Improving Thermal Extraction Method Reproducibility through Instrument Temperature Calibration in the Sample Position
Applications | 2016 | GERSTELInstrumentation
The accurate determination of volatile organic compounds (VOC) and condensable organic compounds (FOG) emissions from automotive interior materials is essential for meeting regulatory requirements and ensuring passenger health and comfort. Variations in extraction temperature and inert gas flow can significantly affect emission results, undermining analytical robustness and inter-laboratory reproducibility.
This study evaluates how small deviations in thermal extraction parameters influence emission data obtained by the VDA 278 method. A range of common automotive materials—including polypropylene granulate, polyurethane foam, leather, duroplastic plastics and paint stripes—were analyzed under controlled variations of desorption temperature and flow to quantify their effects on VOC and FOG emission rates.
All samples were tested according to VDA 278 specifications for thermal desorption coupled to GC/MS. Key steps included:
Extraction temperature was shown to be the most critical parameter. A deviation of ± 2 °C at 90 °C or 120 °C led to up to ± 20 % shifts in emission rates for polypropylene and polyurethane. Leather and duroplastic plastics exhibited wider variability attributed to sample heterogeneity. Variations in desorption flow (60–100 mL/min) affected paint stripe emissions more dramatically than pelletized samples; a ± 5–10 mL/min change still maintained VOC and FOG within ± 20 % for most materials, whereas thin paint layers displayed flow-dependent peak patterns and larger deviations.
By establishing acceptable tolerances—± 2 °C for temperature and ± 5–10 mL/min for flow—laboratories can ensure consistent VDA 278 analyses. In-situ temperature measurement in the desorption tube confirms true sample conditions. The MAESTRO VDA 278 calibration tool automates fine adjustments, reducing manual effort and preventing systematic biases across instruments and sites.
Integration of real-time temperature and flow monitoring with automated feedback control may further tighten method precision. Expanding calibration routines to additional temperatures and adapting them for emerging materials will broaden applicability. Advanced data analytics could predict emission behavior based on material characteristics, guiding both material design and sampling strategies.
This work demonstrates that tight control of desorption temperature and gas flow is vital for reproducible thermal extraction results per VDA 278. The combination of precise in-tube temperature measurement, MAESTRO calibration routines and robust flow control ensures inter-instrument and inter-laboratory reliability, meeting ± 20 % emission rate criteria.
GC/MSD, Thermal desorption, GC/SQ
IndustriesEnergy & Chemicals
ManufacturerAgilent Technologies, GERSTEL
Summary
Importance of the topic
The accurate determination of volatile organic compounds (VOC) and condensable organic compounds (FOG) emissions from automotive interior materials is essential for meeting regulatory requirements and ensuring passenger health and comfort. Variations in extraction temperature and inert gas flow can significantly affect emission results, undermining analytical robustness and inter-laboratory reproducibility.
Study objectives and overview
This study evaluates how small deviations in thermal extraction parameters influence emission data obtained by the VDA 278 method. A range of common automotive materials—including polypropylene granulate, polyurethane foam, leather, duroplastic plastics and paint stripes—were analyzed under controlled variations of desorption temperature and flow to quantify their effects on VOC and FOG emission rates.
Applied methodology and instrumentation
All samples were tested according to VDA 278 specifications for thermal desorption coupled to GC/MS. Key steps included:
- VOC analysis: 30 minutes at 90°C, 82 mL/min flow, cryofocusing at -150°C
- FOG analysis: 60 minutes at 120°C, 82 mL/min flow, cryofocusing at -150°C
- Instrumentation: GERSTEL TDS 3 with TDS A2 autosampler, Agilent 7890 GC, 5973 MSD, HP Ultra 2 column (50 m × 0.32 mm ID, 0.52 μm film)
- Temperature validation: Type T thermocouple with custom adapter inserted in desorption tube, readings recorded with precision handheld meter
- Software: MAESTRO enhanced with VDA 278 calibration feature
Main results and discussion
Extraction temperature was shown to be the most critical parameter. A deviation of ± 2 °C at 90 °C or 120 °C led to up to ± 20 % shifts in emission rates for polypropylene and polyurethane. Leather and duroplastic plastics exhibited wider variability attributed to sample heterogeneity. Variations in desorption flow (60–100 mL/min) affected paint stripe emissions more dramatically than pelletized samples; a ± 5–10 mL/min change still maintained VOC and FOG within ± 20 % for most materials, whereas thin paint layers displayed flow-dependent peak patterns and larger deviations.
