GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

GC Analysis of Halogenated Pesticides using the Agilent J&W FactorFour VF-1701ms with EZ-Guard

Applications | 2010 | Agilent TechnologiesInstrumentation
GC, GC columns, Consumables
Industries
Manufacturer
Agilent Technologies

Summary

Importance of the topic


Pesticide residues in environmental and food samples pose significant health and ecological concerns, demanding reliable analytical methods.
Gas chromatography with electron capture detection (GC-ECD) remains a gold standard for halogenated pesticide analysis, but complex matrices can contaminate columns and degrade analytes.
Integrating a guard column directly into the analytical system enhances robustness, extends column lifetime, and preserves sensitivity.

Objectives and overview of the study


This application note investigates the performance of the Agilent J&W FactorFour VF-1701ms capillary column equipped with an EZ-Guard integrated guard for the separation of 12 halogenated pesticides.
The goal is to demonstrate improved peak shapes, high detector response, and extended column life when analyzing challenging sample matrices.

Methodology and instrumentation


The study employed the following conditions:
  • Column: Agilent J&W FactorFour VF-1701ms, 30 m × 0.25 mm ID, 0.25 µm film thickness, with 10 m EZ-Guard (part number CP9177)
  • Sample: 2 µL injection of a 0.1 % halogenated pesticide mix in hexane, split ratio 1:50
  • Oven temperature: isothermal at 220 °C
  • Carrier gas: hydrogen at 90 kPa
  • Injector temperature: 275 °C
  • Detector: ECD set at 300 °C

Main results and discussion


Chromatographic separation of 12 pesticides, including hexachlorobenzene, HCH isomers, aldrin, dieldrin, endrin, DDD, DDE, DDT isomers, and endosulfan isomers, exhibited sharp, symmetric peaks and high detector responses.
The EZ-Guard integration prevented leakage and adsorption, maintaining column inertness even for labile compounds like endrin and p,p'-DDT.
The final uncoated meter of the guard column functioned as an integrated transfer line, accelerating detector stabilization.

Benefits and practical applications of the method


The VF-1701ms column with EZ-Guard offers:
  • Extended analytical column lifetime by trapping non-volatile matrix components
  • Elimination of connector leakage and analyte adsorption
  • Consistent high sensitivity and peak symmetry across multiple injections
  • Rapid and straightforward guard replacement to restore performance
  • Improved laboratory productivity and reduced downtime

This configuration is particularly suited for routine QA/QC laboratories analyzing complex environmental and food matrices.

Future trends and opportunities for use


Emerging directions include coupling integrated guard columns with GC-MS for enhanced selectivity, expanding multi-residue pesticide methods, and integrating automated sampling systems.
Adoption of green carrier gases and miniaturized column formats may further boost sustainability and throughput.

Conclusion


The Agilent J&W FactorFour VF-1701ms column with EZ-Guard delivers robust, high-performance separation of halogenated pesticides in challenging matrices.
This approach minimizes maintenance, maintains sensitivity, and increases laboratory efficiency, making it an ideal choice for routine pesticide monitoring.

Instrumentation used


This study utilized an Agilent gas chromatograph fitted with an ECD detector, a VF-1701ms capillary column with integrated EZ-Guard, hydrogen carrier gas, and standard split injection accessories.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis of Halogenated Environmental Pollutants Using Electron Capture Detection
TECHNICAL GUIDE Analysis of Halogenated Environmental Pollutants Using Electron Capture Detection • Sample preparation techniques for liquid, solid, and biota samples. • Chromatographic methods and best practices for halogenated pollutants. • Includes analysis of chlorinated pesticides, PCBs, chlorinated herbicides, haloacetic…
Key words
clpesticides, clpesticidesrtx, rtxchlorinated, chlorinatedpress, presscolumns, columnsextraction, extractiontight, tightconfirmational, confirmationalbhc, bhckit, kitinjection, injectionspe, speendrin, endrinresprep, resprepconnectors
Organochlorine pesticides
Organochlorine pesticides
2011|Agilent Technologies|Applications
Organochlorine pesticides Application Note Environmental Authors Introduction Agilent Technologies, Inc. GC analysis of 16 organochlorine pesticides according to EPA 625 is achieved in less than 23 minutes using Agilent VF-5ms EZ-Guard columns. Conditions Technique : GC Column : Agilent FactorFour…
Key words
organochlorine, organochlorinecourtesy, courtesypesticides, pesticidesprinted, printedinjector, injectorusa, usaauthors, authorsprogram, programenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationpressure, pressuresolvent
Organochlorine Pesticides Analysis in Water by GC/ECD - Consumable workflow ordering guide
Organochlorine Pesticides Analysis in Water by GC/ECD Consumable workflow ordering guide Confidently Identify and Quantify Organochlorine Pesticides Organochlorine pesticides are common target analytes because they persist in soil or sediment, which affects water source quality. They are often measured following…
Key words
organochlorine, organochlorinepesticides, pesticidespesticide, pesticidesurrogate, surrogatestandard, standardmix, mixsupplies, suppliestoxaphene, toxaphenepath, pathmin, minbhc, bhcbreakdown, breakdownchlordane, chlordaneendrin, endrinagilent
Fast and High Throughput GC Analysis of Pesticides
Fast and High Throughput GC Analysis of Pesticides
2010|Agilent Technologies|Applications
Fast and High Throughput GC Analysis of Pesticides Technical Overview Introduction Using large volume injection and rapid mass spectroscopy with an Agilent Rapid-MS column reduces the length of the analysis by up to five times when compared to traditional approaches.…
Key words
hch, hchrapid, rapidheptachlor, heptachlorcolumn, columnrestrictor, restrictorepoxide, epoxidepenatchlorobenzene, penatchlorobenzeneendosulfan, endosulfankcounts, kcountscarbofrit, carbofrittelodrin, telodrinpesticide, pesticideagilent, agilentreduces, reducesalfa
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike