C1 to C10 Ethyl Esters

Applications |  | QuadrexInstrumentation
GC, GC columns, Consumables
Industries
Energy & Chemicals
Manufacturer
Quadrex

Summary

Significance of the Topic

Industrial solvents such as ethyl esters are widely used in coatings, adhesives, and pharmaceuticals. Accurate characterization of C1–C10 ethyl esters is essential for quality control, regulatory compliance, and process optimization in industrial laboratories.

Study Objectives and Overview

This application note demonstrates a gas chromatography (GC) method for rapid separation and analysis of C1 through C10 ethyl esters in solvent mixtures. The goal is to provide a robust protocol that ensures baseline resolution, reproducibility, and efficient sample throughput.

Methodology and Instrumentation

  • Column: BTR-CW bonded Carbowax/BTR, oxygen resistant, 30 m × 0.25 mm I.D. × 0.25 µm film thickness
  • Temperature program: Initial 80 °C (4 min hold), ramp at 8 °C/min to 200 °C
  • Injector and detector temperature: 250 °C
  • Split flow: 40 cc/min
  • Sample volume: 0.4 µL

Used Instrumentation

  • GC capillary column (BTR-CW-30-0.25F)
  • Gas chromatograph with split injector and flame ionization detector

Main Results and Discussion

This method achieves clear separation of C1–C10 ethyl esters within a single run. Retention order follows increasing carbon chain length. Peak shapes are sharp with minimal tailing, and run-to-run reproducibility is high. The conditions reliably resolve even components with close boiling points.

Benefits and Practical Applications

  • High resolution screening of volatile esters
  • Rapid analysis suitable for routine quality control
  • Robust performance in industrial laboratory environments
  • Minimal solvent and sample consumption

Future Trends and Potential Applications

  • Integration with mass spectrometry for enhanced sensitivity and identification
  • Use of shorter, high-efficiency columns for faster throughput
  • Automation of sample preparation and data processing
  • Development of green chromatography methods to reduce solvent use

Conclusion

The described GC method using a BTR-CW capillary column offers a straightforward approach to analyze C1–C10 ethyl esters in industrial solvents. It provides reliable separation, reproducibility, and efficiency, making it well-suited for routine QC applications.

Reference

No external literature references were cited in the original text.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Performance of Method 8270 Using Hydrogen Carrier Gas
Your local gas generation partner Performance of Method 8270 Using Hydrogen Carrier Gas www.peakscientific.com Performance of Method 8270 Using Hydrogen Carrier Gas Introduction USEPA Method 8270 [1] for semivolatiles is used by laboratories to measure a mixture of acids, bases,…
Key words
pass, passscion, scionhydrogen, hydrogengas, gassplit, splitspcc, spcccheck, checkinjection, injectioncompounds, compoundscarrier, carrierpulsed, pulsedsludge, sludgeenviropro, enviroprocalibration, calibrationhexachlorocyclopentadiene
Performance of EPA Method 8270 using Hydrogen Carrier Gas on SCION GC MS
Performance of EPA Method 8270 using Hydrogen Carrier Gas on SCION GC-MS Application Note AN0018 INTRODUCTION United States Environmental Protection Agency (USEPA) Method 8270 is an analytical method for the detection of semi-volatile organic compounds in solid waste matrices, soils,…
Key words
scion, scionbenzo, benzopyrene, pyrenehydrogen, hydrogendichloromethane, dichloromethaneminimise, minimiserobust, robustanalyser, analyserpropylamine, propylaminenitroso, nitrosopentachlorophenol, pentachlorophenolcarrier, carrierperylene, peryleneion, ionsource
Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS
Application Note # CA 284780 Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS Abstract The analysis analysis of of semi-volatile semi-volatileorganic organiccompounds compounds(SOCs) (SOCs) using EPA EPA Method Method 8270 8270presents presentschallenges…
Key words
scion, scionbruker, brukerreports, reportscbs, cbsmsws, mswsphone, phonethescion, thescionscan, scanbenzo, benzosim, simmethod, methodfull, fullepa, epaascompound, ascompoundbasedscanning
Fast USEPA 8270 Semivolatiles Analysis Using the 6890N/5975 inert GC/MSD
Fast USEPA 8270 Semivolatiles Analysis Using the 6890N/5975 inert GC/MSD Application Environmental Author Mike Szelewski Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808 Abstract The analysis of semivolatiles using EPA Method 8270 presents challenges due to the simultaneous measurement…
Key words
msd, msdrtl, rtltemp, temprrfs, rrfsacross, acrosslinearity, linearityinlet, inletinert, inertfaster, fasterrun, runtime, timepulse, pulsesampling, samplingoven, ovenoutlet
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike