM) yielded an average %RSD of 8%, demonstrating system inertness and reproducibility. Full extraction ion chromatograms at 200 fg confirm clear analyte peaks well above noise.
Further advancements may include higher-capacity PTV liners for large-volume injection, enhanced detectors with even lower noise floors, automated method optimization routines, and integration of retention time locking with AI-driven compound identification. Expanding these techniques to complex solid and waste matrices through tailored cleanup workflows could broaden femtogram-level detection across diverse analytical contexts.
By combining a cold splitless PTV inlet, trace ion detection, and a triple-axis MSD detector in SIM mode, laboratories can achieve femtogram-level detection limits for a broad range of environmental semivolatiles. This approach enhances sensitivity, maintains chromatographic integrity, and simplifies method robustness, supporting more confident trace-level quantitation.
GC/MSD, GC/SQ
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Importance of Topic
The ability to detect semivolatile organic compounds at femtogram levels by GC/MS with a triple-axis detector addresses critical demands in environmental analysis. Ultra-low detection limits enable more reliable monitoring of complex mixtures of pesticides, acids, bases, and neutral compounds in drinking and source water. Enhanced sensitivity supports compliance with stringent regulatory requirements and improves confidence in trace-level quantitation.
Objectives and Study Overview
This work examines how key GC/MSD parameters influence instrument detection limits (IDLs) for over 100 semivolatiles using:
- A programmable temperature vaporizing (PTV) inlet operated in cold splitless mode
- Trace ion detection (TID) filtering
- A triple-axis detector (TAD) running in selected ion monitoring (SIM) mode
The goal is to identify a workflow that consistently delivers femtogram-level detection and reproducible quantitation across a broad analyte panel.
Methods and Instrumentation
Sample preparation was performed by liquid-solid extraction on C18 cartridges or disks, with dichloromethane liquid-liquid extraction as an alternative. Extracts were analyzed by GC/MS under the following conditions:
- Gas chromatograph: Agilent 7890A/6890N with EPC PTV inlet (Agilent multi-baffle liner, no packing)
- Inlet mode: Cold splitless injection of 2 μL, PTV held at 20 °C for 0.05 min then ramped to 350 °C; purge at 30 mL/min after 1.5 min
- Oven program: 40 °C initial, rapid ramp to 110 °C, then 10 °C/min to 320 °C (total run 26 min), retention time locking (RTL) to phenanthrene-d10 at 12.700 min
- Column: HP-5MSi, 30 m × 0.25 mm × 0.25 μm, constant flow 1.4 mL/min
- Mass spectrometer: Agilent 5975C MSD with triple-axis detector, source 300 °C, transfer line 280 °C, mass range 45–450 amu, threshold 0, drawout lens 3 mm standard aperture
- Acquisition: SIM mode via AutoSIM (25 groups, ions per group 2–45, dwell times 5–50 ms, minimum 10 cycles/peak) and full scan for comparison
Main Results and Discussion
Signal-to-noise (S/N) ratios for selected analytes showed: SIM mode improved S/N by 3× to 15× over scan acquisition, enabling reliable integration at 0.1 ppb (0.2 pg on column) for many compounds. Scan mode at 2 pg produced S/N values above 2–7 for most targets, while SIM at 0.2 pg delivered S/N of 15–77. Linearity across 0.2–200 pg (SIM) yielded an average %RSD of 8%, demonstrating system inertness and reproducibility. Full extraction ion chromatograms at 200 fg confirm clear analyte peaks well above noise.
Benefits and Practical Applications
- Sub-picogram detection limits enable trace-level environmental monitoring far below regulatory minima
- Cold splitless PTV inlet reduces analyte degradation and solvent expansion, improving peak shape
- TAD and TID reduce neutral noise and enhance sensitivity for clean matrices
- Retention time locking simplifies method maintenance and high-throughput operation
Future Trends and Possibilities
Further advancements may include higher-capacity PTV liners for large-volume injection, enhanced detectors with even lower noise floors, automated method optimization routines, and integration of retention time locking with AI-driven compound identification. Expanding these techniques to complex solid and waste matrices through tailored cleanup workflows could broaden femtogram-level detection across diverse analytical contexts.
