Xylenes

Applications | 2011 | Agilent TechnologiesInstrumentation
GC, GC columns, Consumables
Industries
Environmental
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


The accurate separation and quantitation of xylene isomers are essential in environmental monitoring of volatile organic compounds (VOCs). Regulatory frameworks such as US EPA Methods 524 and 8260 mandate reliable detection of ortho-, meta-, and para-xylene as well as ethylbenzene in water and air samples to ensure public safety and compliance.

Objectives and Overview of the Application Note


This application note evaluates the performance of the Agilent FactorFour VF-200ms GC column in resolving the four key analytes—ethylbenzene and the three xylene isomers—under conditions aligned with EPA Method requirements. The study highlights retention characteristics at a low isothermal oven temperature and examines the column’s selectivity and robustness.

Methodology


  • Technique: Capillary gas chromatography with hydrogen as carrier gas (≈1.0 mL/min)
  • Injection: Split mode (1:100), 1 µL sample in cyclohexane (0.1 % w/v)
  • Oven temperature: Isothermal at 40 °C
  • Pressure: Constant 60 kPa
  • Detection: Flame ionization detector (FID)

Used Instrumentation


  • Gas chromatograph equipped with Agilent FactorFour VF-200ms column (30 m × 0.25 mm, 1.0 µm film thickness, Part No. CP8860)
  • Split/splitless injector and FID module
  • Hydrogen supply configured for capillary flow at 1.0 mL/min

Main Results and Discussion


Under the selected conditions, the VF-200ms column achieved baseline separation of ethylbenzene, meta-, para-, and ortho-xylene. The chromatogram exhibited sharp, symmetric peaks with distinct retention order, demonstrating the column’s high selectivity for regulatory VOC analysis. The consistent retention times indicate excellent reproducibility across replicates.

Benefits and Practical Applications


  • Regulatory compliance: Meets or exceeds separation requirements for EPA Methods 524 and 8260
  • Enhanced throughput: Low oven temperature reduces cooling times between runs
  • Temperature robustness: Column rated to 350 °C, allowing flexibility for more complex VOC mixtures
  • Quantitative accuracy: Sharp peaks and minimal coelution improve integration precision

Future Trends and Potential Applications


Advances in comprehensive two-dimensional GC (GC×GC) and faster temperature programming are likely to further improve VOC profiling. Coupling with mass spectrometry will enhance sensitivity and confirmatory identification. Emerging stationary phases could offer even greater isomer selectivity for challenging environmental matrices.

Conclusion


The Agilent FactorFour VF-200ms column provides a reliable and efficient solution for separating ethylbenzene and xylene isomers under EPA-style analytical conditions. Its robust performance and wide temperature range make it a valuable tool for environmental laboratories focused on VOC compliance and quality control.

References


  • Agilent Technologies, Application Note A02215, 2011.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Aromatic volatile organic compounds
Aromatic volatile organic compounds
2011|Agilent Technologies|Applications
Aromatic volatile organic compounds Application Note Environmental Authors Introduction Agilent Technologies, Inc. GC analysis of nine aromatic volatiles uses an Agilent FactorFour VF-200ms column to achieve a separation in less than 12 minutes. Conditions Technique : GC Column : Agilent…
Key words
erthylbenzene, erthylbenzenexylene, xylenemeta, metaortho, orthopara, parachlorobenzene, chlorobenzenecourtesy, courtesyaromatic, aromaticbenzene, benzenetoluene, tolueneprinted, printedvolatile, volatileinjector, injectorusa, usaauthors
Aromatic volatiles
Aromatic volatiles
2011|Agilent Technologies|Applications
Aromatic volatiles Application Note Environmental Authors Introduction Agilent Technologies, Inc. Separation of 11 aromatic volatiles on an Agilent VF-200ms column in nine minutes. Conditions Technique : GC Column : Agilent FactorFour VF-200ms, 0.25 mm x 30 m (df = 1.0…
Key words
courtesy, courtesyvolatiles, volatilesaromatic, aromaticprinted, printedinjector, injectorusa, usaauthors, authorsprogram, programenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationpressure, pressuresolvent
Chlorinated hydrocarbons
Chlorinated hydrocarbons
2011|Agilent Technologies|Applications
Chlorinated hydrocarbons Application Note Environmental Authors Introduction Agilent Technologies, Inc. GC analysis of eight chlorinated hydrocarbons uses an Agilent FactorFour VF-200ms column to achieve a separation in 22 minutes. Conditions Technique : GC Column : Agilent FactorFour VF-200ms, 0.25 mm…
Key words
hexachloroethane, hexachloroethanehexachlorobutadiene, hexachlorobutadienehexachlorobenzene, hexachlorobenzenechlorinated, chlorinatedcourtesy, courtesyhydrocarbons, hydrocarbonsprinted, printedinjector, injectorusa, usaauthors, authorsprogram, programenvironmental, environmentalsize, sizetechnique, techniquecarrier
Basic aromatic compounds - Analysis of basic aromatic compounds
Basic aromatic compounds Analysis of basic aromatic compounds Application Note Environmental Authors Introduction Agilent Technologies, Inc. Using GC/MS, an Agilent FactorFour VF-200ms column separates eight basic aromatics in 25 minutes. Conditions Technique : GC Column : Agilent FactorFour VF-200ms, 0.25…
Key words
aromatic, aromaticbasic, basiccarbazole, carbazoledibenzofuran, dibenzofurancourtesy, courtesycompounds, compoundsprinted, printedinjector, injectorusa, usaauthors, authorsprogram, programenvironmental, environmentalsize, sizetechnique, techniquecarrier
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike