Highly Sensitive Analysis of Fragrance Allergens in Cosmetics Using Triple Quadrupole GCMS
Applications | 2023 | ShimadzuInstrumentation
Cosmetic products often contain fragrance compounds that can induce allergic responses in sensitive individuals. Regulatory bodies such as the EU have defined limits for these allergens, and expanding these lists requires analytical methods with high sensitivity and selectivity to ensure consumer safety and compliance.
This work evaluates the performance of the Shimadzu GCMS-TQ8040 NX triple quadrupole GC–MS system operated in multiple reaction monitoring (MRM) mode for the quantitative analysis of 57 regulated fragrance allergens in cosmetic matrices. The study compares method sensitivity, selectivity, and stability against conventional selected ion monitoring (SIM) approaches, using a hair oil sample as a representative complex matrix.
Standards of 57 allergens were prepared in methyl tert-butyl ether at concentrations from 0.05 to 5 mg/kg, with 1,4-dibromobenzene and 4,4′-dibromobiphenyl (5 mg/kg) as internal standards. Hair oil (0.1 g) was diluted to 10 mL and spiked at 0.1 mg/kg. GC conditions included splitless injection (1 µL), helium carrier gas at 40 cm/s linear velocity, and a multi-step oven temperature program from 40 °C to 280 °C. MS parameters were ion source 200 °C, interface 280 °C, emission current 20 µA, data acquisition in MRM mode with 0.3 s loop time.
Expanding regulatory requirements may demand multi-residue methods covering over 80 allergens. Integrating automated sample preparation, two-dimensional GC separations, or high-resolution mass spectrometry could further improve throughput and analytical confidence in routine QA/QC environments.
The GCMS-TQ8040 NX in MRM mode provides a robust, highly sensitive, and selective platform for quantifying fragrance allergens in cosmetics. Compared to SIM, MRM offers superior accuracy in complex matrices, enabling reliable quality control and compliance with evolving regulations.
GC/MSD, GC/MS/MS, GC/QQQ
IndustriesOther
ManufacturerShimadzu
Summary
Significance of the Topic
Cosmetic products often contain fragrance compounds that can induce allergic responses in sensitive individuals. Regulatory bodies such as the EU have defined limits for these allergens, and expanding these lists requires analytical methods with high sensitivity and selectivity to ensure consumer safety and compliance.
Objectives and Study Overview
This work evaluates the performance of the Shimadzu GCMS-TQ8040 NX triple quadrupole GC–MS system operated in multiple reaction monitoring (MRM) mode for the quantitative analysis of 57 regulated fragrance allergens in cosmetic matrices. The study compares method sensitivity, selectivity, and stability against conventional selected ion monitoring (SIM) approaches, using a hair oil sample as a representative complex matrix.
Used Instrumentation
- Shimadzu GCMS-TQ8040 NX triple quadrupole GC–MS
- AOC-30i autoinjector
- SH-1 capillary column (30 m × 0.25 mm I.D., 0.25 µm film)
Methodology and Instrumentation
Standards of 57 allergens were prepared in methyl tert-butyl ether at concentrations from 0.05 to 5 mg/kg, with 1,4-dibromobenzene and 4,4′-dibromobiphenyl (5 mg/kg) as internal standards. Hair oil (0.1 g) was diluted to 10 mL and spiked at 0.1 mg/kg. GC conditions included splitless injection (1 µL), helium carrier gas at 40 cm/s linear velocity, and a multi-step oven temperature program from 40 °C to 280 °C. MS parameters were ion source 200 °C, interface 280 °C, emission current 20 µA, data acquisition in MRM mode with 0.3 s loop time.
Key Results and Discussion
- Calibration curves demonstrated excellent linearity (R² > 0.995) for 56 of 57 compounds at levels down to 0.05 mg/kg; Sclareol showed R² = 0.994. Repeatability at the lowest concentration had area RSD < 10% for all analytes.
- MRM mode significantly reduced interference from co-eluting impurities compared to SIM. For example, methyl salicylate quantitation in hair oil improved from unreliable SIM results to accurate MRM analysis at 0.048 mg/kg.
- Increased sample dilution achievable by MRM minimized injector contamination, enhancing instrument uptime and data consistency.
