Fast and High Throughput GC Analysis of Pesticides
Applications | 2010 | Agilent TechnologiesInstrumentation
Gas chromatography analysis of pesticide residues is essential for environmental monitoring, food safety and regulatory compliance. Techniques that combine rapid separation with enhanced sensitivity help laboratories increase sample throughput while maintaining low detection limits and reliable compound identification.
The study demonstrates how large volume injection (LVI) coupled with an Agilent Rapid-MS pesticide column and fast mass spectrometry reduces analysis time by up to fivefold compared to conventional GC methods. It evaluates separation efficiency, detection limits and overall performance in a multiresidue pesticide mixture.
The Rapid-MS approach achieved baseline separation of 30 target compounds in under 12 minutes, with PCB180 eluting last at 11.54 minutes. Large volume injection improved sensitivity, delivering detection limits at sub-ppb levels. The MS vacuum effect reduced column bleed and thermal decomposition, enhancing signal-to-noise ratios. Chromatographic efficiency remained high despite carrier gas velocities far above conventional limits.
Integration of ultrafast GC with tandem MS architectures and automated sample preparation will further boost throughput and selectivity. Emerging ionization techniques and miniaturized column formats may enable in-field or on-site pesticide monitoring. Data analytics and cloud-based reporting will streamline regulatory workflows.
The combination of large volume injection, Agilent Rapid-MS column technology and fast single-quadrupole MS delivers a powerful analytical platform for multiresidue pesticide analysis. Laboratories benefit from drastically reduced runtimes, improved sensitivity and robust quantitation, meeting the growing demand for high-throughput, trace-level environmental and food safety testing.
GC/MSD, GC/IT
IndustriesManufacturerAgilent Technologies
Summary
Significance of the Topic
Gas chromatography analysis of pesticide residues is essential for environmental monitoring, food safety and regulatory compliance. Techniques that combine rapid separation with enhanced sensitivity help laboratories increase sample throughput while maintaining low detection limits and reliable compound identification.
Objectives and Study Overview
The study demonstrates how large volume injection (LVI) coupled with an Agilent Rapid-MS pesticide column and fast mass spectrometry reduces analysis time by up to fivefold compared to conventional GC methods. It evaluates separation efficiency, detection limits and overall performance in a multiresidue pesticide mixture.
Methodology and Instrumentation
- Column configuration: Agilent 0.53 mm ID Rapid-MS pesticide column (CP8138) with a 2.5 m retention gap and restrictor for high velocity carrier gas operation
- Injection: 50 µL large volume injection into a Carbofrit liner at 5 µL/s
- Oven program: 40 °C hold for 3 min, ramp at 20 °C/min to 250 °C, no final hold
- Injector: Model 1079, temperature ramp from 65 °C to 350 °C
- Detection: Agilent 240-MS single quadrupole mass spectrometer with electron impact ionization; positive and negative chemical ionization options and MS/MS capabilities
- Sample: 0.8 ppb standard mixture of 30 pesticides and PCBs
Main Results and Discussion
The Rapid-MS approach achieved baseline separation of 30 target compounds in under 12 minutes, with PCB180 eluting last at 11.54 minutes. Large volume injection improved sensitivity, delivering detection limits at sub-ppb levels. The MS vacuum effect reduced column bleed and thermal decomposition, enhancing signal-to-noise ratios. Chromatographic efficiency remained high despite carrier gas velocities far above conventional limits.
Benefits and Practical Applications
- High-throughput screening: Run times cut by 4–5× enable more samples per day
- Improved sensitivity: Large volume injection and rapid MS ensure low detection limits for trace analysis
- Robust separations: Reduced column bleed and compound degradation extend column life
- Regulatory compliance: Rapid confirmation of pesticide residues supports food safety and environmental testing
Future Trends and Opportunities
Integration of ultrafast GC with tandem MS architectures and automated sample preparation will further boost throughput and selectivity. Emerging ionization techniques and miniaturized column formats may enable in-field or on-site pesticide monitoring. Data analytics and cloud-based reporting will streamline regulatory workflows.
Conclusion
The combination of large volume injection, Agilent Rapid-MS column technology and fast single-quadrupole MS delivers a powerful analytical platform for multiresidue pesticide analysis. Laboratories benefit from drastically reduced runtimes, improved sensitivity and robust quantitation, meeting the growing demand for high-throughput, trace-level environmental and food safety testing.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Organochlorine Pesticides by GCxGC-ECD
2010|Agilent Technologies|Applications
® Organochlorine Pesticides by GCxGC-ECD LECO Corporation; Saint Joseph, Michigan USA Key Words: GCxGC-ECD, Environmental, Quantification A relatively new way to solve separation problems is to use comprehensive two-dimensional GC (GCxGC). GCxGC is a way to increase peak capacity by…
Key words
hch, hchendrin, endrinendosulfan, endosulfanecd, ecdgamma, gammagcxgc, gcxgcheptachlor, heptachlorchlordane, chlordanealpha, alphaaldehyde, aldehydedecachlorobiphenyl, decachlorobiphenyltetrachloro, tetrachloromethoxychlor, methoxychlorepoxide, epoxidedieldrin
Automated liquid-liquid extraction workflow for direct ultra-trace analysis of pesticides and PAHs in water matrices using GC-MS/MS
2025|Thermo Fisher Scientific|Applications
Application note | 003943 Environmental Automated liquid-liquid extraction workflow for direct ultra-trace analysis of pesticides and PAHs in water matrices using GC-MS/MS Authors Goal Nicholas A. Warner1, Sinéad McDonald2, The goal of this application note is to demonstrate the capabilities…
Key words
benzo, benzohch, hchfluoranthene, fluoranthenetrans, transcis, cisnonachlor, nonachlorendrin, endrinpyrene, pyreneheptachlor, heptachlorchlordane, chlordaneanthracene, anthracenegamma, gammaendosulfan, endosulfanovex, ovexpermethrin
A Guide to Preparing and Analyzing Chlorinated Pesticides
1999|Restek|Guides
Technical Guide A Guide to Preparing and Analyzing Chlorinated Pesticides Inside: Extraction Methods for Liquid, Solid, and Biota Samples Sample Cleanup Methods Analysis of Chlorinated Pesticides Summary 2 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○…
Key words
pack, packbhc, bhcendrin, endrinendosulfan, endosulfanchlordane, chlordanehch, hchclpesticides, clpesticidesheptachlor, heptachlordata, dataeach, eachrtx, rtxester, esterextraction, extractionpress, presstight
Screening for Hazardous Chemicals in Homeland Security and Environmental Samples Using a GC/MS/ECD/FPD with a 731 Compound DRS Database
2006|Agilent Technologies|Applications
Screening for Hazardous Chemicals in Homeland Security and Environmental Samples Using a GC/MS/ECD/FPD with a 731 Compound DRS Database Application Note Homeland Security, Environmental Authors Bruce Quimby Mike Szelewski Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808 UASA Abstract…
Key words
methyl, methylfpd, fpdamdis, amdissim, simester, esterphthalate, phthalatemsd, msdsplit, splitscan, scanecd, ecddfpd, dfpdsplitter, splitterhazardous, hazardousparathion, parathionepc