Ensuring System Suitability for cis-trans FAMEs Analysis by AOAC 2012.13 on an Agilent 8890 GC with Retention Time Locking
Applications | 2019 | Agilent TechnologiesInstrumentation
Accurate profiling of fatty acid methyl esters (FAMEs) in dairy and infant formula is essential for nutritional labeling, quality control, and compliance with regulatory standards. The distinction between cis and trans isomers, especially in C18 unsaturated fatty acids, directly affects product safety and consumer health information.
This application note implements AOAC Official Method 2012.13 on an Agilent 8890 gas chromatograph fitted with a 100 m CP-Sil 88 column. The study aims to:
Sample Preparation and Standards:
Instrument Configuration:
- A full 37-component chromatogram displayed baseline separation except minor C23:0/C20:4n6 overlap, irrelevant for milk analysis.
- C18:1 isomers: six cis-trans peaks were resolved into two groups (cis vs. trans) with acceptable resolution for trans-fat reporting.
- C18:2 isomers: all four linoleic acid methyl ester isomers exhibited baseline separation with sharp, symmetrical peaks.
- C18:3 isomers: trans-containing linolenic esters separated cleanly from alpha-linolenic methyl ester, satisfying nutritional reporting requirements.
- Column performance standard (cis-trans mix) showed retention time RSD of 0.007% and peak area RSD < 2.5% over five replicate injections.
Retention Time Locking:
Advancements in column technology and software-driven retention time calibration are expected to:
The combination of a long cyanopropyl phase column, optimized temperature programming, and retention time locking on the Agilent 8890 GC system delivers robust, precise, and high-resolution analysis of cis-trans FAME isomers. This approach satisfies AOAC 2012.13 requirements and enhances laboratory efficiency and method transferability.
GC
IndustriesFood & Agriculture
ManufacturerAgilent Technologies
Summary
Significance of the Topic
Accurate profiling of fatty acid methyl esters (FAMEs) in dairy and infant formula is essential for nutritional labeling, quality control, and compliance with regulatory standards. The distinction between cis and trans isomers, especially in C18 unsaturated fatty acids, directly affects product safety and consumer health information.
Objectives and Study Overview
This application note implements AOAC Official Method 2012.13 on an Agilent 8890 gas chromatograph fitted with a 100 m CP-Sil 88 column. The study aims to:
- Evaluate separation efficiency for a standard mixture of 37 FAMEs (C4:0–C24:0).
- Assess resolution of mono- and polyunsaturated cis-trans isomers from the C18 family.
- Demonstrate retention time locking (RTL) to maintain method performance after column maintenance.
Methodology and Instrumentation
Sample Preparation and Standards:
- 37-component FAMEs mix (100–600 µg/mL) in dichloromethane.
- Reference mixes for C18:1, C18:2, C18:3 isomers and an eight-component linolenic acid standard.
- Cis-trans column performance standard.
Instrument Configuration:
- Gas chromatograph: Agilent 8890 GC with split/splitless inlet and flame ionization detector (FID).
- Column: CP-Sil 88, 100 m × 0.25 mm, 0.20 µm film thickness.
- Carrier gas: Helium at 0.8 mL/min constant flow.
- Injection: 1 µL, split ratio 10:1, inlet temperature 250 °C.
- Oven program: 60 °C hold 5 min; ramp 15 °C/min to 165 °C (1 min); 2 °C/min to 225 °C (20 min).
- Detector conditions: FID at 250 °C, H₂ 40 mL/min, air 400 mL/min, N₂ makeup 25 mL/min.
Main Results and Discussion
- A full 37-component chromatogram displayed baseline separation except minor C23:0/C20:4n6 overlap, irrelevant for milk analysis.
- C18:1 isomers: six cis-trans peaks were resolved into two groups (cis vs. trans) with acceptable resolution for trans-fat reporting.
- C18:2 isomers: all four linoleic acid methyl ester isomers exhibited baseline separation with sharp, symmetrical peaks.
- C18:3 isomers: trans-containing linolenic esters separated cleanly from alpha-linolenic methyl ester, satisfying nutritional reporting requirements.
- Column performance standard (cis-trans mix) showed retention time RSD of 0.007% and peak area RSD < 2.5% over five replicate injections.
Retention Time Locking:
- RTL wizard generated a pressure calibration curve with three runs (±15% pressure variation) using C18:0 as the target compound.
