GC Analysis of PAHs using an Agilent J&W FactorFour VF-17ms Column with EZ-Guard
Applications | 2010 | Agilent TechnologiesInstrumentation
Polycyclic aromatic hydrocarbons (PAHs) are environmental pollutants of concern due to their carcinogenic properties. Regulatory agencies in the US and EU mandate monitoring of PAHs in diverse matrices, including air, water, sludge, and food. Complex sample matrices often introduce nonvolatile interferences that degrade GC column performance and detection sensitivity.
This study aims to evaluate the performance of the Agilent J&W FactorFour VF-17ms gas chromatography column equipped with an EZ-Guard in the separation and detection of 16 priority PAHs. The focus is on chromatographic selectivity, peak shape, and column longevity when analyzing complex matrices.
The analysis was conducted using GC coupled with flame ionization detection (GC-FID). Key parameters included:
The VF-17ms column with EZ-Guard achieved baseline separation of all 16 target PAH compounds, including isomeric pairs such as phenanthrene/anthracene and benzo[b]fluoranthene/benzo[k]fluoranthene. Peak shapes remained sharp and symmetric, indicating minimal interaction with nonvolatile sample residues. The integrated guard effectively trapped matrix interferences, preserving analytical column performance.
The use of an EZ-Guard extends the lifetime of the analytical column by preventing nonvolatile fouling. Quick and simple replacement of EZ-Guard segments restores performance without compromising resolution. This configuration enhances laboratory throughput and reduces maintenance downtime in routine PAH monitoring for environmental and food safety laboratories.
Advances in guard column technology may further simplify handling of complex samples by offering automated guard replacement. Coupling VF-17ms columns with mass spectrometric detection could enhance sensitivity and selectivity for trace-level PAH quantitation. Ongoing developments in fast GC and multidimensional chromatography present opportunities for even faster and more comprehensive PAH profiling.
The Agilent J&W FactorFour VF-17ms column with EZ-Guard demonstrates excellent performance for the analysis of PAHs in challenging matrices. It delivers high selectivity, maintains peak integrity, and improves column longevity by trapping nonvolatile contaminants, making it a robust choice for regulatory and research laboratories.
Agilent Technologies, Inc. "GC Analysis of PAHs using an Agilent J&W FactorFour VF-17ms Column with EZ-Guard", Application Note SI-02221, October 2010.
GC, GC columns, Consumables
IndustriesEnvironmental, Food & Agriculture
ManufacturerAgilent Technologies
Summary
Significance of the Topic
Polycyclic aromatic hydrocarbons (PAHs) are environmental pollutants of concern due to their carcinogenic properties. Regulatory agencies in the US and EU mandate monitoring of PAHs in diverse matrices, including air, water, sludge, and food. Complex sample matrices often introduce nonvolatile interferences that degrade GC column performance and detection sensitivity.
Objectives and Study Overview
This study aims to evaluate the performance of the Agilent J&W FactorFour VF-17ms gas chromatography column equipped with an EZ-Guard in the separation and detection of 16 priority PAHs. The focus is on chromatographic selectivity, peak shape, and column longevity when analyzing complex matrices.
Methodology and Instrumentation
The analysis was conducted using GC coupled with flame ionization detection (GC-FID). Key parameters included:
- Column: Agilent FactorFour VF-17ms, 30 m × 0.25 mm i.d., 0.25 μm film thickness with 10 m EZ-Guard
- Carrier gas: Helium at 2.0 bar
- Injector: Splitless injection of 2 μL sample (4–27 μg/mL in acetonitrile) at 275 °C, 1 min splitless, then split 100 mL/min
- Detector: FID at 325 °C
- Temperature program: 70 °C (1.7 min), ramp to 180 °C at 20 °C/min, to 280 °C at 8 °C/min (5 min hold), to 350 °C at 4 °C/min (10 min hold)
Main Results and Discussion
The VF-17ms column with EZ-Guard achieved baseline separation of all 16 target PAH compounds, including isomeric pairs such as phenanthrene/anthracene and benzo[b]fluoranthene/benzo[k]fluoranthene. Peak shapes remained sharp and symmetric, indicating minimal interaction with nonvolatile sample residues. The integrated guard effectively trapped matrix interferences, preserving analytical column performance.