Benefits and Practical Applications
- Enhanced sensitivity allows reliable detection at sub-mg/kg levels, supporting stricter regulatory limits.
- Improved selectivity reduces false positives and simplifies data processing in complex cosmetic matrices.
- Higher dilution factors extend maintenance intervals and promote stable laboratory operations.
Future Trends and Applications
Expanding regulatory requirements may demand multi-residue methods covering over 80 allergens. Integrating automated sample preparation, two-dimensional GC separations, or high-resolution mass spectrometry could further improve throughput and analytical confidence in routine QA/QC environments.
Conclusion
The GCMS-TQ8040 NX in MRM mode provides a robust, highly sensitive, and selective platform for quantifying fragrance allergens in cosmetics. Compared to SIM, MRM offers superior accuracy in complex matrices, enabling reliable quality control and compliance with evolving regulations.
Reference
- Regulation (EC) No 1223/2009 of the European Parliament and of the Council on cosmetic products
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Quantitative Analysis of 57 Fragrance Allergens in Cosmetics Using Twin Line MS System
2023|Shimadzu|Applications
GC-MS GCMS-QP2020 NX Application News Quantitative Analysis of 57 Fragrance Allergens in Cosmetics Using Twin Line MS System Yuto Nakasuji and Takahide Hiramatsu User Benefits The twin line MS system enables efficient analysis of fragrance allergens in cosmetics. …
Key words
デフォルト, デフォルトfragrance, fragrancetwin, twinsalicylate, salicylatearea, areaallergens, allergenshair, hairdamascone, damasconecosmetics, cosmeticsamyl, amylneral, neralnews, newsbenzyl, benzylcitronellol, citronellolline
Highly Selective GC/TQ Analysis of 57 Allergens in Cosmetic Products
2024|Agilent Technologies|Applications
Application Note Materials Testing and Research Highly Selective GC/TQ Analysis of 57 Allergens in Cosmetic Products Featuring the Agilent 8890 GC and Agilent 7000E GC/TQ systems Authors Stéphane Découflet, Hansel Koelblen, and Michael Rothaupt Agilent Technologies, Inc. Abstract With the…
Key words
alpha, alpharelative, relativebenzyl, benzylcounts, countscinnamaldehyde, cinnamaldehydesclareol, sclareolpinene, pineneacquisition, acquisitionsalicylate, salicylateresponse, responsecinnamyl, cinnamylamyl, amylsuper, supermenthol, mentholterpinene
Analysis of Fragrance Allergens - No. 476
2023|Shimadzu|Applications
ERAS-1000-0476 GC-MS SH Series SH-1 476 Analysis of Fragrance Allergens Keywords: EC 1223/2009, G/TBT/N/EU/ 924, IFRA 18-a, 19 1,5 3 4 2 8 2.5 67 10 15 11 9 12, 17 13 IS-1 5.0 7.5 16 1418-b20 21 24 27…
Key words
damascone, damasconebenzyl, benzylsalicylate, salicylateacetate, acetatesantalol, santalolalcohol, alcoholcinnamaldehyde, cinnamaldehydepinene, pineneamyl, amyltetramethylacetyloctahydronaphthalene, tetramethylacetyloctahydronaphthalenedimethylbenzylcarbinyl, dimethylbenzylcarbinylmethylpropional, methylpropionalbenzenepropanol, benzenepropanolamylcinnamyl, amylcinnamylhexadecanolactone
Analysis of Fragrance Allergens
2023|Shimadzu|Applications
ERAS-1000-0474 GC-MS SH Series SH-I-17 474 Analysis of Fragrance Allergens Keywords: EC 1223/2009, G/TBT/N/EU/ 924, IFRA 12 13 11 8 5 2 1 10 6 7 3 4 17 16 14 15 9 18-b 29 30 31 28-b 28-a 24…
Key words
damascone, damasconebenzyl, benzylsalicylate, salicylateacetate, acetatesantalol, santalolalcohol, alcoholcinnamaldehyde, cinnamaldehydepinene, pineneamyl, amyltetramethylacetyloctahydronaphthalene, tetramethylacetyloctahydronaphthalenedimethylbenzylcarbinyl, dimethylbenzylcarbinylmethylpropional, methylpropionalbenzenepropanol, benzenepropanolhexadecanolactone, hexadecanolactoneamylcinnamyl