Vinyl acetate monomer
Applications | | GL SciencesInstrumentation
Residual solvent analysis is a critical quality check in the production of polymers and coatings to ensure product safety and performance. Vinyl acetate monomer, commonly used in adhesives and paints, must be quantified at trace levels to meet regulatory standards and to avoid impacts on downstream processes and material properties.
This application note evaluates the performance of an inert capillary column for the gas chromatographic analysis of vinyl acetate monomer in a toluene matrix. The primary goal is to demonstrate peak shape, sensitivity, and reproducibility under a defined temperature program and inert flow path.
A vinyl acetate standard (10 µg/mL in toluene) was injected (1 µL) into a gas chromatograph equipped with a flame ionization detector (GC/FID). Separation was achieved on a 30 m × 0.32 mm I.D. column with a 5.00 µm film. The oven program held at 100 °C for 8 min, ramped at 20 °C/min to 250 °C, and held for 5 min. Helium at 40 kPa served as the carrier gas. Injection was in split mode with a 9.6 mL/min flow, injector temperature at 150 °C, and FID temperature at 260 °C.
The chromatogram exhibited a single, well-resolved peak for vinyl acetate with no observable tailing or secondary peaks. Baseline stability and signal-to-noise ratio met analytical requirements for trace-level detection. The inert surface treatment of the column minimized adsorption, yielding highly reproducible retention times and peak areas over multiple injections.
Future directions include coupling inert column chemistries with mass spectrometric detectors for enhanced specificity, expanding the method to other residual solvents and monomers, and optimizing fast temperature ramps to further reduce analysis time. Advances in carrier gas management and column materials will continue to improve sensitivity and throughput.
The inert capillary column demonstrated excellent performance for the analysis of vinyl acetate monomer, delivering sharp peaks, consistent quantitation, and reliable throughput. This method is well suited for routine residual solvent testing in industrial laboratories.
No external references provided in the source document.
GC, GC columns, Consumables
IndustriesEnergy & Chemicals
ManufacturerGL Sciences
Summary
Importance of the Topic
Residual solvent analysis is a critical quality check in the production of polymers and coatings to ensure product safety and performance. Vinyl acetate monomer, commonly used in adhesives and paints, must be quantified at trace levels to meet regulatory standards and to avoid impacts on downstream processes and material properties.
Objectives and Study Overview
This application note evaluates the performance of an inert capillary column for the gas chromatographic analysis of vinyl acetate monomer in a toluene matrix. The primary goal is to demonstrate peak shape, sensitivity, and reproducibility under a defined temperature program and inert flow path.
Methodology and Instrumentation
A vinyl acetate standard (10 µg/mL in toluene) was injected (1 µL) into a gas chromatograph equipped with a flame ionization detector (GC/FID). Separation was achieved on a 30 m × 0.32 mm I.D. column with a 5.00 µm film. The oven program held at 100 °C for 8 min, ramped at 20 °C/min to 250 °C, and held for 5 min. Helium at 40 kPa served as the carrier gas. Injection was in split mode with a 9.6 mL/min flow, injector temperature at 150 °C, and FID temperature at 260 °C.
Instrumentation
- Gas chromatograph with FID detector
- InertCap 1 column, 30 m × 0.32 mm I.D., 5.00 µm film
- Helium carrier gas at 40 kPa
Main Results and Discussion
The chromatogram exhibited a single, well-resolved peak for vinyl acetate with no observable tailing or secondary peaks. Baseline stability and signal-to-noise ratio met analytical requirements for trace-level detection. The inert surface treatment of the column minimized adsorption, yielding highly reproducible retention times and peak areas over multiple injections.
Practical Benefits and Applications
- Enhanced inertness reduces sample interaction and maintains peak integrity for reactive analytes
- Robust temperature program ensures rapid analysis and short cycle times
- Reliable quantitation suitable for QA/QC in polymer, coating, and adhesive manufacturing
Future Trends and Applications
Future directions include coupling inert column chemistries with mass spectrometric detectors for enhanced specificity, expanding the method to other residual solvents and monomers, and optimizing fast temperature ramps to further reduce analysis time. Advances in carrier gas management and column materials will continue to improve sensitivity and throughput.
Conclusion
The inert capillary column demonstrated excellent performance for the analysis of vinyl acetate monomer, delivering sharp peaks, consistent quantitation, and reliable throughput. This method is well suited for routine residual solvent testing in industrial laboratories.
References
No external references provided in the source document.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Purity test of Colchicine
|GL Sciences|Applications
InertSearch for GC TM InertCap® Applications Purity test of Colchicine Data No. GA145-0644 I.S. 2 6.0 7.0 Time (min) 8.0 9.0 200000 1 I.S. 100000 2 0 0 2 4 6 8 10 Time ((min)) 12 14 16 Conditions :…
Key words
pharmacopoeia, pharmacopoeiapharmacopeia, pharmacopeiajapanese, japaneseコルヒチン, コルヒチンプロピルアルコール, プロピルアルコールトリクロロメタン, トリクロロメタンエチルアセテート, エチルアセテートクロロホルム, クロロホルムエチル, エチルjapan, japancolchicine, colchicineethyl, ethylガスクロマトグラフィー, ガスクロマトグラフィーイナートサーチ, イナートサーチイナートキャップ
Ethylene oxide in Ethanol
|GL Sciences|Applications
InertSearch for GC TM InertCap® Applications Ethylene oxide in Ethanol Data No. GA147-0591 12000 3 10000 2 8000 uVolt 6000 4000 1 2000 0 0 10 20 30 Time (min) Conditions Analyte : GC/FID Headspace Gas(HT3) 1. Ethylene oxide 1…
Key words
ethylene, ethyleneoxide, oxidepropylene, propylenecheminnovayion, cheminnovayiongasthen, gasthenエチレンオキシド, エチレンオキシドオキシラン, オキシランプロピレン, プロピレンプロピレンオキサイド, プロピレンオキサイドプロピレンオキシド, プロピレンオキシドメチルオキシラン, メチルオキシランエチレン, エチレンヘッドスペース, ヘッドスペースヘッドスペースガス, ヘッドスペースガスoxirane
Organic solvents
|GL Sciences|Applications
InertSearch for GC TM InertCap® Applications Organic solvents Data No. GA140-851 14+15 200000 11 7 13 6 100000 2 16 8 10 4 12 9 1 3 17 5 0 0 2 Conditions Methyl alcohol 4 6 Time (min) 8…
Key words
xylene, xylenebutyl, butylキシレン, キシレンalcohol, alcoholvol, volcyclohexane, cyclohexaneエチレングリ, エチレングリチルアセテ, チルアセテブチルベンゼン, ブチルベンゼンルモノブチルエ, ルモノブチルエモノブチルグリコールエーテル, モノブチルグリコールエーテルsolvents, solventsイソブタノール, イソブタノールメチルイソブチルケトン, メチルイソブチルケトンester
Organic solvents
|GL Sciences|Applications
InertSearch for GC TM InertCap® Applications Organic solvents Data No. GA023-0850 14000 1 12000 10000 9 13 12 14 15 17 8000 6000 3 6 4 4000 11 2 78 10 166 5 19 18 2000 0 10 Acetic acid…
Key words
キシレン, キシレンacetate, acetatebutyl, butylalcohol, alcoholアニソール, アニソールテトラヒドロフラン, テトラヒドロフランヘプタン, ヘプタンethyl, ethylエチルベンゼン, エチルベンゼンアセトン, アセトンエタノール, エタノールanisole, anisoleacetic, aceticxylene, xyleneheptane