Effect of Split Ratio on USEPA Method 8260 Compounds

Applications |  | EST AnalyticalInstrumentation
GC/MSD, GC/SQ, Purge and Trap
Industries
Environmental
Manufacturer
EST Analytical, Restek, Agilent Technologies

Summary

Importance of the Topic


This application note addresses the critical challenge of analyzing volatile organic compounds (VOCs) in aqueous samples at trace levels according to USEPA Method 8260. Reliable monitoring of VOCs in environmental and industrial water matrices is essential for regulatory compliance, public health and quality control. Split injection strategies in capillary gas chromatography/mass spectrometry (GC/MS) enable moisture management and prevent column overload, directly impacting detection limits and data quality.

Objectives and Study Overview


This study evaluates the effect of three split ratios (40:1, 80:1 and 100:1) on analytical performance—linearity, compound response and method detection limits (MDLs)—for Method 8260 analytes. A nine‐point calibration (0.5–200 ppb) and replicate low‐level standards establish linear range and MDLs for each split condition.

Methodology and Instrumentation


  • Sample Preparation: Purge and trap extraction of 5 mL water samples spiked with internal standard.
  • Purge and Trap System: EST Analytical Evolution concentrator equipped with Vocarb 3000 trap and Moisture Reduction Trap (MoRT).
  • GC/MS Platform: Agilent 7890A GC with Restek Rxi‐624Sil MS column (30 m × 0.25 mm × 1.4 µm) coupled to 5975C inert XL MS detector.
  • Injection Conditions: Split inlet at 220 °C, head pressure ~12 psi, split ratios 40:1, 80:1, 100:1.
  • Oven Program: 45 °C (1 min) → 15 °C/min to 220 °C, total run time ~14 min.
  • MS Settings: Electron impact ionization, source 230 °C, quadrupole 150 °C, scan m/z 35–300.

Main Results and Discussion


Average performance across analytes:
  • Curve linearity (R²): 40:1 → 0.993; 80:1 → 0.992; 100:1 → 0.991.
  • Average response factor: ~1.02 (40:1), ~1.05 (80:1), ~1.00 (100:1).
  • Average MDL: 0.22 ppb (40:1), 0.38 ppb (80:1), 0.48 ppb (100:1).
The 40:1 split delivers the lowest MDLs due to higher analyte transfer to the column. Higher split ratios reduce moisture loading but sacrifice sensitivity. The built‐in moisture control valve of the Evolution concentrator minimizes water breakthrough, enabling low-split operation without compromising instrument integrity.

Benefits and Practical Applications


  • Optimized sensitivity: Lower split ratio yields sub‐ppb detection limits required by regulatory methods.
  • Robust moisture management: Dedicated trap and valve design reduce MS contamination.
  • High throughput: Short run times with capillary efficiency maintain resolution over a wide analyte range.

Future Trends and Applications


Advances may include automated split‐ratio optimization, dynamic split control during runs and integration with high‐capacity trap materials. Coupling purge‐and‐trap with two‐dimensional GC and high‐resolution MS will further enhance separation of complex VOC mixtures at ultra-trace levels.

Conclusion


All three split ratios meet USEPA 8260 requirements, but 40:1 provides superior detection limits without excessive moisture interference. The EST Evolution concentrator’s moisture control enables low-split injections, maximizing sensitivity while maintaining instrument health.

Reference


United States Environmental Protection Agency Method 8260B, Revision 2, December 1996.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Effect of Split Ratio on USEPA Method 8260 Compounds
         JSB is an authorised partner of   Effect of Split Ratio on USEPA Method 8260 Compounds #104 Application Note - Environmental Author Anne Jurek Applications Chemist EST Analytical Cincinnati, OH…
Key words
acetate, acetatetert, tertbutylbenzene, butylbenzenemethyl, methylsur, surether, etherbutyl, butylethyl, ethylchloride, chlorideamyl, amylmethacrylate, methacrylateest, estvinyl, vinylpurge, purgerile
A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260 Application Note Environmental Author Abstract Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Environmental laboratories are…
Key words
jurek, jurekanne, annecenturion, centuriondesorb, desorbacetate, acetateest, estsampling, samplingpurge, purgebake, bakebeer, beertert, tertflavor, flavorbutylbenzene, butylbenzeneextraction, extractionmicro
Automated Sampling of Methanol Extractions
Automated Sampling of Methanol Extractions
|Agilent Technologies|Applications
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Automated Sampling of Methanol Extractions Application Note Environmental Author Abstract Anne Jurek Applications Chemist EST Analytical Cincinnati, OH The United States Environmental Protection Agency (USEPA) Method 8260 is used…
Key words
tert, tertbutylbenzene, butylbenzenejurek, jurekether, etherpurge, purgesand, sandanne, annesur, surchloride, chloridebake, bakemethacrylate, methacrylatedesorb, desorbmethyl, methylrinse, rinsexylene
Optimizing 8260 Sample Throughput - Solutions for the Determination of Bromomethane by Purge and Trap
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK Optimizing 8260 Sample Throughput - Solutions for the Determination of Bromomethane by Purge and Trap Application Note Environmental Author Abstract Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Bromomethane…
Key words
bromomethane, bromomethanetrap, trappurge, purgebake, bakedesorb, desorbtemp, temprecovery, recoveryjurek, jurekacetate, acetateanne, annecustom, customest, esttert, tertvaried, variedrec'ry
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike