Application 210-00 Agilent Reformulated Fuel Analyzer
Technical notes | 2002 | Agilent TechnologiesInstrumentation
Accurate detection of oxygenates in gasoline and naphtha is critical for meeting modern fuel quality regulations, ensuring optimal combustion performance, and minimizing environmental impact. Methods compliant with ASTM D5599 and EN 1601 provide standardized protocols for quantifying alcohols and ethers at low concentration levels, supporting quality control in refinery operations and regulatory compliance.
This application note presents an analytical workflow for simultaneous quantification of key oxygenates in reformulated fuels. The study aims to demonstrate a rapid, robust chromatographic method using an Oxygenate Flame Ionization Detector (O-FID) capable of detecting compounds such as methanol, ethanol, MTBE, TAME, and other alcohol/ether species down to 0.06 wt%.
The method employs gas chromatography with O-FID detection under conditions aligned with ASTM D5599 and EN 1601. Samples are introduced into the Agilent Reformulated Fuel Analyzer, separated on a specialty column, and analyzed in a single 15-minute run. Retention times for 15 oxygenates are clearly resolved, enabling precise quantitation.
The method achieves baseline separation of water, lower alcohols (C1–C4), ethers (MTBE, ETBE, TAME, DIPE), and dimethoxyethane in under 15 minutes. Detection limits of 0.06 wt% ensure compliance with stringent fuel specifications. Chromatograms exhibit sharp, reproducible peaks with minimal matrix interferences. Optional method configurations extend compatibility with related ASTM protocols (D3606, D4815, D5580, D5769).
Advances in detector sensitivity and column technology may further reduce analysis time and improve limit of quantitation. Integration with automated sample handling and data processing will streamline workflows. Emerging alternative oxygenates and bio-derived solvents may require method adaptation and validation against updated fuel standards.
The described GC–O-FID method on the Agilent Reformulated Fuel Analyzer offers a streamlined, reliable solution for quantifying a broad range of oxygenates in gasoline and naphtha. Its compliance with ASTM D5599 and EN 1601 ensures applicability in diverse laboratory environments, supporting fuel quality monitoring and regulatory compliance.
GC
IndustriesEnergy & Chemicals
ManufacturerAgilent Technologies, Wasson-ECE Instrumentation
Summary
Significance of the topic
Accurate detection of oxygenates in gasoline and naphtha is critical for meeting modern fuel quality regulations, ensuring optimal combustion performance, and minimizing environmental impact. Methods compliant with ASTM D5599 and EN 1601 provide standardized protocols for quantifying alcohols and ethers at low concentration levels, supporting quality control in refinery operations and regulatory compliance.
Objectives and study overview
This application note presents an analytical workflow for simultaneous quantification of key oxygenates in reformulated fuels. The study aims to demonstrate a rapid, robust chromatographic method using an Oxygenate Flame Ionization Detector (O-FID) capable of detecting compounds such as methanol, ethanol, MTBE, TAME, and other alcohol/ether species down to 0.06 wt%.
Methodology and instrumentation
The method employs gas chromatography with O-FID detection under conditions aligned with ASTM D5599 and EN 1601. Samples are introduced into the Agilent Reformulated Fuel Analyzer, separated on a specialty column, and analyzed in a single 15-minute run. Retention times for 15 oxygenates are clearly resolved, enabling precise quantitation.
Used instrumentation
- Agilent Reformulated Fuel Analyzer platform
- Oxygenate Flame Ionization Detector (O-FID)
- Gas chromatograph with dedicated oxygenate separation column
Key results and discussion
The method achieves baseline separation of water, lower alcohols (C1–C4), ethers (MTBE, ETBE, TAME, DIPE), and dimethoxyethane in under 15 minutes. Detection limits of 0.06 wt% ensure compliance with stringent fuel specifications. Chromatograms exhibit sharp, reproducible peaks with minimal matrix interferences. Optional method configurations extend compatibility with related ASTM protocols (D3606, D4815, D5580, D5769).
