Agilent JetClean self-cleaning ion source for GC/MS
Others | 2019 | Agilent TechnologiesInstrumentation
The performance and reliability of GC/MS analyses critically depend on maintaining a clean ion source. Manual cleaning routines are labor-intensive and cause significant instrument downtime, reducing laboratory productivity and increasing operational costs. Automated self-cleaning solutions can transform maintenance activities into productive measurement time, ensuring consistent data quality.
This application note presents the Agilent JetClean self-cleaning ion source designed for Agilent single and triple quadrupole GC/MS systems. It aims to demonstrate how a controlled hydrogen flow removes matrix deposits without disassembling the source, thereby increasing instrument uptime, preserving analytical performance, and streamlining laboratory workflows.
The JetClean technology integrates with Agilent 5977A/B, 7000B/C/D, 7010A/B and 5977B Intuvo GC/MS instruments. A precisely regulated hydrogen purge is applied automatically during or between runs to prevent buildup of contaminants such as Rhodamine dye or polyaromatic hydrocarbons.
A test with a heavily contaminated lens marked with Rhodamine 6 showed complete removal of deposits in a single cleaning cycle, restoring original sensitivity.
Users reported eliminating routine manual source removal for periods up to nine months, yielding uninterrupted data acquisition and stable signal intensities.
The self-cleaning source offers:
Cost-saving calculators show significant annual savings in operator wages and lost sample revenues when compared to traditional cleaning protocols.
As analytical demands grow, automated maintenance solutions like JetClean will be integral to high-throughput laboratories. Future developments may include adaptive cleaning schedules driven by real-time contamination monitoring, integration with laboratory information management systems (LIMS), and expansion to other mass spectrometry platforms.
The Agilent JetClean self-cleaning ion source offers a transformative approach to GC/MS maintenance by automating contaminant removal, preserving data quality, and maximizing instrument uptime. Its adoption can streamline workflows, reduce operating costs, and support advanced analytical applications without manual intervention.
GC/MSD
IndustriesManufacturerAgilent Technologies
Summary
Significance of the Topic
The performance and reliability of GC/MS analyses critically depend on maintaining a clean ion source. Manual cleaning routines are labor-intensive and cause significant instrument downtime, reducing laboratory productivity and increasing operational costs. Automated self-cleaning solutions can transform maintenance activities into productive measurement time, ensuring consistent data quality.
Objectives and Overview
This application note presents the Agilent JetClean self-cleaning ion source designed for Agilent single and triple quadrupole GC/MS systems. It aims to demonstrate how a controlled hydrogen flow removes matrix deposits without disassembling the source, thereby increasing instrument uptime, preserving analytical performance, and streamlining laboratory workflows.
Methodology and Used Instrumentation
The JetClean technology integrates with Agilent 5977A/B, 7000B/C/D, 7010A/B and 5977B Intuvo GC/MS instruments. A precisely regulated hydrogen purge is applied automatically during or between runs to prevent buildup of contaminants such as Rhodamine dye or polyaromatic hydrocarbons.
Main Results and Discussion
A test with a heavily contaminated lens marked with Rhodamine 6 showed complete removal of deposits in a single cleaning cycle, restoring original sensitivity.
- Polycyclic aromatic hydrocarbons (PAHs) in palm oil extracts were measured over 200 injections across five days after a simple toluene extraction.
- Linearity was maintained over a 1–100 ng/mL calibration range with R2 > 0.997 on both Day 1 and Day 5.
- Precision (area %RSD) remained below 5% for key PAH analytes throughout the study, confirming robust, reproducible performance.
Users reported eliminating routine manual source removal for periods up to nine months, yielding uninterrupted data acquisition and stable signal intensities.
Benefits and Practical Applications
The self-cleaning source offers:
- Up to 90% reduction in manual cleaning events, dramatically cutting labor expenses and revenue loss from downtime.
- Consistent run-to-run reproducibility and lower detection limits in demanding applications such as environmental PAH monitoring and metabolomics.
- Enhanced convenience with the ability to clean during acquisition or idle periods without dismantling the source.
Cost-saving calculators show significant annual savings in operator wages and lost sample revenues when compared to traditional cleaning protocols.
Future Trends and Potential Applications
As analytical demands grow, automated maintenance solutions like JetClean will be integral to high-throughput laboratories. Future developments may include adaptive cleaning schedules driven by real-time contamination monitoring, integration with laboratory information management systems (LIMS), and expansion to other mass spectrometry platforms.
Conclusion
The Agilent JetClean self-cleaning ion source offers a transformative approach to GC/MS maintenance by automating contaminant removal, preserving data quality, and maximizing instrument uptime. Its adoption can streamline workflows, reduce operating costs, and support advanced analytical applications without manual intervention.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Agilent JetClean self-cleaning ion source
2019|Agilent Technologies|Brochures and specifications
Less Cleaning, More GC/MS Results Agilent JetClean self-cleaning ion source Keep Your Ion Source Clean— Automatically Matrix deposits inevitably build up during routine analysis. In the past, you would have to remove the ion source, scrub the lens and other…
Key words
jetclean, jetcleancleaning, cleaningcleanings, cleaningsper, persource, sourceyearly, yearlyrevenue, revenueyear, yearself, selfwage, wagesavings, savingsbenzo, benzowages, wagesion, ionoctober
How Much Could Your Lab Save by Cutting Cleaning Time up to 90%?
|Agilent Technologies|Others
How Much Could Your Lab Save by Cutting Cleaning Time up to 90%? See for yourself Plug in the numbers that are relevant to your lab and see the actual impact the Select your currency Agilent JetClean self-cleaning ion source…
Key words
jetclean, jetcleancleaning, cleaningyearly, yearlyrevenue, revenuecleanings, cleaningsper, perwage, wagewages, wagesdowntime, downtimesavings, savingscost, costevent, eventloss, losssee, seeyour
Significant Robustness Improvements of PAHs Analysis in Palm Oil Using the JetClean Self-Cleaning Ion Source in a GC/MS/MS System
2016|Agilent Technologies|Posters
Significant Robustness Improvements of PAHs Analysis in Palm Oil Using the JetClean Self-Cleaning Ion Source in a GC/MS/MS System ASMS 2016 Poster WP 209 Michel Lesieur1; Elizabeth Almasi2; Terry Sheehan2; 1Agilent Technologies, Les Ulis, France; 2Agilent Technologies, Santa Clara, CA…
Key words
cleaning, cleaningpalm, palmjetclean, jetcleantoluene, tolueneblank, blankpahs, pahsself, selfsource, sourceoil, oilremained, remainedcounts, countsion, ionoïl, oïlprudently, prudentlyimprecisions
PAHS ANALYSIS IN PALM OIL SIGNIFICANT ROBUSTNESS IMPROVEMENTS
2016|Agilent Technologies|Applications
PAHS ANALYSIS IN PALM OIL SIGNIFICANT ROBUSTNESS IMPROVEMENTS Technology Advantage: Agilent JetClean Self-Cleaning Ion Source in a GC/MS/MS System Introduction Regulated limits for polyaromatic hydrocarbons (PAHs) in food have been steadily lowered as the result of increased awareness about their…
Key words
jetclean, jetcleantoluene, toluenecleaning, cleaningblank, blankpalm, palmjul, julaug, augfeb, febapr, aprmar, marsource, sourcenov, novjun, jundec, decpahs