Preparing Gas Phase Standards from Neat Chemicals to Support Air Monitoring in the Workplace
Applications | 2016 | ENTECHInstrumentation
Accurate calibration standards are essential for reliable workplace air monitoring by gas chromatography–mass spectrometry (GC-MS). Commercial high-pressure cylinder mixtures typically deliver up to 1 PPMv, leaving a gap for mid- to high-PPMv applications aligned with OSHA Permissible Exposure Limits (PELs). The described approach empowers laboratories to generate stable, custom gas-phase standards from neat liquids, covering a broad concentration range and ensuring precise instrument calibration and quality assurance.
This application note demonstrates a step-by-step workflow for preparing low- to high-PPMv standards using Entech ESP 2.0 software, inert canisters, and a digital dilution system. It targets:
The workflow consists of:
The protocol enabled generation of calibration curves spanning 0.05–2 × PELs for compounds such as methylene chloride, toluene, and 2-hexanone. Single-component standards (e.g., BFT in He) achieved 77.8 PPMv in 6 L canisters, with subsequent dilution to 1 PPMv for filling over 400 HDS monitors per batch. Equilibration times of 2–3 h ensured homogeneity, while weighting factors allowed tailoring concentrations relative to toxicity and linear range constraints of MS detectors.
Advancements may include automation of dosing via robotic handling, integration of cloud-based compound libraries, expansion to isotopically labeled surrogates for matrix-specific accuracy, and miniaturized sampling devices. Further development of ESP 2.0 could support direct upload of calibration data into LIMS and real-time monitoring networks.
The described method provides a versatile, reproducible, and economical solution for generating gas-phase calibration standards from neat chemicals. By leveraging software-guided calculations and inert sampling media, laboratories can address a wide array of workplace air monitoring needs, ensuring compliance, data integrity, and operational efficiency.
Sample Preparation
IndustriesEnvironmental
ManufacturerENTECH
Summary
Significance of the Topic
Accurate calibration standards are essential for reliable workplace air monitoring by gas chromatography–mass spectrometry (GC-MS). Commercial high-pressure cylinder mixtures typically deliver up to 1 PPMv, leaving a gap for mid- to high-PPMv applications aligned with OSHA Permissible Exposure Limits (PELs). The described approach empowers laboratories to generate stable, custom gas-phase standards from neat liquids, covering a broad concentration range and ensuring precise instrument calibration and quality assurance.
Objectives and Study Overview
This application note demonstrates a step-by-step workflow for preparing low- to high-PPMv standards using Entech ESP 2.0 software, inert canisters, and a digital dilution system. It targets:
- Creating multi-component liquid cocktails from neat compounds
- Generating high-concentration gas standards in Silonite-coated canisters
- Performing secondary dilutions into canisters or Bottle-Vac samplers
- Supporting calibration of GC-MS and HDS Personal Monitors for workplace air analysis
Methodology and Instrumentation
The workflow consists of:
- Cocktail design: ESP 2.0 calculates volumes and weights to transfer equal mole amounts or user-defined weighting factors per analyte.
- Primary gas fill: Injected cocktail into a 6 L Silonite-coated canister at 2–4 psig, then pressurized to target absolute pressure and equilibrated.
- Secondary dilution: Transferring defined pressures into smaller canisters or Bottle-Vacs using the Entech Digital Dilution System (DDS) to achieve lower PPM levels.
- HDS Personal Monitor calibration: Preparing a 1 PPMv Bromofluorotoluene surrogate in helium, charging 50 cc monitors, and applying an internal standard for inertness verification.
Used Instrumentation
- Entech ESP 2.0 Standards Preparation Software
- Silonite-coated stainless steel canisters (6 L, 15 L) and Bottle-Vac samplers
- Entech 4700 Precision Diluter and Digital Dilution System
- GC-MS systems configured for TO-15 and HDS monitor analysis
Key Results and Discussion
The protocol enabled generation of calibration curves spanning 0.05–2 × PELs for compounds such as methylene chloride, toluene, and 2-hexanone. Single-component standards (e.g., BFT in He) achieved 77.8 PPMv in 6 L canisters, with subsequent dilution to 1 PPMv for filling over 400 HDS monitors per batch. Equilibration times of 2–3 h ensured homogeneity, while weighting factors allowed tailoring concentrations relative to toxicity and linear range constraints of MS detectors.
