GC Analysis of Basic and Neutral Compounds on Agilent J&W FactorFour VF-35ms with EZ-Guard
Applications | 2010 | Agilent TechnologiesInstrumentation
This summary reviews a safeguard strategy against complex-matrix contamination in gas chromatography, focusing on basic and neutral compound analysis. Guard columns prevent early system fouling and passivation, ensuring reliable results and reduced downtime.
This study evaluates the performance of the Agilent J&W FactorFour VF-35ms column equipped with an integrated 10 m EZ-Guard section for comprehensive separation of 14 key basic-neutral analytes commonly encountered in environmental testing methods such as EPA 8270.
Sample injections of 1 µL (20 µg/mL in methylene chloride) were made in split mode (1:75). The GC temperature was ramped from 50 °C to 300 °C at 10 °C/min. Hydrogen carrier gas was maintained at 80 kPa. Injector and FID detector temperatures were set at 275 °C and 325 °C, respectively.
The column demonstrated sharp, well-resolved peaks for all 14 analytes, confirming the inertness of the EZ-Guard segment. Integration of the uncoated guard as an transfer line enhanced detector stabilization and boosted overall productivity by preventing active site interactions and leaks.
Use of the EZ-Guard extends analytical column lifetime under challenging matrix conditions, reduces maintenance frequency, and preserves chromatographic performance for environmental and multicomponent assays.
Emerging strategies may integrate guard functionality within column hardware, explore alternative stationary phases for broader selectivity, and advance miniature or portable GC systems for rapid field screening.
The Agilent J&W FactorFour VF-35ms column with EZ-Guard effectively balances robustness and performance for basic and neutral compound analysis, enhancing system uptime and data quality in environmental testing.
No external literature citations were provided in the original document.
GC, GC columns, Consumables
IndustriesManufacturerAgilent Technologies
Summary
Importance of the Topic
This summary reviews a safeguard strategy against complex-matrix contamination in gas chromatography, focusing on basic and neutral compound analysis. Guard columns prevent early system fouling and passivation, ensuring reliable results and reduced downtime.
Objectives and Study Overview
This study evaluates the performance of the Agilent J&W FactorFour VF-35ms column equipped with an integrated 10 m EZ-Guard section for comprehensive separation of 14 key basic-neutral analytes commonly encountered in environmental testing methods such as EPA 8270.
Methodology and Sample Preparation
Sample injections of 1 µL (20 µg/mL in methylene chloride) were made in split mode (1:75). The GC temperature was ramped from 50 °C to 300 °C at 10 °C/min. Hydrogen carrier gas was maintained at 80 kPa. Injector and FID detector temperatures were set at 275 °C and 325 °C, respectively.
Used Instrumentation
- Agilent J&W FactorFour VF-35ms GC column (30 m × 0.25 mm ID, 0.25 µm film) with 10 m EZ-Guard (CP9027)
- Gas chromatograph with flame ionization detector (FID)
- Hydrogen (H2) carrier gas at 80 kPa
- Injector temperature at 275 °C, detector temperature at 325 °C
Main Results and Discussion
The column demonstrated sharp, well-resolved peaks for all 14 analytes, confirming the inertness of the EZ-Guard segment. Integration of the uncoated guard as an transfer line enhanced detector stabilization and boosted overall productivity by preventing active site interactions and leaks.
Benefits and Practical Applications
Use of the EZ-Guard extends analytical column lifetime under challenging matrix conditions, reduces maintenance frequency, and preserves chromatographic performance for environmental and multicomponent assays.
Future Trends and Potential Applications
Emerging strategies may integrate guard functionality within column hardware, explore alternative stationary phases for broader selectivity, and advance miniature or portable GC systems for rapid field screening.
Conclusion
The Agilent J&W FactorFour VF-35ms column with EZ-Guard effectively balances robustness and performance for basic and neutral compound analysis, enhancing system uptime and data quality in environmental testing.
References
No external literature citations were provided in the original document.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
GC Analysis of Base/Neutral Compounds using an Agilent J&W FactorFour VF-17ms Column with EZ-Guard
2010|Agilent Technologies|Applications
GC Analysis of Base/Neutral Compounds using an Agilent J&W FactorFour VF-17ms Column with EZ-Guard Application Note Author Introduction Peter Heijnsdijk Agilent Technologies, Inc. Guard columns are well-known for extending column lifetime when using samples with a complex matrix. Unfortunately, a…
Key words
guard, guardlifetime, lifetimeezguard, ezguardheijnsdijk, heijnsdijkneutral, neutralfactorfour, factorfourpreserving, preservingmulticomponent, multicomponentpeter, peterextending, extendingcolumns, columnsuptime, uptimecolumn, columntight, tightextend
GC Analysis of PAHs using an Agilent J&W FactorFour VF-17ms Column with EZ-Guard
2010|Agilent Technologies|Applications
GC Analysis of PAHs using an Agilent J&W FactorFour VF-17ms Column with EZ-Guard Application Note Author Introduction Peter Heijnsdijk Agilent Technologies, Inc. Polycyclic aromatic hydrocarbons (PAHs) can be formed during incomplete combustion or pyrolysis of organic matter, industrial processes and…
Key words
guard, guardcolumn, columnpahs, pahsheijnsdijk, heijnsdijkadhere, adherepeter, peterarising, arisingrestore, restorenonvolatile, nonvolatileincomplete, incompleteoriginating, originatingisothermal, isothermalcomponents, componentsmeter, meterextends
GC Analysis of Halogenated Pesticides using the Agilent J&W FactorFour VF-1701ms with EZ-Guard
2010|Agilent Technologies|Applications
GC Analysis of Halogenated Pesticides using the Agilent J&W FactorFour VF-1701ms with EZ-Guard Application Note Author Introduction Peter Heijnsdijk Agilent Technologies, Inc. Pesticides analyses can be challenging due to complex sample matrices. Guard columns are often used to protect the…
Key words
guard, guardheijnsdijk, heijnsdijkpesticides, pesticidesboosting, boostingpeter, peterlifetimes, lifetimesleakage, leakagehalogenated, halogenatedmaximizing, maximizingcolumn, columndieldrin, dieldrinendrin, endrininertness, inertnessendosulfan, endosulfandemonstrating
Agilent GC AND GC/MS Columns Catalog
2016|Agilent Technologies|Others
Buy online: www.agilent.com/chem/store Contact us: www.agilent.com/chem/contactus THE ESSENTIAL CHROMATOGRAPHY & SPECTROSCOPY CATALOG For more information This information is subject to change without notice. © Agilent Technologies, Inc. 2016 Printed in Canada March 8, 2016 5991-5213EN GC AND GC/MS Your Essential…
Key words
agilent, agilentcolumns, columnslength, lengthfilm, filmrinsable, rinsableultimetal, ultimetalltm, ltmitalicized, italicizedcolumn, columnbonded, bondedchromosorb, chromosorbtest, testdescriptions, descriptionslinked, linkedmesh