IMPROVED Agilent J&W CP-Wax 52 CB GC columns
Others | 2016 | Agilent TechnologiesInstrumentation
Polar compounds analysis requires highly inert GC flow paths to achieve low detection limits and reliable quantification. Active sites within columns can cause peak tailing, reduced sensitivity, and compromised reproducibility, impacting applications in environmental monitoring, food safety, and pharmaceuticals.
This application note describes the development and evaluation of the improved Agilent J&W CP-Wax 52 CB GC columns. The aim is to demonstrate enhanced inertness, extended lifetime under harsh thermal conditions, and consistent column-to-column performance while maintaining selectivity comparable to standard CP-Wax 52 CB columns for seamless method transfer.
A modified Grob test mixture in dichloromethane was used to assess column inertness over 50 hours of conditioning at 250 °C. FID chromatograms compared the improved CP-Wax 52 CB, the standard version, and various competitor wax columns. Retention time stability and peak shapes for active polar analytes were monitored to evaluate performance.
The improved CP-Wax 52 CB columns maintained inertness after extended high-temperature exposure, whereas competitor columns exhibited rapid activity loss. Peak shapes for active polar analytes remained sharp and symmetrical. Column-to-column reproducibility and retention time stability were significantly enhanced, and the new columns matched the selectivity and retention indices of the standard version, enabling straightforward upgrades.
Continued advances in column deactivation technologies will broaden the scope of trace analysis for increasingly complex polar matrices. Integration with inert flow path components, automated system validation, and novel stationary phase chemistries are expected to drive improvements in metabolomics, forensic, and regulatory testing. Scaling these solutions for high-throughput workflows will further enhance analytical robustness.
The improved Agilent J&W CP-Wax 52 CB GC columns deliver superior inertness, consistent performance, and ease of method transition, addressing key challenges in polar compound analysis. Laboratories can achieve greater sensitivity, reproducibility, and operational efficiency without compromising established workflows.
GC columns, Consumables
IndustriesManufacturerAgilent Technologies
Summary
Importance of the Topic
Polar compounds analysis requires highly inert GC flow paths to achieve low detection limits and reliable quantification. Active sites within columns can cause peak tailing, reduced sensitivity, and compromised reproducibility, impacting applications in environmental monitoring, food safety, and pharmaceuticals.
Objectives and Study Overview
This application note describes the development and evaluation of the improved Agilent J&W CP-Wax 52 CB GC columns. The aim is to demonstrate enhanced inertness, extended lifetime under harsh thermal conditions, and consistent column-to-column performance while maintaining selectivity comparable to standard CP-Wax 52 CB columns for seamless method transfer.
Methodology
A modified Grob test mixture in dichloromethane was used to assess column inertness over 50 hours of conditioning at 250 °C. FID chromatograms compared the improved CP-Wax 52 CB, the standard version, and various competitor wax columns. Retention time stability and peak shapes for active polar analytes were monitored to evaluate performance.
Used Instrumentation
- GC System: Agilent 7890B coupled with 5977A MSD
- Column: Improved Agilent J&W CP-Wax 52 CB, 30 m × 0.25 mm, 0.25 µm (p/n CP8713i)
- Autosampler: Agilent 7683B with 5 µL syringe, 1 µL injection volume
- Carrier Gas: Helium at 0.7 mL/min (constant flow)
- Inlet: Split/splitless at 250 °C, split ratio 100:1
- Oven Program: 50 °C (5 min); ramp 5 °C/min to 250 °C, hold 5 min
- MS Settings: EI mode, scan range 40–400 amu, source 230 °C, quad 150 °C, transfer line 250 °C
Main Results and Discussion
The improved CP-Wax 52 CB columns maintained inertness after extended high-temperature exposure, whereas competitor columns exhibited rapid activity loss. Peak shapes for active polar analytes remained sharp and symmetrical. Column-to-column reproducibility and retention time stability were significantly enhanced, and the new columns matched the selectivity and retention indices of the standard version, enabling straightforward upgrades.