Benefits and Practical Applications
- Cost-effective use of neat reagents versus expensive high-pressure cylinders
- Flexibility to include hundreds of analytes not commercially available
- Custom concentration ranges aligned with regulatory limits
- Enhanced QA/QC through surrogate internal standards and inert sampler verification
- Scalable workflows supporting sub-PPB to high-PPMv calibration
Future Trends and Opportunities
Advancements may include automation of dosing via robotic handling, integration of cloud-based compound libraries, expansion to isotopically labeled surrogates for matrix-specific accuracy, and miniaturized sampling devices. Further development of ESP 2.0 could support direct upload of calibration data into LIMS and real-time monitoring networks.
Conclusion
The described method provides a versatile, reproducible, and economical solution for generating gas-phase calibration standards from neat chemicals. By leveraging software-guided calculations and inert sampling media, laboratories can address a wide array of workplace air monitoring needs, ensuring compliance, data integrity, and operational efficiency.
References
- Cardin D., Bossart J. Preparing Gas Phase Standards from Neat Chemicals to Support Air Monitoring in the Workplace. Application Note A-3740-01, Entech Instruments.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Helium Diffusion Sampling for Personal Monitoring by EPA Method TO15
|ENTECH|Applications
Helium Diffusion Sampling for Personal Monitoring by EPA Method TO15 INSTRUMENTS INC. Application Note: A-3737-03 Authors Daniel J. Cardin Thomas X. Robinson Daniel B. Cardin Entech Instruments, Inc. Simi Valley, CA 93065 Abstract A new, simplified approach for filling air…
Key words
hds, hdssampling, samplingpersonal, personalsampler, samplersamplers, samplersdiffusion, diffusionhelium, heliummonitors, monitorsweight, weightentech, entechair, aircanister, canisterevacuated, evacuatedmonitor, monitorprior
HDS™ Personal Monitor - The Next Generation in IH Monitoring
2019|ENTECH|Brochures and specifications
HDS™ Personal Monitor HDS PERSONAL Monitor The Next Generation in IH Monitoring. Laboratory Verification of Sample Collection Although the HDS™ Personal Monitors are very consistent, there is no reason to have to assume proper sample collection for analysis as with…
Key words
hds, hdspersonal, personalmonitor, monitorsampling, samplingentech, entechbft, bfthelium, heliummonitors, monitorssurrogate, surrogatediacetyl, diacetylunaffected, unaffectedactively, activelyfield, fieldcollection, collectionntroducing
Improving Canister Sampling Accuracy and Reliability Using Helium Diffusion Sampling into Silonite Coated Stainless Steel Canisters
|Agilent Technologies|Applications
Improving Canister Sampling Accuracy and Reliability Using Helium Diffusion Sampling into Silonite Coated Stainless Steel Canisters INSTRUMENTS INC. Application Note: A-3731-01 Authors Daniel B. Cardin Thomas X. Robinson Daniel J. Cardin Entech Instruments, Inc. Simi Valley, CA 93065 Abstract A…
Key words
sampling, samplingcanister, canisterhds, hdsdiffusion, diffusioncanisters, canistersair, airhelium, heliumover, overgauge, gaugesampler, samplervacuum, vacuumsilonite, silonitevac, vacatmospheric, atmosphericorifice
Sampling with MiniCans and Bottle-Vacs
|ENTECH|Applications
Sampling with MiniCans and Bottle-Vacs Application Note: A-3730-01 Authors Dan Cardin Entech Instruments, Inc. Simi Valley, CA 93065 Contents: Introduction MiniCans or Bottle-Vacs? Selecting Container Size - Required Detection Limits - Sample IntegrationTime Quick “Grab” Sampling Step-by-Step Grab Sampling ShortTerm…
Key words
sampling, samplingbottle, bottlecanisters, canistersstep, stepminicans, minicansvacs, vacsgrab, grabvac, vacminican, minicanorifice, orificesamplers, samplersvalve, valveentech, entechcanister, canisterhds