Analysis of Volatile Organic Compounds in Water Using Dynamic Headspace-GCMS in Accordance with US EPA Method 8260D
Posters | 2024 | Shimadzu | ASMSInstrumentation
Volatile organic compounds (VOCs) are pervasive environmental contaminants that can infiltrate groundwater and pose significant human health risks, including carcinogenic and neurotoxic effects. Reliable quantitation of VOCs in water is essential for regulatory compliance and safeguarding public health under frameworks such as US EPA Method 8260D.
This study assesses a dynamic headspace autosampler with a cold‐trap interface coupled to a single quadrupole GC-MS, evaluating its performance as an alternative to traditional purge-and-trap methodology for the analysis of 65 target VOCs in water. Key aims include demonstrating sensitivity, reproducibility, and adherence to US EPA QC requirements.
The dynamic headspace GC-MS approach offers enhanced sensitivity, streamlined sample preparation and lower carryover risk compared to traditional purge-and-trap. Compliance with stringent QC metrics and reliable low-level quantitation make it well suited for environmental monitoring laboratories, QA/QC programs and industrial water testing.
Ongoing developments in headspace automation, novel trap materials and high-resolution mass spectrometry promise further improvements in detection limits and analyte breadth. Integration of AI-driven data processing could expedite routine analysis and enable real‐time monitoring of emerging contaminants in diverse water matrices.
The study confirms that dynamic headspace autosampling coupled with single quadrupole GC-MS provides a robust, sensitive and EPA-compliant alternative to purge-and-trap for VOC analysis in water. It delivers comparable performance with simplified workflow and high-throughput potential for regulatory and research applications.
HeadSpace, GC/MSD, GC/SQ
IndustriesEnvironmental
ManufacturerShimadzu
Summary
Importance of the topic
Volatile organic compounds (VOCs) are pervasive environmental contaminants that can infiltrate groundwater and pose significant human health risks, including carcinogenic and neurotoxic effects. Reliable quantitation of VOCs in water is essential for regulatory compliance and safeguarding public health under frameworks such as US EPA Method 8260D.
Objectives and study overview
This study assesses a dynamic headspace autosampler with a cold‐trap interface coupled to a single quadrupole GC-MS, evaluating its performance as an alternative to traditional purge-and-trap methodology for the analysis of 65 target VOCs in water. Key aims include demonstrating sensitivity, reproducibility, and adherence to US EPA QC requirements.
Methodology and instrument setup
- Instrument: Shimadzu GCMS-QP2020 NX with HS-20 NX dynamic headspace module (trap model).
- Headspace parameters: 60 °C oven, –10 °C trap cooling, 250 °C desorption, 10 injection cycles for analyte enrichment.
- GC conditions: Helium carrier gas, split ratio 5:0, temperature program from 35 °C to 220 °C.
- MS detection: Selected ion monitoring (SIM) mode; source at 230 °C, interface at 220 °C; data acquired in LabSolutions GCMS with AI‐powered peak integration and Environmental QC options.
- Calibration and QC: Six‐point quadratic calibration (0.5–20 ng/mL) with internal standards and surrogates at 5 ng/mL; QC checks per US EPA Method 8260D including ICAL, ICV and CCV.
Results and discussion
- BFB tuning met all relative abundance criteria for mass fragments, confirming instrument readiness.
- Calibration performance: 64 of 65 VOCs passed initial calibration (ICAL), 61 passed initial calibration verification (ICV), and 59 passed continuing calibration verification (CCV).
- Lower limit of quantitation (LLOQ): Achieved 0.5 ng/mL for 59 analytes with signal-to-noise ratios above 10, ensuring trace-level quantitation.
- Method detection limits (MDLs): Calculated MDLs at 0.5 ng/mL and 1.0 ng/mL ranged approximately from 0.01 to 1.26 ng/mL across the analyte panel.
- No significant carryover or contamination observed in method blanks, validating system cleanliness.
Benefits and practical applications
The dynamic headspace GC-MS approach offers enhanced sensitivity, streamlined sample preparation and lower carryover risk compared to traditional purge-and-trap. Compliance with stringent QC metrics and reliable low-level quantitation make it well suited for environmental monitoring laboratories, QA/QC programs and industrial water testing.
Future trends and potential applications
Ongoing developments in headspace automation, novel trap materials and high-resolution mass spectrometry promise further improvements in detection limits and analyte breadth. Integration of AI-driven data processing could expedite routine analysis and enable real‐time monitoring of emerging contaminants in diverse water matrices.
Conclusion
The study confirms that dynamic headspace autosampling coupled with single quadrupole GC-MS provides a robust, sensitive and EPA-compliant alternative to purge-and-trap for VOC analysis in water. It delivers comparable performance with simplified workflow and high-throughput potential for regulatory and research applications.
Reference
- US EPA Method 8260D: Volatile Organic Compounds by Gas Chromatography/Mass Spectrometry, Revision 4, February 2017.
- Definition and Procedure for the Determination of the Method Detection Limit, Federal Register, 1984, Appendix B to Part 136.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of Volatile Organic Compounds (VOCs) in Water Using Trap-Headspace-GCMS in Accordance with US EPA Method 8260D Criteria
2024|Shimadzu|Applications
HS-20 NX (Trap Model) Headspace Autosampler GCMS-QP2020 NX Single Quadrupole Gas Chromatograph Mass Spectrometer Application News Analysis of Volatile Organic Compounds (VOCs) in Water Using Trap-Headspace-GCMS in Accordance with US EPA Method 8260D Criteria Elvi Horiyanto, Elgin Ting, Chua Chun…
Key words
loq, loqratio, ratioarea, areaical, icalpass, passtrap, trapheadspace, headspacetap, tapgcms, gcmstuning, tuningicv, icvvocs, vocsbfb, bfbunknown, unknownmdl
A consolidated method for the analysis of VOCs in soil by HS-GC-MS in analytical testing laboratories
2020|Thermo Fisher Scientific|Applications
APPLICATION NOTE 73742 A consolidated method for the analysis of VOCs in soil by HS-GC-MS in analytical testing laboratories Authors: David Lee and Cristian Cojocariu Thermo Fisher Scientific, Runcorn, UK Keywords: Gas chromatography, GC, valve and loop, headspace, TriPlus 500…
Key words
butylbenzene, butylbenzenefluorobenzene, fluorobenzenexylene, xylenechloroethene, chloroethenebromofluorobenzene, bromofluorobenzenesoil, soilchlorobenzene, chlorobenzenename, namedichlorodifluoromethane, dichlorodifluoromethaneinternal, internaltrichlorofluoromethane, trichlorofluoromethanedibromomethane, dibromomethanebromobenzene, bromobenzenechloroethane, chloroethanetetrachloroethene
Shimadzu Guide to US EPA Method 8260 for Analysis of Volatile Organic Compounds in Ground Water and Solid Waste
2015|Shimadzu|Applications
No. SSI-GCMS-1503 Gas Chromatograph Mass Spectrometer No. GCMS-1503 Shimadzu Guide to US EPA Method 8260 for Analysis of Volatile Organic Compounds in Ground Water and Solid Waste ■ Introduction Environmental contamination has been at the forefront of government policy and…
Key words
pass, passistd, istdacetate, acetatebfb, bfbtert, tertbutylbenzene, butylbenzenerestek, restekccv, ccvrsd, rsdmethyl, methylsurr, surrbutyl, butylvinyl, vinylethyl, ethylamyl
Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS
2004|PerkinElmer|Applications
w w w. p e r k i n e l m e r. c o m This application note demonstrates the use of a new sample-introduction technology (incorporated in the PerkinElmer® TurboMatrix™ HS-110 Trap) for U.S. EPA Method 8260B.…
Key words
target, targetheadspace, headspacetrap, trapspccs, spccsspcc, spccstandard, standardccc, cccinternal, internalcheck, checkrrf, rrfsurrogate, surrogatebromoform, bromoformperkinelmer, perkinelmerrsd, rsdcompounds