Solvents - Fast analysis of volatile organic compounds to EPA 524

Applications | 2011 | Agilent TechnologiesInstrumentation
GC/MSD, GC columns, Consumables
Industries
Environmental
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Rapid detection of volatile organic compounds (VOCs) in environmental water samples is critical for regulatory compliance under EPA Method 524. Traditional capillary columns often suffer from long run times and extensive stationary phase bleed at elevated temperatures, compromising sensitivity and instrument performance.

Objectives and Study Overview


The study evaluates an Agilent CP-Select 624 CB column (0.15 mm × 25 m, 0.84 µm film) coupled with purge and trap sample preparation to achieve:
  • High chromatographic resolution for a broad range of chlorinated and aromatic solvents
  • Reduced analysis time compared to standard 0.32 mm columns
  • Minimized column bleed to enhance MS detection limits

Methodology and Instrumentation


Sample preparation:
  • Extraction of 50 mL water samples via purge and trap unit
  • Concentration of analytes at part-per-billion levels
Chromatographic conditions:
  • Agilent CP-Select 624 CB column (0.15 mm × 25 m, df = 0.84 µm)
  • Carrier gas: helium at 220 kPa
  • Temperature program: 35 °C (1 min) to 190 °C at 5 °C/min
Detection:
  • Mass selective detector (MSD)

Main Results and Discussion


  • Separation and identification of 53 target VOCs within 30 minutes
  • Sharp, well-resolved peaks with minimal background interference
  • Stationary phase bleed reduced by a factor of 10–15 versus 0.32 mm columns
  • Enhanced sensitivity and lower detection limits in MS detection

The combination of narrow-bore column geometry and optimized film thickness facilitated rapid elution and accurate quantification at ppb levels.

Benefits and Practical Applications


  • Increased sample throughput for environmental monitoring laboratories
  • Reduced maintenance due to lower detector contamination
  • Accurate quantification supports regulatory compliance
  • Method fully aligns with EPA 524 requirements for drinking water analysis

Future Trends and Applications


  • Integration with automated high-throughput purge and trap systems
  • Extension to emerging contaminants and complex sample matrices
  • Development of portable GC-MS platforms for on-site analysis
  • Application of data analytics and machine learning for automated peak identification

Conclusion


The optimized GC-MS method using the Agilent CP-Select 624 CB column and purge and trap extraction delivers rapid, sensitive, and reliable VOC analysis in water samples, enhancing laboratory efficiency and regulatory compliance.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
EPA Method 524.2 Drinking Water
EPA Method 524.2 Drinking Water
|Quadrex|Applications
GC CAPILLARY COLUMN APPLICATIONS: Environmental EPA Method 524.2 Drinking Water 1. 1,1-dichloroethylene 2. methylene chloride 3. trans 1,2-dichloroethylene 4. 1,1-dichloroethane 5. 2,2-dichloropropane 6. cis 1,2-dichloroethylene 7. bromochloromethane 8. chloroform 9. 1,1,1-trichloroethane 10. 1,1-dichloropropylene 11. carbon tetrachloride 12. benzene 13. 1,2-dichloroethane…
Key words
methylpolysiloxane, methylpolysiloxanecyanopropyl, cyanopropylphenyl, phenylmsd, msdepa, epadrinking, drinkingenvironmental, environmentalwater, watermethod
Determination of Volatile Organic Compounds in Water by Purge and Trap Gas Chromatography/Mass Spectrometry
Application Note Environmental Determination of Volatile Organic Compounds in Water by Purge and Trap Gas Chromatography/Mass Spectrometry Using the Chinese Ministry of Environmental Protection method HJ 639-2012 Author Youjuan Zhang Agilent Technologies (Shanghai) Co. Ltd. Shanghai 200131 P. R. China…
Key words
bake, bakepurge, purgetemperature, temperaturebutylbenzene, butylbenzenesim, simscan, scandesorb, desorbsweep, sweeplod, lodloq, loqrinse, rinsechloride, chloridetime, timeflow, flowdibromofluoromethane
Effect of Split Ratio on USEPA Method 8260 Compounds
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Effect of Split Ratio on USEPA Method 8260 Compounds Application Note Environmental Author Abstract Anne Jurek Applications Chemist EST Analytical Cincinnati, OH USEPA Method 8260 involves purging analytes out…
Key words
acetate, acetatetert, tertbutylbenzene, butylbenzenemethyl, methylether, etherbutyl, butylsur, surethyl, ethylamyl, amylchloride, chloridemethacrylate, methacrylatevinyl, vinyljurek, jurekxylene, xyleneanne
US EPA Method 524.2: Successful Measurement of Purgeable Organic Compounds in Drinking Water by Agilent 8860/5977B GC/MSD
Application Note Environmental US EPA Method 524.2: Successful Measurement of Purgeable Organic Compounds in Drinking Water by Agilent 8860/5977B GC/MSD Authors Bruce D. Quimby and Anastasia A. Andrianova Agilent Technologies, Inc. Abstract An Agilent 8860/5977B GC/MSD system coupled with a…
Key words
rrf, rrftrichloromonofluoromethane, trichloromonofluoromethaneistd, istdtribromomethane, tribromomethanebfb, bfbchloride, chlorideethyl, ethylmethyl, methylbutylbenzene, butylbenzenemethacrylate, methacrylatetetrachloroethylene, tetrachloroethyleneaquatek, aquateksurr, surrtert, tertlumin
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike