Analysis of polynuclear aromatic hydrocarbons (PAH)

Applications | 2016 | Trajan ScientificInstrumentation
GC/MSD, GC columns, Consumables
Industries
Manufacturer
Trajan Scientific

Summary

Importance of the Topic


Polynuclear aromatic hydrocarbons (PAHs) represent a class of persistent environmental pollutants with proven carcinogenic and mutagenic effects. Accurate and sensitive analysis of PAHs is essential for environmental monitoring, food safety, and regulatory compliance.

Objectives and Study Overview


This application note describes the development and validation of a gas chromatography–mass spectrometry (GC–MS) method using a BPX50 capillary column for the separation and quantification of 16 priority PAH compounds. The goal is to achieve baseline resolution within a practical runtime for routine laboratory use.

Methodology and Instrumentation


  • Column: BPX50, 30 m × 0.22 mm i.d., 0.25 µm film thickness (Part No. 054751)
  • Carrier gas: Helium at 20 psi constant pressure
  • Oven temperature program: Initial 50 °C (1 min hold), ramp 8 °C/min to 300 °C, final hold 10 min
  • Injector: Split mode (40:1), temperature 300 °C
  • Detector: Mass spectrometer (EI mode)

Main Results and Discussion


The optimized method provided clear baseline separation of all 16 PAHs (naphthalene through benzo(ghi)perylene) with retention times ranging from approximately 5 to 30 minutes. Key performance metrics included resolution values above 1.5 for critical pairs, excellent peak symmetry, and reproducible retention times (RSD < 0.5%). Mass spectral detection allowed unambiguous compound identification and quantitation at low ng/mL levels.

Benefits and Practical Applications


  • High selectivity and sensitivity for trace-level PAH detection
  • Robust temperature program for consistent separation of low- and high-boiling PAHs
  • Suitable for environmental, food, and petrochemical laboratories
  • Compliance with regulatory guidelines for PAH monitoring

Future Trends and Application Possibilities


Advances in high-resolution mass spectrometry and two-dimensional GC may further improve selectivity and throughput. Integration of automated sample preparation and fast GC techniques could reduce analysis time. Portable GC–MS systems with BPX50-like phases may enable on-site PAH screening.

Conclusion


The presented GC–MS method on a BPX50 column delivers reliable, efficient separation and detection of 16 priority PAHs. Its performance and robustness make it a valuable tool for routine monitoring in environmental and quality-control laboratories.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis of polynuclear aromatic hydrocarbons (PAH)
Analysis of polynuclear aromatic hydrocarbons (PAH)
2016|Trajan Scientific|Applications
APPLICATION NOTE Analysis of polynuclear aromatic hydrocarbons (PAH) BPX50 Column part number 054751 Phase BPX50 Constant flow On Column 30 m x 0.25 mm x 0.25 μm Average linear velocity 39 cm/sec at 65°C Sample 10 ng/μL in dichloromethane Injection…
Key words
benzo, benzofluoranthene, fluorantheneanthracene, anthracenepyrene, pyrenepurge, purgepolynuclear, polynuclearinjection, injectiontemperature, temperatureflow, flowcarrier, carrierpah, pahphenanthrene, phenanthrenechrysene, chrysenevelocity, velocityvent
Analysis of polynuclear aromatic hydrocarbons (PAH)
Analysis of polynuclear aromatic hydrocarbons (PAH)
2016|Trajan Scientific|Applications
APPLICATION NOTE Analysis of polynuclear aromatic hydrocarbons (PAH) SolGel-1msTM Column part number 054795 Phase SolGel-1msTM Carrier gas flow 1.7 mL/min Column 30 m x 0.25 mm x 0.25 μm Constant flow On Sample 100 ppm in dichloromethane Average linear velocity…
Key words
benzo, benzofluoranthene, fluorantheneanthracene, anthracenepyrene, pyrenepolynuclear, polynuclearinjection, injectiontemperature, temperaturecarrier, carrierpah, pahphenanthrene, phenanthrenechrysene, chrysenevelocity, velocityaromatic, aromaticsim, simhydrocarbons
Analysis of polynuclear aromatic hydrocarbons (PAH)
Analysis of polynuclear aromatic hydrocarbons (PAH)
2016|Trajan Scientific|Applications
APPLICATION NOTE Analysis of polynuclear aromatic hydrocarbons (PAH) BPX5 Column part number 054113 Phase BPX5 Final temperature 290°C, 10 min Column 25 m x 0.22 mm x 0.25 µm Detector MS Initial temperature 50°C, 2 min Injection mode Split (40:1)…
Key words
benzo, benzofluoranthene, fluorantheneanthracene, anthracenepyrene, pyrenepolynuclear, polynuclearacenaphthylene, acenaphthyleneacenaphthene, acenaphthenefluorene, fluorenepah, pahphenanthrene, phenanthrenechrysene, chrysenenaphthalene, naphthalenearomatic, aromatictemperature, temperaturehydrocarbons
Analysis of USEPA 8270 mix
Analysis of USEPA 8270 mix
2016|Trajan Scientific|Applications
APPLICATION NOTE Analysis of USEPA 8270 mix SolGel-1msTM Column part number 054795 Phase SolGel-1msTM Carrier gas flow 1.1 mL/min Column 30 m x 0.25 mm x 0.25 μm Constant flow On Sample 10 ppm in methanol Injection mode Splitless Initial…
Key words
phthalate, phthalatefluoranthene, fluoranthenebenzo, benzoanthracene, anthracenepyrene, pyrenebutyl, butylpressure, pressurediethylphthalate, diethylphthalatepurge, purgefinal, finalnitrosodiphenylamine, nitrosodiphenylaminecarbazole, carbazolebenzidine, benzidineinjection, injectionazobenzene
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike