Overcoming Analytical Challenges in PBDEs Quantification in Environmental Samples using GC-Orbitrap MS
Presentations | | Thermo Fisher ScientificInstrumentation
Polybrominated diphenyl ethers (PBDEs) are widely used flame retardants that persist, bioaccumulate and pose health risks including endocrine disruption and cancer.
Their quantification in complex environmental matrices is challenged by low concentrations, isomeric interferences and thermal degradation, highlighting the need for robust high-resolution methods.
This study evaluates the performance of the Thermo Scientific Exactive GC Orbitrap GC-MS system combined with a TG-PBDE capillary column for routine targeted quantification of 27 PBDE congeners in environmental samples.
Key questions addressed:
Soxhlet extraction of 2 g sample with toluene (24 h) followed by alumina cleanup using n-hexane/DCM and concentration under nitrogen.
Internal standards (^13C-labeled PBDEs) and recovery surrogates were spiked.
Excellent chromatographic separation of all PBDE congeners in under 11 minutes, including baseline resolution of BDE-49 and BDE-71.
Calibration curves showed linearity (R²>0.998) and residual RSD <13% across five levels.
Instrument detection limits ranged from 6 fg to 250 fg on column depending on congener.
Mass accuracy was consistently within ±2 ppm with accurate isotopic patterns.
Repeatability tests (n=12) confirmed robust performance for routine injections.
Environmental sample analyses revealed predominant congeners:
The combination of the Exactive GC Orbitrap GC-MS system with a dedicated TG-PBDE column offers a robust, sensitive and high-throughput solution for routine targeted quantification of PBDEs in environmental samples, meeting the demands of regulatory and research laboratories.
No explicit references listed in the source document.
GC/MSD, GC/MS/MS, GC/HRMS, GC/Orbitrap
IndustriesEnvironmental
ManufacturerThermo Fisher Scientific
Summary
Significance of the Topic
Polybrominated diphenyl ethers (PBDEs) are widely used flame retardants that persist, bioaccumulate and pose health risks including endocrine disruption and cancer.
Their quantification in complex environmental matrices is challenged by low concentrations, isomeric interferences and thermal degradation, highlighting the need for robust high-resolution methods.
Objectives and Study Overview
This study evaluates the performance of the Thermo Scientific Exactive GC Orbitrap GC-MS system combined with a TG-PBDE capillary column for routine targeted quantification of 27 PBDE congeners in environmental samples.
Key questions addressed:
- Minimization of thermal degradation (e.g. BDE-209)
- Resolution of critical isomer pairs (BDE-49 vs. BDE-71)
- Sensitivity, selectivity and linearity metrics
- Instrument detection limits and mass accuracy
- Repeatability and suitability for routine analysis
Methodology
Soxhlet extraction of 2 g sample with toluene (24 h) followed by alumina cleanup using n-hexane/DCM and concentration under nitrogen.
Internal standards (^13C-labeled PBDEs) and recovery surrogates were spiked.
Instrumentation
- Thermo Scientific TRACE 1310 GC with rapid heating modular injector
- Thermo Scientific TG-PBDE 15 m × 0.25 mm × 0.10 µm capillary column
- Thermo Scientific Exactive GC Orbitrap GC-MS system with ExtractaBrite ion source
- Electron ionization (35 eV), resolution 60 000 FWHM at m/z 200, full scan and targeted SIM (68–1000 m/z)
- Chromeleon 7.2 CDS software for data acquisition and isotopic dilution quantification
Main Results and Discussion
Excellent chromatographic separation of all PBDE congeners in under 11 minutes, including baseline resolution of BDE-49 and BDE-71.
Calibration curves showed linearity (R²>0.998) and residual RSD <13% across five levels.
Instrument detection limits ranged from 6 fg to 250 fg on column depending on congener.
Mass accuracy was consistently within ±2 ppm with accurate isotopic patterns.
Repeatability tests (n=12) confirmed robust performance for routine injections.
