Analysis of Volatile Organic Compounds using GC/MS
Applications | | ShimadzuInstrumentation
Monitoring volatile organic compounds (VOCs) at trace levels is critical for environmental protection, water quality assessment, and regulatory compliance. Accurate measurement of VOCs supports contamination tracking, public health safeguards, and industrial quality control.
This application study evaluates a combined Purge & Trap gas chromatography/mass spectrometry (GC/MS) approach for quantifying a suite of 21 VOCs, including gaseous analytes such as chloromethane, vinyl chloride, and 1,3-butadiene. The goal is to demonstrate robust detection and separation at low parts-per-trillion (ppt) concentrations in aqueous matrices.
The total ion chromatogram and individual SIM traces confirmed baseline separation and clear peak identification for all 20 monitored VOCs at 10 ng/L levels. Early-eluting compounds overlapped with solvent or water peaks, causing baseline fluctuations, but selective ion monitoring allowed reliable quantitation. Retention times ranged from 5.5 to 25 min, demonstrating effective chromatographic resolution.
Emerging developments may focus on automated trap loading, portable GC/MS systems for field deployment, expanded compound libraries for broader screening, and coupling with data analytics for real-time monitoring and rapid environmental assessments.
The optimized Purge & Trap GC/MS method provides sensitive, selective, and reproducible quantification of a wide range of VOCs at ppt concentrations. This approach is well suited for routine environmental monitoring, water quality control, and industrial compliance applications.
GC/MSD, Purge and Trap, GC/SQ
IndustriesEnvironmental
ManufacturerShimadzu
Summary
Importance of the topic
Monitoring volatile organic compounds (VOCs) at trace levels is critical for environmental protection, water quality assessment, and regulatory compliance. Accurate measurement of VOCs supports contamination tracking, public health safeguards, and industrial quality control.
Objectives and study overview
This application study evaluates a combined Purge & Trap gas chromatography/mass spectrometry (GC/MS) approach for quantifying a suite of 21 VOCs, including gaseous analytes such as chloromethane, vinyl chloride, and 1,3-butadiene. The goal is to demonstrate robust detection and separation at low parts-per-trillion (ppt) concentrations in aqueous matrices.
Methodology
- Sample preparation by dissolving gaseous standards into chilled methanol
- Purge & Trap conditions: purge temperature 35°C, purge time 2.5 min, flow rate 30 mL/min, no dry purge
- Cryo-focusing at –190°C, desorption at 200°C for 3 min, bake at 200°C for 20 min
- SIM detection with a dwell time of 0.2 s to enhance selectivity
Instrumentation
- Gas chromatograph: Aquatic column, 60 m × 0.25 mm I.D., 1.0 µm film thickness; temperature program from 40°C (5 min) to 220°C at 7°C/min
- Mass spectrometer: scan range m/z 29–300, selected ion monitoring for target masses
- Purge & Trap module: GL-TRAP 1, with line and valve temperatures at 180°C
Main results and discussion
The total ion chromatogram and individual SIM traces confirmed baseline separation and clear peak identification for all 20 monitored VOCs at 10 ng/L levels. Early-eluting compounds overlapped with solvent or water peaks, causing baseline fluctuations, but selective ion monitoring allowed reliable quantitation. Retention times ranged from 5.5 to 25 min, demonstrating effective chromatographic resolution.
Benefits and practical applications
- Detection limits at ppt levels enable trace analysis in drinking water and industrial effluents
- SIM mode reduces matrix interferences and improves quantitation accuracy
- Purge & Trap integration minimizes manual sample handling and increases throughput
Future trends and opportunities
Emerging developments may focus on automated trap loading, portable GC/MS systems for field deployment, expanded compound libraries for broader screening, and coupling with data analytics for real-time monitoring and rapid environmental assessments.
Conclusion
The optimized Purge & Trap GC/MS method provides sensitive, selective, and reproducible quantification of a wide range of VOCs at ppt concentrations. This approach is well suited for routine environmental monitoring, water quality control, and industrial compliance applications.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of Volatile Hazardous Air Pollutants and Ozone Precursors Using the Cryofocusing GC-MS/MS (MRM) Method
2019|Shimadzu|Applications
LAAN-C-XX-E038 Application Note No. 56 Analysis of Volatile Hazardous Air Pollutants and Ozone Precursors Using the Cryofocusing GC-MS/MS (MRM) Method Yasuji Watanabe *1, Akira Suzuki *2 Environment Environment 1. Introduction Environmental monitoring of volatile organic compounds in the atmosphere has…
Key words
unit, unitmrm, mrminterference, interferencesubstance, substancedrift, driftunknown, unknownpseudo, pseudodichlorobenzene, dichlorobenzenelow, lowbaseline, baselinesim, simchloride, chloridecyclopentane, cyclopentanepinene, pinenexylene
Analysis of Residual Solvents – Class 1, Class 2A, Class 2B – in Pharmaceuticals Using Headspace-GC/MS
2014|Shimadzu|Applications
LAAN-A-MS-E034 Application News M268 Gas Chromatography Mass Spectrometry Analysis of Residual Solvents – Class 1, Class 2A, Class 2B – in Pharmaceuticals Using Headspace-GC/MS No. Residual solvents in pharmaceuticals are defined as organic volatile chemicals that are used or produced…
Key words
eic, eicsim, simtetralin, tetralinnitromethane, nitromethanetetrachloride, tetrachloridemethylcyclohexane, methylcyclohexaneequilibrating, equilibratingtrichloroethene, trichloroethenecyclohexane, cyclohexanepyridine, pyridinechlorobenzene, chlorobenzenetetrahydrofuran, tetrahydrofurannews, newsethylbenzene, ethylbenzenechloroform
US EPA Method 8260 With the Tekmar Lumin and the Agilent 7890B GC/5977A MS
2018|Agilent Technologies|Applications
Application Note US EPA Method 8260 With the Tekmar Lumin and the Agilent 7890B GC/5977A MS Author Roger Bardsley Applications Chemist, Teledyne Tekmar Abstract US EPA Method 8260 is used primarily by environmental labs for the analysis of Volatile Organic…
Key words
purge, purgetemp, tempbake, bakeppb, ppbsurrogate, surrogatelumin, luminrsd, rsdwater, watertert, tertacetate, acetatemdl, mdlether, etherdesorb, desorbvariable, variabletime
Analysis of Potential Genotoxic Impurities in Active Pharmaceutical Ingredients (5) - Analysis of Alkyl Halides -
2012|Shimadzu|Applications
LAAN-J-MS-E046 GC-MS Gas Chromatograph - Mass Spectrometer Analysis of Potential Genotoxic Impurities in Active Pharmaceutical Ingredients (5) - Analysis of Alkyl Halides - 46 Alkyl halides are used as an alkylating agent for raw ingredients in the synthesis of pharmaceuticals…
Key words
iodomethane, iodomethanedichloroethylene, dichloroethylenechloromethane, chloromethanevinyl, vinylhalides, halidesspeed, speedchloride, chloridetime, timealkyl, alkylevent, eventmode, modecis, cisscan, scaninject, injectalkylating