Analysis of Trace Oxygenates in Petroleum-Contaminated Wastewater, Using Purge-and-Trap GC-MS (U.S. EPA Methods 5030B & 8260)
Applications | 2012 | RestekInstrumentation
Gasoline oxygenates such as MTBE and TBA are widely used fuel additives that enhance octane rating and reduce emissions. However, their prevalence in underground storage tanks poses a significant risk of contaminating ground and surface waters. Accurate monitoring of these compounds at trace levels in complex petroleum matrices is essential for environmental protection and regulatory compliance.
This study aimed to evaluate the performance of a purge-and-trap GC-MS approach, based on modified EPA Methods 5030B and 8260B, for reliable identification and quantification of trace oxygenates in petroleum-contaminated wastewater. A new capillary stationary phase (Rtx-VMS) was tested for its ability to resolve coeluting analytes such as MTBE and TBA.
The analytical protocol combined:
Purge conditions were optimized at 40 °C for 11 min with a 38 mL/min flow to enhance oxygenate recovery. The GC oven ramp (35 °C hold to 90 °C at 4 °C/min, then to 220 °C at 45 °C/min) was fine-tuned to avoid critical coelutions. Calibration followed five-point curves (5–80 ppb) with internal (methyl-d3-tert-butyl ether) and surrogate standards as per EPA 8260.
The Rtx-VMS column achieved baseline separation of target oxygenates and gasoline interferents. Relative standard deviations for response factors were generally below 12%, meeting EPA acceptance criteria. Most oxygenates exhibited recoveries between 90% and 105% in a 1 ppm gasoline matrix spiked at 5 ppb. Exceptions included acetone (62% recovery) and a slightly elevated RSD for tert-butyl alcohol. Resolution of MTBE and TBA was confirmed, preventing false positives due to shared diagnostic ions.
By integrating an optimized purge-and-trap stage with GC-MS, laboratories can:
Advancements in stationary phase chemistry and mass spectrometric techniques will further improve selectivity and sensitivity for emerging oxygenate compounds. Automation of sample introduction and data processing could enhance throughput. Additionally, coupling with high-resolution MS may address challenges posed by increasingly complex environmental matrices.
Modifying EPA purge-and-trap GC-MS methods with an Rtx-VMS column offers a robust solution for trace oxygenate analysis in petroleum-contaminated water. The approach delivers reliable separation, quantification, and compliance with regulatory performance criteria.
[1] English C., Cox C., Dorman F., Patwardhan D., “Analysis of Gasoline Oxygenates Using a New Capillary Column Stationary Phase,” Pittsburgh Conference 2001, Session 199.
[2] Happel A.M., Beckenbach E.H., Halden R.U., “MTBE Impacts to California Groundwater Resources,” LLNL Report UCRL-AR-130897, 1988.
[3] U.S. EPA, Method 8260, “Volatile Organic Compounds by GC/MS,” Rev. 0, July 1992.
[4] U.S. EPA, Method 8260B, “Volatile Organic Compounds by GC/MS,” Rev. 2, Dec. 1996.
[5] English C.M., Dorman F.L., Stidsen G.B., “Analysis of Gasoline Oxygenates by EPA Method 8260B,” Pittsburgh Conference 2003, Session 590-6P.
GC/MSD, Purge and Trap, GC/SQ
IndustriesEnvironmental
ManufacturerAgilent Technologies, OI Analytical
Summary
Importance of the Topic
Gasoline oxygenates such as MTBE and TBA are widely used fuel additives that enhance octane rating and reduce emissions. However, their prevalence in underground storage tanks poses a significant risk of contaminating ground and surface waters. Accurate monitoring of these compounds at trace levels in complex petroleum matrices is essential for environmental protection and regulatory compliance.
Objectives and Study Overview
This study aimed to evaluate the performance of a purge-and-trap GC-MS approach, based on modified EPA Methods 5030B and 8260B, for reliable identification and quantification of trace oxygenates in petroleum-contaminated wastewater. A new capillary stationary phase (Rtx-VMS) was tested for its ability to resolve coeluting analytes such as MTBE and TBA.
Methodology and Instrumentation
The analytical protocol combined:
- Purge-and-trap sample preparation (O.I. Analytical 4560 concentrator with #10 Tenax®/silica gel/carbon trap)
- GC separation on a 30 m×0.25 mm ID×1.4 µm Rtx-VMS column
- Detection by full-scan MS (Agilent 5971A coupled to 5890 Series II GC)
Purge conditions were optimized at 40 °C for 11 min with a 38 mL/min flow to enhance oxygenate recovery. The GC oven ramp (35 °C hold to 90 °C at 4 °C/min, then to 220 °C at 45 °C/min) was fine-tuned to avoid critical coelutions. Calibration followed five-point curves (5–80 ppb) with internal (methyl-d3-tert-butyl ether) and surrogate standards as per EPA 8260.
Key Results and Discussion
The Rtx-VMS column achieved baseline separation of target oxygenates and gasoline interferents. Relative standard deviations for response factors were generally below 12%, meeting EPA acceptance criteria. Most oxygenates exhibited recoveries between 90% and 105% in a 1 ppm gasoline matrix spiked at 5 ppb. Exceptions included acetone (62% recovery) and a slightly elevated RSD for tert-butyl alcohol. Resolution of MTBE and TBA was confirmed, preventing false positives due to shared diagnostic ions.
Benefits and Practical Applications
By integrating an optimized purge-and-trap stage with GC-MS, laboratories can:
- Detect and quantify trace fuel oxygenates in complex samples with high confidence
- Reduce the need for secondary confirmatory analyses
- Support regulatory monitoring of contaminated sites and wastewater streams
Future Trends and Opportunities
Advancements in stationary phase chemistry and mass spectrometric techniques will further improve selectivity and sensitivity for emerging oxygenate compounds. Automation of sample introduction and data processing could enhance throughput. Additionally, coupling with high-resolution MS may address challenges posed by increasingly complex environmental matrices.
Conclusion
Modifying EPA purge-and-trap GC-MS methods with an Rtx-VMS column offers a robust solution for trace oxygenate analysis in petroleum-contaminated water. The approach delivers reliable separation, quantification, and compliance with regulatory performance criteria.
Instrumentation Used
- O.I. Analytical 4560 Purge-and-Trap Concentrator with 4551A Autosampler
- Agilent 5890 Series II GC coupled to 5971A MS detector
- Rtx-VMS capillary column (30 m×0.25 mm ID×1.4 µm)
References
[1] English C., Cox C., Dorman F., Patwardhan D., “Analysis of Gasoline Oxygenates Using a New Capillary Column Stationary Phase,” Pittsburgh Conference 2001, Session 199.
[2] Happel A.M., Beckenbach E.H., Halden R.U., “MTBE Impacts to California Groundwater Resources,” LLNL Report UCRL-AR-130897, 1988.
[3] U.S. EPA, Method 8260, “Volatile Organic Compounds by GC/MS,” Rev. 0, July 1992.
[4] U.S. EPA, Method 8260B, “Volatile Organic Compounds by GC/MS,” Rev. 2, Dec. 1996.
[5] English C.M., Dorman F.L., Stidsen G.B., “Analysis of Gasoline Oxygenates by EPA Method 8260B,” Pittsburgh Conference 2003, Session 590-6P.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Fast, Optimized GC Purge-and-Trap Analysis of Volatiles for Soil and Water Methods
2018|Agilent Technologies|Applications
Featured Application: Optimized volatiles analysis on an Rtx-VMS column with Restek CRMs means better VOC data in less time Fast, Optimized GC Purge-and-Trap Analysis of Volatiles for Soil and Water Methods • Save set-up time—we’ve optimized volatiles analysis for you!…
Key words
tert, tertether, etherbutylbenzene, butylbenzenemethyl, methylxylene, xylenebutyl, butylchloride, chlorideethyl, ethylamyl, amylcarbon, carbonalcohol, alcoholmethacrylate, methacrylatestyrene, styrenetba, tbavms
Optimizing the Analysis of Volatile Organic Compounds
2003|Restek|Guides
Technical Guide 1 Optimizing the Analysis of Volatile Organic Compounds Inside: EPA Method Definitions State GRO Methods Contract Laboratory Program (CLP) The Love Canal Scandal Purge and Trap Theory Sequences and Flow Paths of the Purge and Trap Unit Purge…
Key words
pack, packxylene, xylenertx, rtxtert, tertpid, pidpurge, purgebutylbenzene, butylbenzeneelcd, elcdeach, eachchloride, chloridetrap, trapbenzene, benzenedata, dataacetate, acetatevgc
Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters
2015|Agilent Technologies|Applications
Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters Application Note P a g e |1 Abstract Water quality laboratories across the nation are faced with both a rising level of…
Key words
ether, etherbutylbenzene, butylbenzenechloride, chlorideethyl, ethylpurge, purgeamyl, amyltemp, tempbutyl, butylvinyl, vinylmethyl, methyldichlorodifluoromethane, dichlorodifluoromethanersd, rsdchloromethane, chloromethanemethacrylate, methacrylateavg
Volatiles by EPA Method 8260 on Rxi®-624Sil MS (30 m, 0.25 mm ID, 1.40 μm)
|Agilent Technologies|Applications
Volatiles by EPA Method 8260 on Rxi®-624Sil MS (30 m, 0.25 mm ID, 1.40 µm) Resolution of critical pairs, low bleed, and high inertness make this a great column for volatiles! Column Sample Conc.: Injection Inj. Temp.: Purge and Trap…
Key words
butylbenzene, butylbenzeneacetate, acetatexylene, xylenetert, tertether, etherbutyl, butyldbcp, dbcpethyl, ethylchloroprene, chloroprenemibk, mibkdipe, dipemethacrylonitrile, methacrylonitriledibromofluoromethane, dibromofluoromethanepropionitrile, propionitrilecumene