Benefits and Practical Applications
- Reliable low-level detection of polar compounds with minimal peak tailing
- Extended column lifetime under rigorous temperature cycling
- Simplified method transfer due to identical selectivity with standard columns
- Reduced maintenance and revalidation efforts, increasing laboratory productivity
- Enhanced confidence in environmental, food, and pharmaceutical analyses
Future Trends and Applications
Continued advances in column deactivation technologies will broaden the scope of trace analysis for increasingly complex polar matrices. Integration with inert flow path components, automated system validation, and novel stationary phase chemistries are expected to drive improvements in metabolomics, forensic, and regulatory testing. Scaling these solutions for high-throughput workflows will further enhance analytical robustness.
Conclusion
The improved Agilent J&W CP-Wax 52 CB GC columns deliver superior inertness, consistent performance, and ease of method transition, addressing key challenges in polar compound analysis. Laboratories can achieve greater sensitivity, reproducibility, and operational efficiency without compromising established workflows.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
IMPROVED Agilent J&W CP-Wax 52 CB GC columns
2016|Agilent Technologies|Others
ENHANCE YOUR POLAR PEAK SHAPE ACCURATE. SENSITIVE. REPEATABLE. IMPROVED Agilent J&W CP-Wax 52 CB GC columns Now your analysis of polar compounds can be even more worry free Flow path inertness is vital to any GC analysis. As the GC…
Key words
polar, polarocimene, ocimenecaryophyllene, caryophyllenecolumn, columnpinene, pinenelinalool, linaloolacetate, acetatevendor, vendorinertness, inertnessaving, avingbornanol, bornanolnreliable, nreliablelimits, limitsimproved, improvedcompounds
Agilent J&W CP-Wax 52 CB Columns with Improved Inertness Consistency from Column to Column
2016|Agilent Technologies|Technical notes
Agilent J&W CP-Wax 52 CB Columns with Improved Inertness Consistency from Column to Column Technical Overview Introduction Polyethylene glycol (PEG) also known as Wax-type GC columns have been the industry standard for many years for the analysis of a wide…
Key words
inertness, inertnessvendor, vendorcolumn, columnwax, waxprobes, probesmaltol, maltolbenchmarking, benchmarkingtest, testimproved, improvedagilent, agilentcolumns, columnsconsistency, consistencystandard, standardimprovement, improvementdecanal
Agilent GC AND GC/MS Columns Catalog
2016|Agilent Technologies|Others
Buy online: www.agilent.com/chem/store Contact us: www.agilent.com/chem/contactus THE ESSENTIAL CHROMATOGRAPHY & SPECTROSCOPY CATALOG For more information This information is subject to change without notice. © Agilent Technologies, Inc. 2016 Printed in Canada March 8, 2016 5991-5213EN GC AND GC/MS Your Essential…
Key words
agilent, agilentcolumns, columnslength, lengthfilm, filmrinsable, rinsableultimetal, ultimetalltm, ltmcolumn, columnitalicized, italicizedbonded, bondedchromosorb, chromosorbtest, testdescriptions, descriptionsmesh, meshlinked
GC AND GC/MS - Your Essential Resource for Columns & Supplies
2016|Agilent Technologies|Brochures and specifications
Buy online: www.agilent.com/chem/store Contact us: www.agilent.com/chem/contactus THE ESSENTIAL CHROMATOGRAPHY & SPECTROSCOPY CATALOG For more information This information is subject to change without notice. © Agilent Technologies, Inc. 2014 Printed in Canada October XX, 2014 5991-5213EN GC AND GC/MS Your Essential…
Key words
agilent, agilentsupplies, suppliescolumns, columnslength, lengthparts, partscrosslab, crosslabcolumn, columnfilm, filmsuggested, suggesteddescription, descriptionoem, oemrinsable, rinsableultimetal, ultimetaltips, tipsnut