IMPROVED Agilent J&W CP-Wax 52 CB GC columns
Others | 2016 | Agilent TechnologiesInstrumentation
Polar compounds often interact with active sites in traditional GC systems, leading to peak tailing, reduced sensitivity and unreliable quantification. Ensuring inert flow paths and column surfaces is essential for accurate, repeatable analysis of these challenging analytes.
This application note presents the development and evaluation of improved Agilent J&W CP-Wax 52 CB GC columns. The primary goal is to demonstrate enhanced inertness, minimal column bleed and consistent chromatographic performance for polar analytes, while maintaining selectivity and ease of method transfer.
The improved CP-Wax 52 CB columns were tested against standard versions and competitor wax columns using a demanding polar test probe mixture.
Longevity testing at 250 °C for 50 h showed that improved CP-Wax 52 CB columns retained inertness and produced sharp, symmetrical peaks for active polar compounds. In contrast, competitor wax columns exhibited rapid degradation in peak shape and increased bleed. Retention times on improved columns matched those of standard CP-Wax 52 CB, confirming method compatibility and ease of upgrade.
The improved CP-Wax 52 CB columns offer:
These features support sensitive environmental, flavor and fragrance, pharmaceutical and petrochemical analyses where polar compounds are critical.
Further improvements in surface deactivation techniques may drive detection limits lower. Integration with high-resolution MS and tailored stationary phases for specific polar classes will expand application scope. Automation in column testing and reporting could streamline quality assurance and method validation.
Improved Agilent J&W CP-Wax 52 CB GC columns deliver robust inertness, low bleed and consistent chromatographic performance for polar compound analysis. Their compatibility with existing methods simplifies upgrades, ensuring reliable, sensitive and repeatable results across diverse analytical applications.
GC columns, Consumables
IndustriesManufacturerAgilent Technologies
Summary
Importance of Topic
Polar compounds often interact with active sites in traditional GC systems, leading to peak tailing, reduced sensitivity and unreliable quantification. Ensuring inert flow paths and column surfaces is essential for accurate, repeatable analysis of these challenging analytes.
Objectives and Study Overview
This application note presents the development and evaluation of improved Agilent J&W CP-Wax 52 CB GC columns. The primary goal is to demonstrate enhanced inertness, minimal column bleed and consistent chromatographic performance for polar analytes, while maintaining selectivity and ease of method transfer.
Methodology and Instrumentation
The improved CP-Wax 52 CB columns were tested against standard versions and competitor wax columns using a demanding polar test probe mixture.
- GC/MS System: Agilent 7890B GC coupled to 5977A MSD
- Columns: Standard and improved CP-Wax 52 CB, 30 m×0.25 mm, 0.25 µm
- Autosampler: Agilent 7683B, 1 µL injection
- Carrier Gas: Helium, constant 0.7 mL/min
- Inlet Conditions: Split/splitless at 250 °C, split 100:1
- Oven Program: 50 °C (5 min), ramp 5 °C/min to 250 °C (5 min hold)
- Mass Spectrometry: EI scan 40–400 amu, source 230 °C, quad 150 °C, transfer line 250 °C
Main Results and Discussion
Longevity testing at 250 °C for 50 h showed that improved CP-Wax 52 CB columns retained inertness and produced sharp, symmetrical peaks for active polar compounds. In contrast, competitor wax columns exhibited rapid degradation in peak shape and increased bleed. Retention times on improved columns matched those of standard CP-Wax 52 CB, confirming method compatibility and ease of upgrade.
Benefits and Practical Applications
The improved CP-Wax 52 CB columns offer:
- Superior peak shape for weakly retained, polar analytes
- Extended inertness lifetime under repeated high-temperature cycling
- Excellent column-to-column reproducibility and retention time stability
- Unchanged selectivity, theoretical plates and retention indices for seamless method transfer
These features support sensitive environmental, flavor and fragrance, pharmaceutical and petrochemical analyses where polar compounds are critical.
Future Trends and Possibilities
Further improvements in surface deactivation techniques may drive detection limits lower. Integration with high-resolution MS and tailored stationary phases for specific polar classes will expand application scope. Automation in column testing and reporting could streamline quality assurance and method validation.
Conclusion
Improved Agilent J&W CP-Wax 52 CB GC columns deliver robust inertness, low bleed and consistent chromatographic performance for polar compound analysis. Their compatibility with existing methods simplifies upgrades, ensuring reliable, sensitive and repeatable results across diverse analytical applications.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
IMPROVED Agilent J&W CP-Wax 52 CB GC columns
2016|Agilent Technologies|Others
ENHANCE YOUR POLAR PEAK SHAPE ACCURATE. SENSITIVE. REPEATABLE. IMPROVED Agilent J&W CP-Wax 52 CB GC columns Now your analysis of polar compounds can be even more worry free Flow path inertness is vital to any GC analysis. As the GC…
Key words
polar, polarocimene, ocimenecaryophyllene, caryophyllenecolumn, columnpinene, pinenelinalool, linaloolacetate, acetatevendor, vendoraving, avinginertness, inertnessbornanol, bornanolnreliable, nreliablelimits, limitsimproved, improvedcompounds
Agilent J&W CP-Wax 52 CB Columns with Improved Inertness Consistency from Column to Column
2016|Agilent Technologies|Technical notes
Agilent J&W CP-Wax 52 CB Columns with Improved Inertness Consistency from Column to Column Technical Overview Introduction Polyethylene glycol (PEG) also known as Wax-type GC columns have been the industry standard for many years for the analysis of a wide…
Key words
inertness, inertnessvendor, vendorcolumn, columnwax, waxprobes, probesmaltol, maltolbenchmarking, benchmarkingtest, testimproved, improvedagilent, agilentcolumns, columnsconsistency, consistencystandard, standardimprovement, improvementdecanal
Agilent GC AND GC/MS Columns Catalog
2016|Agilent Technologies|Others
Buy online: www.agilent.com/chem/store Contact us: www.agilent.com/chem/contactus THE ESSENTIAL CHROMATOGRAPHY & SPECTROSCOPY CATALOG For more information This information is subject to change without notice. © Agilent Technologies, Inc. 2016 Printed in Canada March 8, 2016 5991-5213EN GC AND GC/MS Your Essential…
Key words
agilent, agilentcolumns, columnslength, lengthfilm, filmrinsable, rinsableultimetal, ultimetalltm, ltmcolumn, columnitalicized, italicizedbonded, bondedchromosorb, chromosorbtest, testdescriptions, descriptionsmesh, meshlinked
GC AND GC/MS - Your Essential Resource for Columns & Supplies
2016|Agilent Technologies|Brochures and specifications
Buy online: www.agilent.com/chem/store Contact us: www.agilent.com/chem/contactus THE ESSENTIAL CHROMATOGRAPHY & SPECTROSCOPY CATALOG For more information This information is subject to change without notice. © Agilent Technologies, Inc. 2014 Printed in Canada October XX, 2014 5991-5213EN GC AND GC/MS Your Essential…
Key words
agilent, agilentsupplies, suppliescolumns, columnslength, lengthparts, partscrosslab, crosslabcolumn, columnfilm, filmsuggested, suggesteddescription, descriptionoem, oemrinsable, rinsableultimetal, ultimetaltips, tipsnut