Benefits and practical applications
By establishing acceptable tolerances—± 2 °C for temperature and ± 5–10 mL/min for flow—laboratories can ensure consistent VDA 278 analyses. In-situ temperature measurement in the desorption tube confirms true sample conditions. The MAESTRO VDA 278 calibration tool automates fine adjustments, reducing manual effort and preventing systematic biases across instruments and sites.
Future trends and potential applications
Integration of real-time temperature and flow monitoring with automated feedback control may further tighten method precision. Expanding calibration routines to additional temperatures and adapting them for emerging materials will broaden applicability. Advanced data analytics could predict emission behavior based on material characteristics, guiding both material design and sampling strategies.
Conclusion
This work demonstrates that tight control of desorption temperature and gas flow is vital for reproducible thermal extraction results per VDA 278. The combination of precise in-tube temperature measurement, MAESTRO calibration routines and robust flow control ensures inter-instrument and inter-laboratory reliability, meeting ± 20 % emission rate criteria.
Reference
- Thermal Desorption Analysis of Organic Emissions for the Characterization of Non-Metallic Materials for Automobiles. VDA 278. Verband der Automobilindustrie, 2011.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Measurement of Chemical Emissions from Spray Polyurethane Foam (SPF) Insulation Using an Automated Micro-Scale Chamber Coupled Directly with the Analysis System
2016|Agilent Technologies|Applications
GERSTEL Application Note No. 188, 2016 Measurement of Chemical Emissions from Spray Polyurethane Foam (SPF) Insulation Using an Automated Micro-Scale Chamber Coupled Directly with the Analysis System Yunyun Nie1, Eike Kleine-Benne1, Kurt Thaxton2 1. GERSTEL GmbH & Co. KG, Eberhard-Gerstel-Platz…
Key words
spf, spfemission, emissionfoam, foamchamber, chamberbdmaee, bdmaeeemissions, emissionstmibpa, tmibpatcpp, tcppinsulation, insulationcell, cellmicro, micropolyurethane, polyurethanedesorption, desorptionsink, sinkscale
"What’s inside your car?” - Car interior emissions testing using Thermal desorption GC/MS analysis
|Thermo Fisher Scientific|Presentations
"What’s inside your car?” Car interior emissions testing using Thermal desorption GC/MS analysis Inge de Dobbeleer EU GCMS specialist ThermoFisher Lara Kelly Sales Manager Markes Thermal desorption The Thermal Desorption Process On-line Electrically-cooled focusing trap Direct desorption of materials Canisters…
Key words
genesis, genesisvoc, voccar, caremissions, emissionsdesorption, desorptionfog, fogleather, leatherthermal, thermalemission, emissiontrim, trimchamber, chambercompounds, compoundssubstances, substancestube, tubemin
Automotive interior VOC and FOG emissions
2020|Thermo Fisher Scientific|Applications
TECHNICAL NOTE 73565 Automotive interior VOC and FOG emissions A technical guide for analysis of interior materials by direct desorption TD–GC–MS in accordance with VDA 278 Authors: Cristian Cojocariu1 and Dwain Cardona2, Markes International Ltd3 Thermo Fisher Scientific, Runcorn, UK…
Key words
voc, vocfog, fogcalibration, calibrationhexadecane, hexadecaneshould, shouldtoluene, toluenetenax, tenaxtube, tubesample, samplebaseline, baselinetubes, tubesrun, runtrap, trapmarkes, markescontrol
VDA 278 Analysis Using Thermal Desorption
|CDS Analytical|Applications
#97a VDA 278 Analysis Using Thermal Desorption Application Note Automotive Author: C. Zawodny The German Automobile Association method VDA 278 is a thermal desorption analysis for organic emissions from interior components of a vehicle. This analysis deals with a VOC…
Key words
voc, vocfog, fogtra, tration, tionabundance, abundancetub, tubcarpetfibers, carpetfibersfogstd, fogstdiniti, initivalv, valvfibers, fiberstda, tdafigure, figurecarpet, carpetlong