Oxygenates in Fuels –Analysis Solutions From Trace Levels to Ethanol Fuels
Presentations | 2010 | Agilent TechnologiesInstrumentation
Accurate measurement of trace oxygenates in fuels and intermediates is essential for product quality control environmental compliance and engine performance optimization
This application summary outlines standardized gas chromatographic methods for quantifying oxygenates such as alcohols and ethers in feedstocks intermediates and finished gasoline It reviews three ASTM methods covering crude oils light hydrocarbons and reformulated gasoline and an ethanol fuel specification method
The analysis relies on multidimensional GC combining a nonpolar primary column (DB-1 or HP-5) with a polar oxygenate selective column (Oxy-PLOT or LowOX)
Valve and fluidic switch modes effectively isolate oxygenates from high boiling hydrocarbons allowing quantification down to low mg kg levels in complex matrices
The described solutions provide laboratories with turnkey methods to meet ASTM performance specifications for trace oxygenate analysis in feedstocks intermediates and finished fuels
Advances may include integration of mass spectral detection for unequivocal compound identification development of new stationary phases for even greater selectivity automation of sample handling and miniaturized GC systems for field deployable screening
Multidimensional GC methods with selective oxygenate columns and switching hardware offer robust sensitive and reproducible analysis of trace oxygenates across the fuel supply chain ensuring product integrity and regulatory compliance
GC
IndustriesEnergy & Chemicals
ManufacturerAgilent Technologies
Summary
Significance of the Topic
Accurate measurement of trace oxygenates in fuels and intermediates is essential for product quality control environmental compliance and engine performance optimization
Study Objectives and Overview
This application summary outlines standardized gas chromatographic methods for quantifying oxygenates such as alcohols and ethers in feedstocks intermediates and finished gasoline It reviews three ASTM methods covering crude oils light hydrocarbons and reformulated gasoline and an ethanol fuel specification method
Methodology and Instrumentation
The analysis relies on multidimensional GC combining a nonpolar primary column (DB-1 or HP-5) with a polar oxygenate selective column (Oxy-PLOT or LowOX)
- Two hardware configurations are described rotary valve switching and Capillary Flow Technology fluidic switch
- Typical GC conditions use helium carrier gas injector at 225 °C split ratio 10 1 oven program from 50 °C to 240 °C at 10 °C min and FID detection at 275 °C
- For ethanol/methanol in denatured fuel ethanol a single column PDMS phase with high split ratio and an oven program from 60 °C to 250 °C is used per ASTM D5501
Main Results and Discussion
Valve and fluidic switch modes effectively isolate oxygenates from high boiling hydrocarbons allowing quantification down to low mg kg levels in complex matrices
- In light C2–C5 streams more than 25 oxygenates including ethers aldehydes and alcohols are resolved with good peak shape
- In reformulated gasoline a 12 wt% ethanol matrix shifts early retention times but quantitative precision remains below 2 RSD for QA QC samples
- Methanol interference in denatured ethanol analysis can be managed by optimized integration and column selection achieving RSDs below 1.2 for methanol and 0.03 for ethanol
Benefits and Practical Applications
The described solutions provide laboratories with turnkey methods to meet ASTM performance specifications for trace oxygenate analysis in feedstocks intermediates and finished fuels
- Enhanced selectivity through two-dimensional separation reduces coelution of aromatics and high boiling species
- High throughput QA QC checks for fuel blending compliance and regulatory reporting
- Flexibility to analyze a wide concentration range from sub-mg kg to percent level ethanol in a single injection
Future Trends and Opportunities
Advances may include integration of mass spectral detection for unequivocal compound identification development of new stationary phases for even greater selectivity automation of sample handling and miniaturized GC systems for field deployable screening
Conclusion
Multidimensional GC methods with selective oxygenate columns and switching hardware offer robust sensitive and reproducible analysis of trace oxygenates across the fuel supply chain ensuring product integrity and regulatory compliance
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Fast Determination of Denatured Fuel Ethanol Purity by Two-Dimensional Gas Chromatography
2003|Agilent Technologies|Applications
Fast Determination of Denatured Fuel Ethanol Purity by Two-Dimensional Gas Chromatography Application Petrochemical Authors James D. McCurry Agilent Technologies 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract Two-dimensional gas chromatography not only provides improved resolution, it can also provide much…
Key words
ethanol, ethanolinnowax, innowaxgasoline, gasolinefuel, fueldenatured, denatureddeans, deanshydrocarbons, hydrocarbonsswitch, switchmethanol, methanolalcohol, alcoholepc, epccolumn, columnalcohols, alcoholsnatural, naturaldimensional
Analysis of Alcohols and Aromatics in Gasoline and Fuel Ethanol by a Two-Dimensional GC Method
2014|Agilent Technologies|Posters
Analysis of Alcohols and Aromatics in Gasoline and Fuel Ethanol by a Two-Dimensional GC Method Shannon Coleman Applications Engineer, Agilent Technologies Chemical Analysis Group Introduction HP-1 Column 15 m × 0.25 mm, 0.25 µm Front - FID pA 200 FID…
Key words
cut, cutswitch, switchfid, fidinnowax, innowaxdeans, deansoxygenates, oxygenatesgasoline, gasolineistd, istdcolumn, columndimensional, dimensionalcoelute, coeluteethanol, ethanolfluidic, fluidicmin, minprimary
A Fast, Unified Analysis of Key Components in Bio-Ethanol Fuels Using Heart-Cutting Multi-Dimensional Gas Chromatography
2012|Agilent Technologies|Posters
A Fast, Unified Analysis of Key Components in Bio-Ethanol Fuels Using Heart-Cutting Multi-Dimensional Gas Chromatography James D. McCurry, Agilent Technologies, Wilmington DE, USA Introduction Improvements Using Heart-Cutting Multi-Dimensional GC Capillary Flow Technology Deans Switch Ethanol fuels made from renewable biomass…
Key words
ethanol, ethanolmdgc, mdgcmethanol, methanolbenzene, benzenetoluene, toluenedenatured, denaturedfuel, fuelheart, heartcutting, cuttingcut, cutepc, epcfuels, fuelsflush, flushcommercial, commercialhydrocarbons
A Fast, Unified Analysis of Key Components in Bio-Ethanol Fuels Using Heart-Cutting Multi-Dimensional Gas Chromatography
|Agilent Technologies|Posters
A Fast, Unified Analysis of Key Components in Bio-Ethanol Fuels Using Heart-Cutting Multi-Dimensional Gas Chromatography (K.19) James D. McCurry, Agilent Technologies, Wilmington DE, USA Introduction Improvements Using Heart-Cutting Multi-Dimensional GC Capillary Flow Technology Deans Switch Ethanol fuels made from renewable…
Key words
ethanol, ethanolmdgc, mdgcmethanol, methanolbenzene, benzenetoluene, toluenedenatured, denaturedfuel, fuelheart, heartcutting, cuttingcut, cutepc, epcfuels, fuelsflush, flushcommercial, commercialhydrocarbons