Benefits and Practical Applications
The use of an EZ-Guard extends the lifetime of the analytical column by preventing nonvolatile fouling. Quick and simple replacement of EZ-Guard segments restores performance without compromising resolution. This configuration enhances laboratory throughput and reduces maintenance downtime in routine PAH monitoring for environmental and food safety laboratories.
Future Trends and Potential Applications
Advances in guard column technology may further simplify handling of complex samples by offering automated guard replacement. Coupling VF-17ms columns with mass spectrometric detection could enhance sensitivity and selectivity for trace-level PAH quantitation. Ongoing developments in fast GC and multidimensional chromatography present opportunities for even faster and more comprehensive PAH profiling.
Conclusion
The Agilent J&W FactorFour VF-17ms column with EZ-Guard demonstrates excellent performance for the analysis of PAHs in challenging matrices. It delivers high selectivity, maintains peak integrity, and improves column longevity by trapping nonvolatile contaminants, making it a robust choice for regulatory and research laboratories.
Reference
Agilent Technologies, Inc. "GC Analysis of PAHs using an Agilent J&W FactorFour VF-17ms Column with EZ-Guard", Application Note SI-02221, October 2010.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Separation of EU and EPA Regulated PAHs on Agilent J&W FactorFour VF-17ms Column
2010|Agilent Technologies|Applications
Separation of EU and EPA Regulated PAHs on Agilent J&W FactorFour VF-17ms Column Application Note Author Introduction Laura Provoost Agilent Technologies, Inc. The difficulty in analyzing polycyclic aromatic hydrocarbons (PAHs) is the number of PAHs with the same mass. This…
Key words
benzo, benzotriphenylene, triphenylenefluoranthene, fluoranthenepah, pahpahs, pahsanthracene, anthracenepyrene, pyreneresolve, resolveepa, epadifficult, difficultbenz, benzpolycyclic, polycyclicfluorene, fluorenechrysene, chrysenemin
Fast separation of EU and EPA Regulated PAHs on Agilent J&W FactorFour VF-17ms for PAH
2010|Agilent Technologies|Applications
Fast separation of EU and EPA Regulated PAHs on Agilent J&W FactorFour VF-17ms for PAH Application Note Author Introduction Laura Provoost Agilent Technologies, Inc. When analyzing polycyclic aromatic hydrocarbons (PAHs) the bottleneck is the number of PAHs with the same…
Key words
benzo, benzofluoranthene, fluoranthenetriphenylene, triphenylenepahs, pahsanthracene, anthraceneepa, epabenz, benzfluorene, fluorenepah, pahchrysene, chrysenepyrene, pyreneregulated, regulatedprovoost, provoostmin, minlaura
Fast GC Analysis of EPA and EU Regulated PAHs
2010|Agilent Technologies|Applications
Fast GC Analysis of EPA and EU Regulated PAHs Application Note Authors Introduction Johan Kuipers Agilent Technologies, Inc. After assessing 33 PAHs, the European Commission Scientific Committee on Food (SCF) in 2002 found that due to their toxicity, 15 of…
Key words
abundance, abundancemin, minperylene, peryleneanthracene, anthracenebenzo, benzotime, timephenantrene, phenantrenepah, pahpahs, pahscoelution, coelutionghi, ghiambientale, ambientalebardini, bardinicompromizing, compromizingdaep
PAHs
2011|Agilent Technologies|Applications
PAHs Application Note Environmental Authors Introduction Agilent Technologies, Inc. GC analysis of 16 polyaromatic hydrocarbons according to EPA 610 is achieved in less than 23 minutes using Agilent VF-5ms EZ-Guard columns. Conditions Technique : GC Column : Agilent FactorFour EZ-Guard…
Key words
courtesy, courtesypahs, pahsprinted, printedinjector, injectorusa, usaauthors, authorsprogram, programenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationpressure, pressuresolvent, solventdetector