Benefits and practical applications
- High throughput: complete analysis in ~15 minutes
- Low detection limits: reliable measurement of trace oxygenates
- Method flexibility: supports multiple ASTM and EN methods
- Quality assurance: robust performance for refinery QC and regulatory labs
Future trends and opportunities
Advances in detector sensitivity and column technology may further reduce analysis time and improve limit of quantitation. Integration with automated sample handling and data processing will streamline workflows. Emerging alternative oxygenates and bio-derived solvents may require method adaptation and validation against updated fuel standards.
Conclusion
The described GC–O-FID method on the Agilent Reformulated Fuel Analyzer offers a streamlined, reliable solution for quantifying a broad range of oxygenates in gasoline and naphtha. Its compliance with ASTM D5599 and EN 1601 ensures applicability in diverse laboratory environments, supporting fuel quality monitoring and regulatory compliance.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Application 250-00 Agilent Reformulated Fuel Analyzer
2002|Agilent Technologies|Technical notes
Application 250-00 Agilent Reformulated Fuel Analyzer Determination of Ethers and Alcohols in Gasoline Using ASTM D4815-99 Technical Overview Application Highlights • A Flame Ionization Detector (FID) to identify methyl tert-butylether (MTBE), ethyl tert-butylether (ETBE), tert-amylmethylether (TAME), diisopropyl ether (DIPE), methanol,…
Key words
butanol, butanolpropanoll, propanolltaa, taadipe, dipetame, tameetbe, etbeisobutanol, isobutanolmtbe, mtbeisopropanol, isopropanolethanol, ethanolreformulated, reformulatedsec, secmethanol, methanolethers, ethersgasoline
Application 257-00 Agilent Reformulated Fuel Analyzer
2002|Agilent Technologies|Technical notes
Application 257-00 Agilent Reformulated Fuel Analyzer ASTM D3606, ASTM D4815, and ASTM D5580 Technical Overview Application Highlights • ASTM D3606 Analysis of benzene and toluene Standard D3606 1 Benzene 2 MEK 3 Toluene in finished motor fuels using a Thermal…
Key words
butanol, butanolpropanoll, propanollxylene, xylenetaa, taadipe, dipetame, tameetbe, etbeisobutanol, isobutanolmtbe, mtbefinished, finishedisopropanol, isopropanolbackflush, backflushethylbenzene, ethylbenzenehydrocarbon, hydrocarbonethanol
Running ASTM Method D4815 for Determining Oxygenates in Gasoline on the Agilent 6820 GC
2004|Agilent Technologies|Applications
Running ASTM Method D4815 for Determining Oxygenates in Gasoline on the Agilent 6820 GC Application Gasoline Analysis and Environmental Protection Authors Introduction ChunXiao Wang Agilent Technologies Co., Ltd. (Shanghai) 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai 200131, P.R.C. Unleaded…
Key words
butanol, butanoldme, dmetcep, tcepfid, fidbackflush, backflushdipe, dipetert, tertpropanol, propanolcerity, ceritymtbe, mtberepeatability, repeatabilityvalve, valvetame, tamearea, areainlet
Oxygenates in Fuels –Analysis Solutions From Trace Levels to Ethanol Fuels
2010|Agilent Technologies|Presentations
Oxygenates in Fuels – Analysis Solutions From Trace Levels to Ethanol Fuels Page 1 James D. McCurry Senior Scientist Agilent Technologies Wilmington, DE USA October 13, 2010 Application Summary There is a need to measure trace oxygenates in: • Feedstocks…
Key words
oxygenates, oxygenatesethanol, ethanolgasoline, gasolineastm, astmoxygenate, oxygenateswitch, switchfluidic, fluidicfuel, fuelmethanol, methanolepc, epcether, ethercft, cftaux, auxtrace, tracedipe