Reducing Analysis Time Using GC/MSD and Deconvolution Reporting Software
Applications | 2007 | Agilent TechnologiesInstrumentation
Analysis of complex food matrices often requires multiple runs and detectors to confirm the presence of numerous pesticide residues. Employing gas chromatography coupled to a mass selective detector with deconvolution reporting software streamlines this process, delivering rapid multi-residue screening with high confidence and minimal false positives or negatives.
This application study aimed to demonstrate substantial time savings and maintained data quality by transitioning from traditional element-specific detectors to GC-MS with deconvolution reporting software (DRS). The work focused on market basket samples in a research laboratory setting, comparing detection capabilities, run times, and confirmatory power.
Samples of fruits and vegetables were spiked with internal and surrogate standards, extracted with acidified acetonitrile, centrifuged, and cleaned via solid-phase extraction. A programmable temperature vaporizing inlet in solvent-vent mode enabled large-volume injection. Separation was achieved on an HP-5MS column under constant pressure, followed by mass spectral acquisition at 35-500 amu and 3.12 scans per second. Deconvolution software combined ChemStation quantitation, AMDIS deconvolution, and NIST library searching to produce an expert-verified report in under two minutes.
Analysis of seven market samples revealed residues ranging from trace levels to 0.48 ppm. DRS confirmed all targets and identified additional uncalibrated compounds. Duplicate spikes in strawberries and tomatoes yielded recoveries close to expected values, demonstrating method accuracy and reproducibility. Deconvolution addressed coelution, enabling clean mass spectra matching even in highly complex chromatograms.
Ongoing developments may include expanded deconvolution libraries, integration with high-resolution MS for greater selectivity, and automated data review workflows. Such advances will support regulatory compliance, large-scale monitoring programs, and rapid response to emerging contaminants.
GC-MS with deconvolution reporting software offers a fast, reliable, and versatile solution for multi-residue screening in food and flavor matrices. By combining retention time locking, spectral deconvolution, and library searching, laboratories can achieve high throughput and robust confirmation with minimal false alarms.
GC/MSD, GC/SQ
IndustriesFood & Agriculture
ManufacturerAgilent Technologies
Summary
Significance of the topic
Analysis of complex food matrices often requires multiple runs and detectors to confirm the presence of numerous pesticide residues. Employing gas chromatography coupled to a mass selective detector with deconvolution reporting software streamlines this process, delivering rapid multi-residue screening with high confidence and minimal false positives or negatives.
Objectives and Study Overview
This application study aimed to demonstrate substantial time savings and maintained data quality by transitioning from traditional element-specific detectors to GC-MS with deconvolution reporting software (DRS). The work focused on market basket samples in a research laboratory setting, comparing detection capabilities, run times, and confirmatory power.
Methodology and Instrumentation
Samples of fruits and vegetables were spiked with internal and surrogate standards, extracted with acidified acetonitrile, centrifuged, and cleaned via solid-phase extraction. A programmable temperature vaporizing inlet in solvent-vent mode enabled large-volume injection. Separation was achieved on an HP-5MS column under constant pressure, followed by mass spectral acquisition at 35-500 amu and 3.12 scans per second. Deconvolution software combined ChemStation quantitation, AMDIS deconvolution, and NIST library searching to produce an expert-verified report in under two minutes.
Instrumentation Used
- Agilent 6890N GC with EPC-PTV inlet in solvent vent mode
- Agilent 5973N MSD upgraded with inert source and enhanced electronics
- HP-5MS column (30 m × 0.25 mm × 0.25 μm) under constant pressure
- Retention Time Locked database for 927 analytes
Main Results and Discussion
Analysis of seven market samples revealed residues ranging from trace levels to 0.48 ppm. DRS confirmed all targets and identified additional uncalibrated compounds. Duplicate spikes in strawberries and tomatoes yielded recoveries close to expected values, demonstrating method accuracy and reproducibility. Deconvolution addressed coelution, enabling clean mass spectra matching even in highly complex chromatograms.