- Post-maintenance (0.5 m column trim) relocking restored retention times within 0.009 min of original values.
- Method transfer and interlaboratory alignment are facilitated by stable locked retention times.
Benefits and Practical Applications
- Reliable separation of a wide range of FAMEs and cis-trans isomers meets AOAC 2012.13 performance criteria.
- High repeatability and precision support QA/QC in dairy, infant formula, and vegetable oil analysis.
- RTL reduces downtime after column maintenance and simplifies method transfer between laboratories.
Future Trends and Potential Uses
Advancements in column technology and software-driven retention time calibration are expected to:
- Enable faster separations with shorter columns while maintaining isomer resolution.
- Integrate real-time monitoring of method performance using machine-learning-based peak recognition.
- Expand cis-trans profiling to emerging food matrices, including plant-based alternatives and functionalized lipid ingredients.
Conclusion
The combination of a long cyanopropyl phase column, optimized temperature programming, and retention time locking on the Agilent 8890 GC system delivers robust, precise, and high-resolution analysis of cis-trans FAME isomers. This approach satisfies AOAC 2012.13 requirements and enhances laboratory efficiency and method transferability.
References
- Official Methods of Analysis AOAC International, Method 2012.13, 2012.
- David F.; Sandra P.; Vickers A. K. Column Selection for the Analysis of Fatty Acid Methyl Esters. Agilent Technologies Application Note 5989-3760EN, 2005.
- Zou Y.; Wu H. Improving the Analysis of 37 Fatty Acid Methyl Esters. Agilent Technologies Application Note 5991-8706EN, 2018.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
New GC Column Technology for Old Problems in Fatty Acid Analysis
2020|Agilent Technologies|Presentations
New GC Column Technology for Old Problems in Fatty Acid Analysis Gustavo Serrano Izaguirre, PhD Global Product Manager – GC Columns December 15, 2020 1 December 15, 2020 DE44171.6363541667 What Are We Testing? Fat profiles: • • • • Total…
Key words
new, newtechnology, technologycolumn, columnfastfame, fastfameacid, acidfatwax, fatwaxintuvo, intuvofames, famestrans, transfatty, fattyfats, fatscis, cisfame, fametriglycerides, triglyceridesfid
Fatty Acid and FAME Analysis Using State-of-the-Art Gas Chromatography
2019|Agilent Technologies|Presentations
Fatty Acid and FAME Analysis Using State-of-the-Art Gas Chromatography Gustavo Serrano, Ph.D. Global Product Manager - GC Columns Agilent Technologies Outline • Analysis of Fats and Oils and Column Selection • DB-FATWAX Ultra Inert • DB-FastFAME • DB-HeavyWAX • Conclusions…
Key words
acid, acidacids, acidsfatwax, fatwaxfatty, fattydha, dhafat, fatconfidential, confidentialfame, famefid, fidwax, waxheavywax, heavywaxcolumn, columncis, cisoil, oilfastfame
Fatty Acid Methyl Esters (FAMEs) Analysis on an Agilent 8890 GC and Its Application to Real Samples
2019|Agilent Technologies|Applications
Application Note Food Testing Fatty Acid Methyl Esters (FAMEs) Analysis on an Agilent 8890 GC and Its Application to Real Samples Author Jie Zhang Agilent Technologies, Inc. Shanghai, China Abstract This Application Note applied methods GB5009.168-20161 and 5009.257-20162 to FAMEs…
Key words
fatty, fattytrans, transacids, acidsresponse, responsefames, famesrsd, rsdretention, retentionarea, areamin, minoctadecatrienoic, octadecatrienoictime, timeanpel, anpelesters, estersmethyl, methylcis
Rapid Analysis of Fatty Acid Methyl Esters (FAMEs) Using a High Resolution 90-m Agilent J&W DB‑FastFAME Capillary GC Column
2020|Agilent Technologies|Applications
Application Note Food Testing Rapid Analysis of Fatty Acid Methyl Esters (FAMEs) Using a High Resolution 90-m Agilent J&W DB‑FastFAME Capillary GC Column Author Yun Zou Agilent Technologies (Shanghai) Co. Ltd. Abstract This application note shows an excellent separation of…
Key words
fastfame, fastfametrans, transfatty, fattycarrier, carriercis, cisfames, famescolumn, columnfame, famegas, gasfid, fidseparation, separationacid, acidinlet, inletoven, ovenhydrogen