GC Monitoring of Chlorinated Pesticides in Hazardous Waste Sites
Applications | 1994 | MerckInstrumentation
Monitoring chlorinated pesticides at hazardous waste sites is essential to assess contamination levels, protect environmental and human health, and ensure compliance with regulatory standards such as the US EPA Contract Laboratory Program (CLP). Analytical methods that deliver high resolution, inertness, and linearity enable accurate detection of trace-level organochlorine compounds in complex matrices.
This work examines the performance of SPB-608 fused silica wide-bore capillary columns for gas chromatographic monitoring of organochlorine pesticides and PCBs in accordance with the US EPA CLP March 1990 Statement of Work. The study aims to verify baseline resolution of target analytes, evaluate column inertness through decomposition measurements, and confirm linearity across defined concentration ranges.
The evaluation employed direct injection of CLP-defined standard mixtures into a gas chromatograph equipped with an electron capture detector. GC conditions included a 30m x 0.53mm ID SPB-608 column with 0.5µm film thickness, helium carrier gas at 5mL/min, an oven program ramp from 150°C to 275°C at 6°C/min, and ECD at 300°C.
All seven pesticide analytes in the resolution check mix achieved baseline separation, surpassing the CLP resolution requirement of >60% peak height of the shorter peak. Combined decomposition of endrin and 4,4'-DDT measured 8.86%, well below the 30% threshold. Linearity assessment across three concentration levels yielded a mean relative standard deviation of calibration factors at 9.49%, within CLP limits (<20% for most analytes, <30% for select cases). Absolute retention times remained within the prescribed windows for all compounds.
The demonstrated SPB-608 column performance ensures reliable detection and quantification of chlorinated pesticides at hazardous waste sites, supporting environmental monitoring, regulatory compliance, and risk assessment. Its inert surface and consistent linearity make it suitable for routine analysis in contract laboratories and industrial settings.
Emerging developments may include higher inertness column coatings, faster temperature-programmed analyses, coupling with mass spectrometric detection for enhanced specificity, micro- or nano-scale column formats for increased throughput, and integration with automated sampling systems for real-time monitoring.
SPB-608 fused silica capillary columns meet and exceed US EPA CLP requirements for resolution, inertness, and linearity in the monitoring of chlorinated pesticides and PCBs. Their robust performance and compatibility with standard GC-ECD protocols make them a reliable choice for environmental and hazardous waste analysis.
GC, GC columns, Consumables
IndustriesEnvironmental
ManufacturerMerck
Summary
Importance of the Topic
Monitoring chlorinated pesticides at hazardous waste sites is essential to assess contamination levels, protect environmental and human health, and ensure compliance with regulatory standards such as the US EPA Contract Laboratory Program (CLP). Analytical methods that deliver high resolution, inertness, and linearity enable accurate detection of trace-level organochlorine compounds in complex matrices.
Objectives and Study Overview
This work examines the performance of SPB-608 fused silica wide-bore capillary columns for gas chromatographic monitoring of organochlorine pesticides and PCBs in accordance with the US EPA CLP March 1990 Statement of Work. The study aims to verify baseline resolution of target analytes, evaluate column inertness through decomposition measurements, and confirm linearity across defined concentration ranges.
Methodology
The evaluation employed direct injection of CLP-defined standard mixtures into a gas chromatograph equipped with an electron capture detector. GC conditions included a 30m x 0.53mm ID SPB-608 column with 0.5µm film thickness, helium carrier gas at 5mL/min, an oven program ramp from 150°C to 275°C at 6°C/min, and ECD at 300°C.
Instrumentation Used
- Gas chromatograph with electron capture detector (ECD)
- SPB-608 capillary column, 30m x 0.53mm ID, 0.5µm film
- Helium carrier gas, 5mL/min
- Oven protocol: 150°C (0.5 min) to 275°C at 6°C/min
- Injection volume: 2µL direct inject at 200°C
Main Results and Discussion
All seven pesticide analytes in the resolution check mix achieved baseline separation, surpassing the CLP resolution requirement of >60% peak height of the shorter peak. Combined decomposition of endrin and 4,4'-DDT measured 8.86%, well below the 30% threshold. Linearity assessment across three concentration levels yielded a mean relative standard deviation of calibration factors at 9.49%, within CLP limits (<20% for most analytes, <30% for select cases). Absolute retention times remained within the prescribed windows for all compounds.
Benefits and Practical Applications
The demonstrated SPB-608 column performance ensures reliable detection and quantification of chlorinated pesticides at hazardous waste sites, supporting environmental monitoring, regulatory compliance, and risk assessment. Its inert surface and consistent linearity make it suitable for routine analysis in contract laboratories and industrial settings.
Future Trends and Applications
Emerging developments may include higher inertness column coatings, faster temperature-programmed analyses, coupling with mass spectrometric detection for enhanced specificity, micro- or nano-scale column formats for increased throughput, and integration with automated sampling systems for real-time monitoring.
Conclusion
SPB-608 fused silica capillary columns meet and exceed US EPA CLP requirements for resolution, inertness, and linearity in the monitoring of chlorinated pesticides and PCBs. Their robust performance and compatibility with standard GC-ECD protocols make them a reliable choice for environmental and hazardous waste analysis.
References
- US EPA Contract Laboratory Program Statement of Work for Organics Analysis, Multi-Media, Multi-Concentration, Document 0LM01.0, including revisions 0LM01.1 and 0LM01.1.1.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
GC Monitoring of Chlorinated Pesticides in Hazardous Waste Sites
1999|Merck|Applications
TheReporter Reprinted from Volume 14, No. 5, 1995 T295045 © 1999 Sigma-Aldrich Co. For more information, or current prices, contact your nearest Supelco subsidiary listed below. To obtain further contact information, visit our website (www.sigma-aldrich.com), see the Supelco catalog, or…
Key words
endrin, endrinbhc, bhcendosulfan, endosulfansupelco, supelcochlordane, chlordaneheptachlor, heptachlorketone, ketonepesticides, pesticidesfigure, figurealdehyde, aldehydeepoxide, epoxidetetrachloro, tetrachloroinertness, inertnessclp, clpdecachlorobiphenyl
Bulletin 841C Chlorinated Pesticides Analysis by Capillary GC: Alternative Approaches We manufacture and test SPB-608 capillary columns specifically for analyses of chlorinated pesticides. 30m x 0.25mm ID columns can be used in capillary GC systems; 15m x 0.53mm ID columns…
Key words
bhc, bhcendosulfan, endosulfanendrin, endrinheptachlor, heptachlorpesticides, pesticidesepoxide, epoxidedieldrin, dieldrinaldrin, aldrinaldehyde, aldehydesupelco, supelcosulfate, sulfateextraction, extractionchlordane, chlordaneenvi, envimethoxychlor
Fast Screening for Chlorinated Pesticides by Solid Phase Microextraction/Capillary GC
1999|Merck|Applications
Application Note 58 Fast Screening for Chlorinated Pesticides by Solid Phase Microextraction/Capillary GC Fast analysis of environmental samples increases throughput in the lab and is highly desirable for “quick turnaround” screening methods used in the field or prior to GC/MS…
Key words
endrin, endrinendosulfan, endosulfanspme, spmechlorinated, chlorinatedbhc, bhcchlordane, chlordaneheptachlor, heptachlorpesticides, pesticidesliquid, liquidextraction, extractionmethoxychlor, methoxychloraldehyde, aldehydeepoxide, epoxidemicroextraction, microextractiondieldrin
Fast Screening for Chlorinated Pesticides, Using Solid Phase MicroExtraction (SPME) and Capillary GC
1999|Merck|Applications
TheReporter Reprinted from Volume 14, No. 4, 1995 T295024 © 1999 Sigma-Aldrich Co. For more information, or current prices, contact your nearest Supelco subsidiary listed below. To obtain further contact information, visit our website (www.sigma-aldrich.com), see the Supelco catalog, or…
Key words
endrin, endrinendosulfan, endosulfanspme, spmebhc, bhcchlordane, chlordaneheptachlor, heptachlorsupelco, supelcocapillary, capillaryphase, phasemethoxychlor, methoxychlorpesticides, pesticidesaldehyde, aldehydeepoxide, epoxidechlorinated, chlorinateddieldrin