Screening for Hazardous Chemicals in Homeland Security and Environmental Samples Using a GC/MS/ECD/FPD with a 731 Compound DRS Database

Applications | 2006 | Agilent TechnologiesInstrumentation
GC, GC/MSD, GC/SQ
Industries
Homeland Security
Manufacturer
Agilent Technologies

Summary

Význam tématu


Rapid and accurate screening of hazardous chemicals is essential for homeland security and environmental monitoring. Gas chromatography–mass spectrometry (GC/MS) is a versatile tool but faces challenges with unknown targets, complex matrices, and high sample throughput. Integrating selective detectors (electron capture, flame photometric), microfluidic splitting, retention time locking, synchronous SIM/Scan acquisition, and automated deconvolution improves both speed and reliability.

Cíle a přehled studie / článku


This work presents an Agilent GC/MS system equipped with a microfluidic splitter routing column effluent to an MSD and either a dual-wavelength flame photometric detector or μECD. A 731-compound Hazardous Chemicals Database Library (HCD) with retention time locking is paired with Deconvolution Reporting Software (DRS) to achieve rapid screening and minimize false positives/negatives. Methods are developed in three speed scales (1X, 3X, 7X) to balance resolution, capacity, and analysis time.

Použitá metodika a instrumentace


  1. GC: Agilent 6890N with split/splitless inlet, constant pressure mode, fast oven ramps, optional backflushing using AUX EPC.
  2. Detektory: MSD (Agilent 5975 inert or 5973N), μECD for halogenated analytes, single/dual FPD for phosphorus and sulfur species.
  3. Microfluidic splitter: three-way (MSD/FPD/ECD) or two-way (MSD/DFPD) diffusion-bonded flow division under constant makeup flow.
  4. Data acquisition: synchronous SIM/Scan on ChemStation rev. D.02.00+, deconvolution via AMDIS within DRS.
  5. HCD database: 731 individual compounds covering volatiles, semivolatiles, dioxins/furans, PCBs, and an expanded pesticide library.

Hlavní výsledky a diskuse


  1. Ambient headspace GC/MS screening of 14 halogenated volatiles in water at 2 ppm and 100 ppb, with and without 100 ppm gasoline matrix. ECD traces revealed all targets; DRS delivered high-confidence identifications where conventional scan quantitation failed.
  2. Synchronous SIM/Scan increased target ion signal-to-noise from ~34 to >400 for low-level analytes, restoring qualifier detection and improving sensitivity.
  3. Fast chromatography: 1X method (43 min), 3X (14.3 min), 7X (6.1 min) with retention times matching database predictions within ±0.05 min, enabling flexible throughput.
  4. Pesticide screening: nine organophosphorus compounds (50 ppm) in a 50 000 ppm fuel matrix. At 7X speed, DRS combined with FPD/ECD detected all analytes with high match scores and limited false positives.

Přínosy a praktické využití metody


  • Simultaneous MS and selective detector data allow robust compound confirmation and detection of non-target species.
  • Retention time locking and automated deconvolution reduce calibration workload and matrix interferences.
  • Synchronous SIM/Scan accelerates trace screening for high-priority agents without additional runs.
  • Scalable methods support both emergency response and routine environmental analysis.

Budoucí trendy a možnosti využití


  • Extending the HCD to include emerging contaminants and chemical warfare agents.
  • Integrating automated sample preparation and high-throughput techniques (e.g., SPME, online SPE).
  • Applying machine-learning algorithms to deconvolution outputs for enhanced specificity.
  • Developing field-portable GC/MS platforms with onboard retention time locking and deconvolution.

Závěr


The integrated GC/MS/ECD/FPD approach with retention time locking, synchronous SIM/Scan, and DRS provides a powerful, high-speed screening platform for a broad spectrum of hazardous chemicals in complex matrices. Method flexibility and combined detector signals deliver rapid, reliable identification and semiquantitation for homeland security and environmental applications.

Reference


  1. Retention Time Locking: Concepts and Applications, Agilent Technologies, publication 5966-2469E.
  2. Identifying Pesticides with Full Scan, SIM, μECD, and FPD from a Single Injection, Agilent Technologies, publication 5989-3299EN.
  3. Improving Productivity with Synchronous SIM/Scan, Agilent Technologies, publication 5989-3108EN.
  4. Comprehensive Pesticide Screening by GC/MSD using Deconvolution Reporting Software, Agilent Technologies, publication 5989-1157EN.
  5. Precise Time-Scaling of Gas Chromatographic Methods Using Method Translation and Retention Time Locking, Agilent Technologies, publication 5967-5820E.
  6. New Tools for Rapid Pesticide Analysis in High Matrix Samples, Agilent Technologies, publication 5989-1716EN.
  7. Ambient Headspace GC and GC-MSD Analysis of Nonpolar Volatiles in Water, Agilent Technologies, publication 5968-9455E.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Excellent choices for environmental applications
Excellent choices for environmental applications
2009|Agilent Technologies|Guides
Solutions that meet your demands for: speed accuracy productivity Excellent choices for environmental applications Productivity Tools Applications > Return to Table of Contents > Search entire document Integrated Second Peristaltic Pump for Improved Sample Throughput and Reduced Matrix Effects Technical…
Key words
rtl, rtlpressure, pressuretime, timecolumn, columnaed, aedlocking, lockingmethod, methodagilent, agilentrtlock, rtlockretention, retentionicp, icpinlet, inletcan, canmsd, msdflow
Replacing Multiple 50-Minute GC and GC-MS/SIM Analyses with One 15-Minute Full-Scan GC-MS Analysis for Nontargeted Pesticides Screening and >10x Productivity Gain
Replacing Multiple 50-Minute GC and GC-MS/SIM Analyses with One 15-Minute Full-Scan GC-MS Analysis for Nontargeted Pesticides Screening and >10x Productivity Gain Application Food Safety Author Introduction Chin-Kai Meng and Mike Szelewski Agilent Technologies 2850 Centerville Road Wilmington, DE 19808 To…
Key words
cfsan, cfsanµecd, µecddeconvolution, deconvolutionfda, fdasplitter, splitteramdis, amdisspectrum, spectrumtid, tidpeach, peachginseng, ginsengagilent, agilentfpd, fpdthree, threelibrary, libraryflow
Identifying Pesticides with Full Scan, SIM, μECD, and FPD from a Single Injection
Identifying Pesticides with Full Scan, SIM, µECD, and FPD from a Single Injection Application Food Safety, Environmental Authors Chin-Kai Meng and Bruce Quimby Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract In this application note, a gas…
Key words
fpd, fpdµecd, µecdsplitter, splitterscan, scansim, simepc, epcrestrictor, restrictormsd, msdway, waylibrary, librarysplit, splittemp, tempmode, modefull, fullconstant
Screening for 926 Pesticides and Endocrine Disruptors by GC/MS with Deconvolution Reporting Software and a New Pesticide Library
Screening for 926 Pesticides and Endocrine Disruptors by GC/MS with Deconvolution Reporting Software and a New Pesticide Library Application Note Food and Environmental Authors USA endocrine disrupters in about two minutes per sample. Deconvolution helps identify pesticides that are buried…
Key words
methyl, methylester, esterdrs, drsphthalate, phthalatesulfone, sulfoneamdis, amdisoxon, oxonethyl, ethylisomer, isomeragilent, agilentdatabase, databasedeconvolution, deconvolutionpesticide, pesticidelibrary, librarycas