Conclusion
By combining a cold splitless PTV inlet, trace ion detection, and a triple-axis MSD detector in SIM mode, laboratories can achieve femtogram-level detection limits for a broad range of environmental semivolatiles. This approach enhances sensitivity, maintains chromatographic integrity, and simplifies method robustness, supporting more confident trace-level quantitation.
Reference
- USEPA Method 525.2 (2008).
- Szelewski M., Wilson B., Perkins P. Improvements in the Agilent 6890/5973 GC/MSD System for Use with USEPA Method 8270. Agilent Technologies, 2008.
- Szelewski M. Fast Semivolatiles Analysis Using the Agilent 6890/5973 inert GC/MSD. Agilent Technologies, 2008.
- Szelewski M. Part-per-Trillion Calibration of Semivolatiles Using LVI-PTV-GC/MSD. Agilent Technologies, 2008.
- Szelewski M. Synchronous SIM/Scan Low-Level PAH Analysis Using the Agilent 6890/5975 inert GC/MSD. Agilent Technologies, 2008.
- Prest H., Foote J. The Triple-Axis Detector: Attributes and Operating Advice. Agilent Technologies, 2008.
- Roushall R., Prest H. The 5975C Series MSDs: Method Optimization and Trace Ion Detection. Agilent Technologies, 2008.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Parts-per-Trillion Level Calibration of Semivolatiles Using LVI-PTV-GC/MSD
2007|Agilent Technologies|Applications
Parts-per-Trillion Level Calibration of Semivolatiles Using LVI-PTV-GC/MSD Application Environmental Author Mike Szelewski Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808 USA Abstract The analysis of semivolatiles in the parts-per-trillion range presents challenges due to analyte activity, background contamination, and…
Key words
ptv, ptvtemp, tempwashes, washessolv, solvvent, ventmsd, msdcryo, cryosemivolatiles, semivolatilesrtl, rtlpurge, purgecalibration, calibrationsim, simdichloromethane, dichloromethaneoven, ovensolvent
Environmental Semivolatiles Using an Agilent Multimode Inlet for Maximum Sensitivity
2009|Agilent Technologies|Applications
Environmental Semivolatiles Using an Agilent Multimode Inlet for Maximum Sensitivity Application Note Environmental Author Abstract Mike Szelewski The analysis of semivolatiles at very low levels presents challenges due to analyte Agilent Technologies, Inc. activity, background contamination and instrument sensitivity. Method…
Key words
mmi, mmiinlet, inletparameters, parameterssplitless, splitlesscold, coldhot, hotendrin, endrinmin, minsemivolatiles, semivolatileshold, holdbreakdown, breakdownsolvent, solventaldrin, aldrinnext, nextrrfs
Semivolatiles Retention Time Locked (RTL) Deconvolution Databases for Agilent GC/MSD Systems
2008|Agilent Technologies|Applications
Semivolatiles Retention Time Locked (RTL) Deconvolution Databases for Agilent GC/MSD Systems Application Environmental Author Mike Szelewski Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808 USA Abstract The G1677AA Semivolatiles Retention Time Locked database/library can provide rapid confirmation of environmental…
Key words
entry, entrydrs, drsmsd, msdcas, casamdis, amdisphthalate, phthalatebenzo, benzoconfigurator, configuratorlocked, lockedretention, retentionfluoranthene, fluoranthenename, nametime, timedbl, dblcompound
Reducing Analysis Time Using GC/MSD and Deconvolution Reporting Software
2007|Agilent Technologies|Applications
Reducing Analysis Time Using GC/MSD and Deconvolution Reporting Software Application Food and Flavors Authors Mike Grady, Steve Morrison, and Bob Deets Campbell Soup Company Campbell Place Camden, NJ 08103 USA Mike Szelewski Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE…
Key words
msd, msddrs, drswashes, washessolv, solvcampbell, campbellpermethrin, permethrindeconvolution, deconvolutionfewest, fewestcryo, cryolocked, lockednegatives, negativespositives, positivespostinj, postinjpreinj, preinjvent