Benefits and Practical Applications
- Reduced total analysis time by consolidating multi-detector runs into a single GC-MS method
- Increased throughput with a two-minute automated verification report
- Enhanced confidence through dual confirmation by retention time and spectral deconvolution
- Scalable method transfer across laboratories using retention time locking
- Capability to screen hundreds of compounds without extensive sample handling or additional standards
Future Trends and Potential Applications
Ongoing developments may include expanded deconvolution libraries, integration with high-resolution MS for greater selectivity, and automated data review workflows. Such advances will support regulatory compliance, large-scale monitoring programs, and rapid response to emerging contaminants.
Conclusion
GC-MS with deconvolution reporting software offers a fast, reliable, and versatile solution for multi-residue screening in food and flavor matrices. By combining retention time locking, spectral deconvolution, and library searching, laboratories can achieve high throughput and robust confirmation with minimal false alarms.
References
- Wylie P L, Szelewski M J, Meng C-K Comprehensive pesticide screening by GC-MSD using deconvolution reporting software Agilent Technologies Pub 5989-1157EN
- Anastassiades M, Lehotay S J, Stajnbaher D, Schenck F J Fast multiresidue method employing acetonitrile extraction and dispersive SPE for produce J AOAC Int 86 412-431 2003
- Japan Department of Food Safety Ministry of Health Labour and Welfare Analytical methods for residual agricultural chemicals feed additives and veterinary drugs in food
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Semivolatiles Retention Time Locked (RTL) Deconvolution Databases for Agilent GC/MSD Systems
2008|Agilent Technologies|Applications
Semivolatiles Retention Time Locked (RTL) Deconvolution Databases for Agilent GC/MSD Systems Application Environmental Author Mike Szelewski Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808 USA Abstract The G1677AA Semivolatiles Retention Time Locked database/library can provide rapid confirmation of environmental…
Key words
entry, entrydrs, drsmsd, msdcas, casamdis, amdisphthalate, phthalatebenzo, benzoconfigurator, configuratorlocked, lockedretention, retentionfluoranthene, fluoranthenename, nametime, timedbl, dblcompound
Determination of Allergens in Fragrance Products Using Agilent Deconvolution Reporting Software 
2008|Agilent Technologies|Applications
Determination of Allergens in Fragrance Products Using Agilent Deconvolution Reporting Software Application Brief Wei Luan, Chris Sandy, and Mike Szelewski Flavor and Fragrance The 7th amendment of the European Cosmetics Directive (2003/15/EC) was published in 2003. Manufacturers of cosmetics were…
Key words
fragrance, fragrancewashes, washescinnamaldehyde, cinnamaldehydeallergens, allergensrtl, rtllocked, lockedmol, molbenzyl, benzylpostinj, postinjpreinj, preinjsaver, saveralcohol, alcoholisomethyl, isomethylcitral, citralagilent
Screening for 926 Pesticides and Endocrine Disruptors by GC/MS with Deconvolution Reporting Software and a New Pesticide Library
2006|Agilent Technologies|Applications
Screening for 926 Pesticides and Endocrine Disruptors by GC/MS with Deconvolution Reporting Software and a New Pesticide Library Application Note Food and Environmental Authors USA endocrine disrupters in about two minutes per sample. Deconvolution helps identify pesticides that are buried…
Key words
methyl, methylester, esterdrs, drsphthalate, phthalatesulfone, sulfoneamdis, amdisoxon, oxonethyl, ethylisomer, isomeragilent, agilentdatabase, databasedeconvolution, deconvolutionpesticide, pesticidelibrary, librarycas
A Blind Study of Pesticide Residues in Spiked and Unspiked Fruit Extracts Using Deconvolution Reporting Software
2004|Agilent Technologies|Applications
A Blind Study of Pesticide Residues in Spiked and Unspiked Fruit Extracts Using Deconvolution Reporting Software Application Food Author Christopher P. Sandy Agilent Technologies, Inc. Block A, CSC Eskdale Road Winnersh Triangle Estate Wokingham, Berk RG41 5DZ United Kingdom Abstract…
Key words
drs, drspesticide, pesticideconfirmed, confirmedrtl, rtlspiked, spikedmsd, msddetected, detectedthree, threewere, werepesticides, pesticidesmethyl, methylamdis, amdisendosulfan, endosulfanhch, hchdeconvolution