Environmental sample analyses revealed predominant congeners:
- Sludge: BDE-209, BDE-206, BDE-207, BDE-99
- Filter dust: BDE-209, BDE-47, BDE-99
- Air: BDE-99, BDE-47, BDE-100
- Sediment: BDE-15, BDE-47, BDE-99
Benefits and Practical Applications
- Rapid high-resolution and accurate mass analysis reduces interferences and improves specificity
- Fast chromatographic runs (<11 min) enhance throughput for routine monitoring
- Low detection limits and reliable quantification support compliance with environmental regulations
- Integration with CDS software streamlines isotopic dilution workflows for accurate results
Future Trends and Potential Applications
- Extension to other persistent organic pollutants and emerging contaminants
- Automation of sample preparation and coupling with online extraction techniques
- Miniaturized and portable high-resolution MS platforms for field analysis
- Advanced data analytics and machine learning for predictive environmental monitoring
Conclusion
The combination of the Exactive GC Orbitrap GC-MS system with a dedicated TG-PBDE column offers a robust, sensitive and high-throughput solution for routine targeted quantification of PBDEs in environmental samples, meeting the demands of regulatory and research laboratories.
References
No explicit references listed in the source document.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Overcoming analytical challenges for polybrominated diphenyl ethers (PBDEs) analysis in environmental samples using gas chromatography – Orbitrap mass spectrometry
2018|Thermo Fisher Scientific|Applications
APPLICATION NOTE 10644 Overcoming analytical challenges for polybrominated diphenyl ethers (PBDEs) analysis in environmental samples using gas chromatography – Orbitrap mass spectrometry Authors Goal Jane Cooper,1 Jordi Parera,2 Esteban Abad,2 Richard Law,¹ and Cristian Cojocariu1 1 Thermo Fisher Scientific, Runcorn,…
Key words
bde, bdepbdes, pbdespbde, pbdecounts, countsexactive, exactivebdes, bdesisotopic, isotopicquantification, quantificationwere, weredibdes, dibdesmonobde, monobdetribdes, tribdesnumber, numberheptabdes, heptabdesenvironmental
Fast, ultra-sensitive analysis of PBDEs in food using advanced GC-MS/MS technology
2018|Thermo Fisher Scientific|Applications
APPLICATION NOTE 10674 Fast, ultra-sensitive analysis of PBDEs in food using advanced GC-MS/MS technology Authors Aaron Lamb1, Cristian Cojocariu1, and Sean Panton2 Thermo Fisher Scientific, Runcorn, UK 1 2 Fera Science, Ltd., York, UK Keywords Food safety, polybrominated diphenyl ethers…
Key words
pbde, pbdedeviation, deviationhrms, hrmsaei, aeipbdes, pbdescounts, countsratio, ratioion, ionpenta, pentatetra, tetrabde, bdehexa, hexapass, passcongeners, congenershepta
Determination of Polybrominated Diphenylethers (PBDE) in Sediment and Sewage Sludge
2004|Agilent Technologies|Applications
Determination of Polybrominated Diphenylethers (PBDE) in Sediment and Sewage Sludge Application Environmental Authors Introduction Sonia Wanner Centre d'Analyse et de Recherche (CAR) 76 route du Rhin F-67441 Illkirch France With increasing frequency, environmental laboratories are asked to analyze PBDEs (flame…
Key words
sludge, sludgesewage, sewagepbde, pbdepolybrominated, polybrominateddecabde, decabdeabundance, abundancepdbe, pdbefrance, franceoctabde, octabdebrominated, brominatedpentabde, pentabdecongeners, congenerssediment, sedimentdiphenyl, diphenylbdes
Meeting the requirements of US and European water standards
2020|Thermo Fisher Scientific|Guides
Analytical solutions: Meeting the requirements of US and European water standards Foreward The quality of the water in our rivers, lakes and seas is of significant concern for both the organisms that inhabit these environments and those that rely on…
Key words
eic, eicwater, waterdirective, directiveemerging, emergingframework, frameworkcontaminants, contaminantsdrinking, drinkingpass, passthermo, thermoscientific, scientificcompound, compoundcompounds, compoundsorbitrap, orbitrapenvironmental, environmentalrsd