Agilent JetClean: In-situ GC/MS Ion Source Cleaning and Conditioning

Technical notes | 2016 | Agilent TechnologiesInstrumentation
GC/MSD, GC/SQ, GC/QQQ
Industries
Manufacturer
Agilent Technologies

Summary

Importance of Topic


Gas chromatography–mass spectrometry (GC/MS) is a cornerstone technique in analytical chemistry for trace and target compound analysis. Maintaining optimal performance of the GC inlet, column, and ion source is critical to preserve sensitivity, reproducibility, and instrument uptime. Traditional manual source cleaning interrupts workflows and requires significant operator time. The Agilent JetClean system introduces an in situ hydrogen‐driven cleaning and conditioning approach that minimizes downtime and extends maintenance intervals.

Objectives and Overview


This application note presents two JetClean operating modes—Acquire & Clean and Clean Only—and evaluates their impact on GC/MS performance. The objectives are to demonstrate how in situ hydrogen treatment reduces manual cleaning frequency, sustains detection limits and response factors across many sample batches, and streamlines laboratory workflows.

Methods and Instrumentation


  • Instruments: Agilent 5975, 5977A/B, 7000B/C and 7010 GC/MS systems equipped with the JetClean ion source assembly.
  • Acquire & Clean mode: hydrogen flows 0.13–0.53 mL/min delivered continuously during sample acquisition; emission current adjusted 10–35 µA.
  • Clean Only mode: offline hydrogen treatment up to 3.5 mL/min with emission currents up to 100 µA; temperature range 150–350 °C; duration 1–120 min.
  • Monitoring: use scan mode (45–300 m/z) or SIM for key ions (e.g., m/z 57, 91) to track residue removal kinetics.
  • Safety and configuration: hydrogen leak testing, line purging, auto‐tune validation for low air and water backgrounds.

Main Results and Discussion


  • Acquire & Clean mode extended the interval between manual MS service from approximately 25 to 90 sample batches while maintaining response factors near initial calibration levels.
  • Clean Only mode rapidly removed hydrocarbon and dye deposits from the ion source drawout lens, restoring optical clarity and signal stability within minutes.
  • Optimizing hydrogen flow and emission settings balanced cleaning efficiency against potential spectral artifacts; lower flows preserved fragile compound spectra, while higher settings accelerated residue removal.

Benefits and Practical Applications


  • Substantial reduction in manual maintenance time and instrument downtime.
  • Improved reproducibility and extended consumable life for ion sources and GC columns.
  • Ability to integrate source cleaning into automated sample sequences for preventive maintenance guided by performance metrics.

Future Trends and Possibilities


Emerging developments may include adaptive, data‐driven hydrogen cleaning protocols that respond to real‐time performance indicators and predictive maintenance scheduling through LIMS integration. Expansion of in situ cleaning strategies to other ionization sources and mass spectrometer platforms is anticipated.

Conclusion


Agilent JetClean offers a robust, flexible solution for GC/MS source maintenance, combining concurrent hydrogen cleaning during acquisition with standalone treatment cycles. This approach extends manual service intervals, preserves analytical performance, and reduces operator effort compared to traditional manual cleaning methods.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Maintaining Sensitivity and Reproducibility with the JetCleanSelf-Cleaning Ion Source for Pesticides in Food and Feed
Maintaining Sensitivity and Reproducibility with the JetClean Self-Cleaning Ion Source for Pesticides in Food and Feed Jessica Westland1; Vivian Chen2 and Elizabeth Almasi3; 1Agilent Technologies, Wilmington, DE 19808, USA 2Agilent Technologies, Shanghai Company Limited; 200131,CHINA; 3Agilent ASMS 2016 TP 216…
Key words
jetclean, jetcleansource, sourceendosulfan, endosulfanepoxide, epoxidemethyl, methylheptachlor, heptachlorendrin, endrinstd, stdacquire, acquireirgarol, irgaroltune, tunevivian, viviantemperature, temperatureelizabeth, elizabethdbcp
Maintaining Sensitivity and Reproducibility with the JetClean Self Cleaning Ion Source for Pesticides in Food and Feed
Maintaining Sensitivity and Reproducibility with the JetClean Self-Cleaning Ion Source for Pesticides in Food and Feed NACRW 2016 P-064 Jessica Westland1 and Elizabeth Almasi2 1Agilent Technologies, Wilmington, DE 19808 USA; 2Agilent Technologies, Santa Clara, CA 95051 USA Introduction Experimental Operation…
Key words
jetclean, jetcleansource, sourceendosulfan, endosulfanmethyl, methyldbcp, dbcpstd, stdheptenophos, heptenophoschlorobenzilate, chlorobenzilatepesticides, pesticidesparathion, parathionhexachlorobenzene, hexachlorobenzeneepoxide, epoxideheptachlor, heptachlorendrin, endrinacquire
Maintaining Sensitivity and Reproducibility with the Agilent JetClean Self-Cleaning Ion Source for Pesticides in Food and Feed
Maintaining Sensitivity and Reproducibility with the Agilent JetClean Self-Cleaning Ion Source for Pesticides in Food and Feed Application Note Author Abstract Jessica Westland Approximately 200 various pesticides were analyzed in organic honey extract on Agilent Technologies, Inc. the Agilent 7010A…
Key words
jetclean, jetcleanagilent, agilenthoney, honeymin, mintime, timewithout, withoutcleaning, cleaningself, selfmasshunter, masshunterendrin, endrinpuu, puusource, sourcetransitions, transitionsuse, usepesticide
JetClean Self cleaning ion source
JetClean Self cleaning ion source
2016|Agilent Technologies|Presentations
ANY ANALYTE, ANY METHOD, ONE PARTNER ONLY AGILENT COVERS ALL YOUR MASS SPEC NEEDS July 13, 2016 Page 1 JetClean Self-cleaning ion source Elizabeth Almasi Applications Manager June 9, 2016 What are your plans for the weekend? Cleaning is a…
Key words
jetclean, jetcleancleaning, cleaningjanuary, januaryoctober, octoberapril, aprilaugust, augustseptember, septembernovember, novembermarch, marchfebruary, februaryjune, juneclean, cleandecember, decemberself, selfsource
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike