Fast Analysis of FAMEs Using Conventional GC Instrumentation
Applications | 2013 | Thermo Fisher ScientificInstrumentation
Gas chromatography remains a cornerstone technique in analytical chemistry for separating complex mixtures. The analysis of fatty acid methyl esters in biological and industrial samples is crucial for nutritional, environmental and quality control applications. Reducing analysis time without sacrificing resolution can significantly increase laboratory throughput and cost efficiency.
This study compares two polyethylene glycol GC columns with different internal diameters and film thicknesses for the fast separation of a 14 component FAME standard C8–C24. The goal is to transfer a conventional method to a fast GC column and evaluate changes in analysis speed, resolution, pressure and reproducibility.
A 500 µg/mL standard solution of 14 FAMEs was prepared in dichloromethane. Three methods were evaluated:
Method transfer equations were applied to maintain equivalent temperature gradients, phase ratios and hold times between column formats.
Transitioning to the fast column reduced analysis time by 30% while improving resolution by approximately 7%. Increasing the carrier gas velocity further delivered a 50% reduction in run time with no loss in chromatographic performance. Column head pressures rose from 170 kPa to 430 kPa but remained within the operating limits of conventional GC systems. Retention time reproducibility was excellent, with relative standard deviations below 0.1% for all FAME components across six replicates.
Further miniaturization of column dimensions and optimization of temperature ramp profiles may yield even faster separations. This approach can be extended to other analyte classes in environmental, food and pharmaceutical testing. Integration with mass spectrometry or advanced detector technologies offers opportunities for comprehensive high-throughput profiling.
By transferring a conventional FAME analysis to a fast GC column and optimizing carrier gas velocity, analysis time was halved without compromising resolution or reproducibility. Key parameters for successful method transfer include column dimensions, film thickness, linear velocity and temperature programming.
No additional literature references were provided.
GC
IndustriesFood & Agriculture
ManufacturerThermo Fisher Scientific
Summary
Significance of the Topic
Gas chromatography remains a cornerstone technique in analytical chemistry for separating complex mixtures. The analysis of fatty acid methyl esters in biological and industrial samples is crucial for nutritional, environmental and quality control applications. Reducing analysis time without sacrificing resolution can significantly increase laboratory throughput and cost efficiency.
Objectives and Study Overview
This study compares two polyethylene glycol GC columns with different internal diameters and film thicknesses for the fast separation of a 14 component FAME standard C8–C24. The goal is to transfer a conventional method to a fast GC column and evaluate changes in analysis speed, resolution, pressure and reproducibility.
Methodology
A 500 µg/mL standard solution of 14 FAMEs was prepared in dichloromethane. Three methods were evaluated:
- Standard method on a 30 m × 0.25 mm × 0.25 µm column with a 19.5 min oven program.
- Fast method on a 20 m × 0.15 mm × 0.15 µm column with a 13.13 min program maintaining linear velocity at 30 cm/s.
- Faster method on the same fast column with increased linear velocity at 43 cm/s and a 10.25 min program.
Method transfer equations were applied to maintain equivalent temperature gradients, phase ratios and hold times between column formats.
Instrumentation Used
- Thermo Scientific TRACE GC Ultra with split/splitless injector and flame ionization detector.
- TraceGOLD TG-WaxMS columns in standard and fast formats.
- Thermo Scientific autosampler with fixed needle syringe and standard consumables.
- Data acquisition and processing with Thermo Scientific Xcalibur software.
Key Results and Discussion
Transitioning to the fast column reduced analysis time by 30% while improving resolution by approximately 7%. Increasing the carrier gas velocity further delivered a 50% reduction in run time with no loss in chromatographic performance. Column head pressures rose from 170 kPa to 430 kPa but remained within the operating limits of conventional GC systems. Retention time reproducibility was excellent, with relative standard deviations below 0.1% for all FAME components across six replicates.
Benefits and Practical Applications
- Enhanced sample throughput and reduced per-sample costs.
- Maintained high resolution and reproducibility.
- Straightforward method transfer requiring no hardware changes.
- Compatibility with existing laboratory workflows and GC systems.
Future Trends and Potential Applications
Further miniaturization of column dimensions and optimization of temperature ramp profiles may yield even faster separations. This approach can be extended to other analyte classes in environmental, food and pharmaceutical testing. Integration with mass spectrometry or advanced detector technologies offers opportunities for comprehensive high-throughput profiling.
Conclusion
By transferring a conventional FAME analysis to a fast GC column and optimizing carrier gas velocity, analysis time was halved without compromising resolution or reproducibility. Key parameters for successful method transfer include column dimensions, film thickness, linear velocity and temperature programming.
Reference
No additional literature references were provided.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Fast Analysis of FAMEs Using Conventional GC Instrumentation
2013|Thermo Fisher Scientific|Applications
Anila I. Khan, Thermo Fisher Scientific, Runcorn, UK Key Words Fast GC analysis, FAMEs, Fast GC column, polyethylene glycol (PEG), TraceGOLD TG-WaxMS Abstract This application note compares the performance of a 0.15 mm internal diameter (i.d.) GC column. with that…
Key words
original, originalcolumn, columnwaxms, waxmsvelocity, velocitynew, newtracegold, tracegoldconditions, conditionsthermo, thermofast, fastmean, meanscientific, scientificmethod, methodlinear, lineartransfer, transferiii
Fast Analysis of the Florida TRPH C8–C40 Standard Using Conventional GC Instrumentation
2013|Thermo Fisher Scientific|Applications
Anila I. Khan, Thermo Fisher Scientific, Runcorn, UK Key Words Fast GC analysis, Florida TRPH C8-C40 standard, 100% dimethylpolysiloxane, Fast GC column, TraceGOLD TG-1MS Abstract This application note compares the performance of a 0.15 mm internal diameter (i.d.) GC column…
Key words
original, originalcolumn, columnvelocity, velocitymethod, methodnew, newconditions, conditionsfast, fastmean, meanthermo, thermoscientific, scientificlinear, lineartrph, trphtransfer, transferiii, iiirsd
Optimizing GC Parameters for Faster Separations with Conventional Instrumentation
2013|Thermo Fisher Scientific|Technical notes
Anila I. Khan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Key Words TraceGOLD fast GC analysis Abstract Today’s laboratories are increasingly looking for ways to speed up analysis time in order to increase sample throughput and reduce analysis costs, without compromising…
Key words
column, columnvelocity, velocityminutes, minutesthickness, thicknessfilm, filmtracegold, tracegoldlength, lengthchromatographic, chromatographiclinear, lineartemperature, temperatureramp, rampcarrier, carrierreduce, reduceefficiency, efficiencyoriginal
Fast Analysis of Organochlorine Pesticides Standard Using Conventional GC Instrumentation
2013|Thermo Fisher Scientific|Applications
Anila I. Khan, Thermo Fisher Scientific, Runcorn, UK Key Words Fast GC analysis, organochlorine pesticides, OCP, EPA Method 8081, Fast GC column, TraceGOLD TG-5SilMS Abstract This application note compares the performance of a 0.15 mm internal diameter (i.d.) GC column…
Key words
original, originalorganochlorine, organochlorineocp, ocpcolumn, columniii, iiinew, newconditions, conditionsmethod, methodfast, fastmean, meanthermo, thermoscientific, scientificstandard, standardvelocity, velocitypesticide