Fast Analysis of Organochlorine Pesticides Standard Using Conventional GC Instrumentation
Applications | 2013 | Thermo Fisher ScientificInstrumentation
Analysis of organochlorine pesticides is critical for environmental monitoring, food safety, and regulatory compliance. Fast and reliable chromatographic methods support high sample throughput in routine and research laboratories. By optimizing column dimensions and transfer parameters, laboratories can achieve significant reductions in analysis time without compromising separation quality or reproducibility.
This study demonstrates the transfer of an EPA Method 8081 organochlorine pesticide standard analysis from a conventional 30 m × 0.25 mm × 0.25 µm GC column to a Thermo Scientific TraceGOLD TG-5SilMS fast GC column (20 m × 0.15 mm × 0.15 µm). The goals were to reduce run time, maintain resolution, and validate retention time reproducibility under varying linear velocities.
Working standard preparation involved diluting a 200 µg/mL OCP mix in hexane/toluene to 50 µg/mL in hexane. Chromatographic conditions were:
Column comparison and method transfer employed equations to match temperature gradient, phase ratio, linear velocity, and column length. Three methods were evaluated:
Method II achieved a 30 % reduction in analysis time with only a 4 % decrease in resolution compared to method I. Increasing linear velocity in method III further reduced run time by 50 % relative to the standard while improving resolution by approximately 7 %. Column head pressures remained within instrument limits (190 to 472 kPa). Six replicate injections for each method yielded excellent retention time reproducibility (%RSD generally below 0.2 % for key OCP analytes).
Transferring OCP analysis to a fast GC column offers:
This approach is well suited for high-throughput environmental, food, and quality control laboratories following EPA guidelines.
Emerging developments may include further optimization of column materials and dimensions, implementation of alternative carrier gases (hydrogen or nitrogen), and integration with mass spectrometry for enhanced selectivity. Advances in software algorithms for automated method transfer and real-time pressure monitoring will support even faster and more robust pesticide analyses.
Method transfer to a Thermo Scientific TraceGOLD TG-5SilMS fast GC column significantly reduces analysis time for organochlorine pesticides while maintaining or improving chromatographic resolution and reproducibility. This strategy enables laboratories to increase throughput without altering their GC systems or compromising data quality.
EPA Method 8081 Organochlorine Pesticides by Gas Chromatography
GC, GC columns, Consumables
IndustriesEnvironmental
ManufacturerThermo Fisher Scientific
Summary
Significance of Topic
Analysis of organochlorine pesticides is critical for environmental monitoring, food safety, and regulatory compliance. Fast and reliable chromatographic methods support high sample throughput in routine and research laboratories. By optimizing column dimensions and transfer parameters, laboratories can achieve significant reductions in analysis time without compromising separation quality or reproducibility.
Study Objectives and Overview
This study demonstrates the transfer of an EPA Method 8081 organochlorine pesticide standard analysis from a conventional 30 m × 0.25 mm × 0.25 µm GC column to a Thermo Scientific TraceGOLD TG-5SilMS fast GC column (20 m × 0.15 mm × 0.15 µm). The goals were to reduce run time, maintain resolution, and validate retention time reproducibility under varying linear velocities.
Methodology and Instrumentation
Working standard preparation involved diluting a 200 µg/mL OCP mix in hexane/toluene to 50 µg/mL in hexane. Chromatographic conditions were:
- Instrument: Thermo Scientific TRACE GC Ultra with FID detection
- Injection: Split mode, 30:1, injector at 280 °C
- Detector: FID at 300 °C, air 350 mL/min, H2 35 mL/min, N2 30 mL/min
- Software: Thermo Scientific Xcalibur
Column comparison and method transfer employed equations to match temperature gradient, phase ratio, linear velocity, and column length. Three methods were evaluated:
- Standard (I): 30 m × 0.25 mm × 0.25 µm, He 1.2 mL/min, 30 °C/min ramps, 15.17 min run time
- Fast (II): 20 m × 0.15 mm × 0.15 µm, He 0.6 mL/min, 45 °C/min and 15 °C/min ramps, 10.91 min run time
- Faster (III): same column, He 1.0 mL/min, 60 °C/min and 20 °C/min ramps, 8.08 min run time
Results and Discussion
Method II achieved a 30 % reduction in analysis time with only a 4 % decrease in resolution compared to method I. Increasing linear velocity in method III further reduced run time by 50 % relative to the standard while improving resolution by approximately 7 %. Column head pressures remained within instrument limits (190 to 472 kPa). Six replicate injections for each method yielded excellent retention time reproducibility (%RSD generally below 0.2 % for key OCP analytes).
Benefits and Practical Applications
Transferring OCP analysis to a fast GC column offers:
- Up to 50 % shorter run times
- Consistent separation quality and resolution
- High reproducibility in retention times
- No modifications to existing GC configuration
This approach is well suited for high-throughput environmental, food, and quality control laboratories following EPA guidelines.
Future Trends and Opportunities
Emerging developments may include further optimization of column materials and dimensions, implementation of alternative carrier gases (hydrogen or nitrogen), and integration with mass spectrometry for enhanced selectivity. Advances in software algorithms for automated method transfer and real-time pressure monitoring will support even faster and more robust pesticide analyses.
Conclusion
Method transfer to a Thermo Scientific TraceGOLD TG-5SilMS fast GC column significantly reduces analysis time for organochlorine pesticides while maintaining or improving chromatographic resolution and reproducibility. This strategy enables laboratories to increase throughput without altering their GC systems or compromising data quality.
Reference
EPA Method 8081 Organochlorine Pesticides by Gas Chromatography
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Fast Analysis of Organochlorine Pesticides Standard Using Conventional GC Instrumentation
2013|Thermo Fisher Scientific|Applications
Anila I. Khan, Thermo Fisher Scientific, Runcorn, UK Key Words Fast GC analysis, organochlorine pesticides, OCP, EPA Method 8081, Fast GC column, TraceGOLD TG-5SilMS Abstract This application note compares the performance of a 0.15 mm internal diameter (i.d.) GC column…
Key words
original, originalorganochlorine, organochlorineocp, ocpcolumn, columniii, iiinew, newconditions, conditionsmethod, methodthermo, thermofast, fastmean, meanscientific, scientificstandard, standardvelocity, velocitypesticide
Fast Analysis of the Florida TRPH C8–C40 Standard Using Conventional GC Instrumentation
2013|Thermo Fisher Scientific|Applications
Anila I. Khan, Thermo Fisher Scientific, Runcorn, UK Key Words Fast GC analysis, Florida TRPH C8-C40 standard, 100% dimethylpolysiloxane, Fast GC column, TraceGOLD TG-1MS Abstract This application note compares the performance of a 0.15 mm internal diameter (i.d.) GC column…
Key words
original, originalcolumn, columnvelocity, velocitymethod, methodnew, newconditions, conditionsfast, fastmean, meanthermo, thermoscientific, scientificlinear, lineartrph, trphtransfer, transferiii, iiirsd
Fast Analysis of FAMEs Using Conventional GC Instrumentation
2013|Thermo Fisher Scientific|Applications
Anila I. Khan, Thermo Fisher Scientific, Runcorn, UK Key Words Fast GC analysis, FAMEs, Fast GC column, polyethylene glycol (PEG), TraceGOLD TG-WaxMS Abstract This application note compares the performance of a 0.15 mm internal diameter (i.d.) GC column. with that…
Key words
original, originalcolumn, columnwaxms, waxmsvelocity, velocitynew, newtracegold, tracegoldconditions, conditionsfast, fastmean, meanthermo, thermoscientific, scientificmethod, methodlinear, lineartransfer, transferiii
Fast Analysis of FAMEs Using Conventional GC Instrumentation
2013|Thermo Fisher Scientific|Applications
Anila I. Khan, Thermo Fisher Scientific, Runcorn, UK Key Words Fast GC analysis, FAMEs, Fast GC column, polyethylene glycol (PEG), TraceGOLD TG-WaxMS Abstract This application note compares the performance of a 0.15 mm internal diameter (i.d.) GC column. with that…
Key words
original, originalcolumn, columnwaxms, waxmsvelocity, velocitynew, newtracegold, tracegoldconditions, conditionsthermo, thermofast, fastmean, meanscientific, scientificmethod, methodlinear, lineartransfer, transferiii