US EPA Method 8260 With the Tekmar Lumin and the Agilent 7890B GC/5977A MS

Applications | 2018 | Agilent TechnologiesInstrumentation
GC/MSD, Purge and Trap, GC/SQ
Industries
Environmental
Manufacturer
Agilent Technologies, Teledyne LABS

Summary

Significance of Topic


Volatile organic compounds (VOCs) analysis in aqueous and soil samples is essential for environmental monitoring and regulatory compliance. Efficient water removal prior to GC/MS analysis improves peak resolution and instrument longevity.

Objectives and Overview


This study evaluates the performance of the Tekmar Lumin Purge and Trap concentrator equipped with a Moisture Control System for VOC analysis following US EPA Methods 8260 and 5030, using an Agilent 7890B GC with 5977A Mass Spectrometer. Goals include assessing water vapor reduction, cycle time, detection limits, accuracy, precision, and chromatographic quality without dry purge.

Methodology


Sample Preparation and Calibration
  • Standards prepared from Restek MegaMix, acetate, oxygenate and VOA mixes; ketones and t-butanol spiked at elevated levels.
  • Calibration curves for target compounds ranged from sub-ppb to hundreds of ppb; internal standards and surrogate standards were added for quantification and quality control.

Used Instrumentation


  • Tekmar Lumin Purge and Trap concentrator with integrated Moisture Control System and AQUATek 100 autosampler
  • Agilent 7890B Gas Chromatograph equipped with a DB-624 UI column (20 m × 0.18 mm, 1 µm) and helium carrier gas
  • Agilent 5977A Mass Spectrometer operating in full-scan mode (35–270 m/z) with transfer line and source maintained at elevated temperature

Main Results and Discussion


  • The Moisture Control System removed water vapor effectively, eliminating the need for a dry purge and preventing chromatographic artifacts.
  • Optimized trap cooling design reduced sample cycle time, increasing throughput over a 12-hour period.
  • Calibration linearity was excellent, with relative standard deviations below 16% for most compounds; method detection limits ranged from 0.04 to 1.4 ppb.
  • Accuracy averaged near 100% for over 100 VOCs and surrogates; precision (%RSD) remained below 15% in all cases.
  • Chromatograms of 0.5 ppb and 30 ppb standards demonstrated sharp, well-resolved peaks without water interference.
  • Internal and surrogate standard responses remained stable over nine hours, with %RSD below 5% for most analytes.

Benefits and Practical Applications


  • Advanced water management extends GC/MS uptime and reduces maintenance requirements.
  • Elimination of dry purge simplifies workflows and shortens analysis cycles.
  • Enhanced sensitivity and reproducibility support trace-level environmental monitoring and compliance testing.

Future Trends and Opportunities


Further integration of moisture control technologies into multidimensional and high-throughput GC/MS platforms is anticipated. Automation of sample preparation and data analysis will streamline routine environmental monitoring. Future P&T designs may incorporate real-time moisture sensing and adaptive purge parameters for optimized performance.

Conclusion


The Tekmar Lumin Purge and Trap concentrator with Moisture Control System enables reliable VOC analysis by US EPA Methods 8260 and 5030 without dry purge. The approach delivers low detection limits, robust accuracy and precision, and increased laboratory productivity through reduced cycle times.

References


  • US EPA Office of Solid Waste SW-846 Method 8260B Revision 2 December 1996
  • US EPA Office of Solid Waste SW-846 Method 8260C Revision 3 August 2006
  • US EPA Office of Solid Waste SW-846 Method 5030B Revision 2 December 1996
  • US EPA Office of Solid Waste SW-846 Method 5030C Revision 3 May 2003

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
US EPA Method 8260 with the Tekmar Lumin and the Agilent 7890B GC / 5977A MS
US EPA Method 8260 with the Tekmar Lumin and the Agilent 7890B GC / 5977A MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method 8260 is used primarily by environmental labs…
Key words
temp, tempppb, ppbpurge, purgebake, bakeprecision, precisionrsd, rsdaccuracy, accuracyacetate, acetatetert, tertlumin, luminmdl, mdlether, etherdesorb, desorbsurrogate, surrogatevariable
US EPA Method 8260 with the Tekmar Lumin P&T Concentrator and Thermo Scientific™ TRACE™ 1310 GC and ISQ™ MS
US EPA Method 8260 with the Tekmar Lumin P&T Concentrator and Thermo Scientific™ TRACE™ 1310 GC and ISQ™ MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method 8260 is used primarily…
Key words
temp, tempppb, ppbpurge, purgebake, bakelumin, luminprecision, precisionrsd, rsdaccuracy, accuracyacetate, acetatemcs, mcstert, tertmdl, mdlether, etherdesorb, desorbvariable
US EPA Method 8260 with the Teledyne Tekmar Lumin P&T Concentrator, AQUATek LVA and the Agilent 7890B GC/5977A MSD
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1901 P+T VOC ENV US EPA Method 8260 with the Teledyne Tekmar Lumin P&T Concentrator, AQUATek LVA and the Agilent 7890B GC/5977A MSD Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US…
Key words
temp, temppurge, purgeaquatek, aquatekbake, bakelva, lvalumin, lumindesorb, desorbacetate, acetateppb, ppbtert, terttekmar, tekmarteledyne, teledyneether, etherconcentrator, concentratorvariable
US EPA Method 524.2 with the Teledyne Tekmar Lumin P&T Concentrator and Agilent 7890B GC/5977A MS
Application Note US EPA Method 524.2 with the Teledyne Tekmar Lumin P&T Concentrator and Agilent 7890B GC/5977A MS Author Roger Bardsley Applications Chemist, Teledyne Tekmar Abstract US EPA Method 524.2 is used primarily by environmental labs for the analysis of…
Key words
lumin, lumintemp, tempbake, bakepurge, purgetekmar, tekmarteledyne, teledyneppb, ppbwater, waterether, etherdesorb, desorbvariable, variablersd, rsdmcs, mcstime, timemdl
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike