Shortening Cycle Times for Analyzing Volatile Compounds
Applications | 2011 | ShimadzuInstrumentation
Volatile organic compounds in water are key indicators of environmental pollution and industrial contamination. Rapid and reliable detection of these analytes supports regulatory compliance, environmental monitoring, and public health protection. Innovations that shorten analysis time while maintaining sensitivity and reproducibility can greatly enhance laboratory throughput and cost-effectiveness.
This study aimed to reduce total cycle time for headspace gas chromatography–mass spectrometry (GC-MS) analysis of 28 target VOCs in water. By leveraging a fast-cooling GC oven and optimized headspace sampling, the work evaluates whether sub-six-minute measurement per sample is achievable without compromising analytical performance.
Samples were heated and equilibrated using an automated headspace sampler operated in constant mode. VOCs were separated on a narrow-bore Rtx-624 capillary column and detected by mass spectrometry in selected-ion monitoring (SIM) mode.
All 28 VOCs, including three internal standards, were detected within a GC-MS analysis window of under 6 minutes. Column cooling to starting temperature required 3.8 minutes, including a 1 minute equilibration. Typical limits of quantitation were 0.1 µg/L (5 µg/L for 1,4-dioxane) with seven-fold replicate measurements yielding relative standard deviations below 7 % for all target compounds. Chromatograms showed sharp peaks and baseline stability, demonstrating that accelerated temperature ramps and fast headspace cycles do not degrade chromatographic resolution or sensitivity.
The fast-GC/MS approach significantly increases sample throughput to approximately six analyses per hour. This efficiency benefits environmental testing laboratories, drinking water quality monitoring, and industrial process control by reducing turnaround time and instrument idle periods.
Continued development of rapid-cooling oven technologies and multiplexed headspace sampling could further boost throughput. Integration with automated data processing and advanced mass spectrometry detectors promises enhanced sensitivity for trace-level monitoring. Expansion to other volatile analyte classes and coupling with field-deployable GC-MS systems may broaden applications in on-site environmental diagnostics.
The combination of a double-shot cooling GC oven and optimized headspace sampling enables sub-six-minute analysis cycles for a broad range of VOCs in water without sacrificing sensitivity or repeatability. This method offers a robust solution for high-throughput environmental and industrial analyses.
Shimadzu Corporation. Shortening Cycle Times for Analyzing Volatile Compounds by Headspace GC/MS, LAAN-J-MS-E034, First Edition, November 2011.
GC/MSD, HeadSpace, GC/SQ
IndustriesEnvironmental
ManufacturerShimadzu, PerkinElmer
Summary
Significance of the Topic
Volatile organic compounds in water are key indicators of environmental pollution and industrial contamination. Rapid and reliable detection of these analytes supports regulatory compliance, environmental monitoring, and public health protection. Innovations that shorten analysis time while maintaining sensitivity and reproducibility can greatly enhance laboratory throughput and cost-effectiveness.
Objectives and Study Overview
This study aimed to reduce total cycle time for headspace gas chromatography–mass spectrometry (GC-MS) analysis of 28 target VOCs in water. By leveraging a fast-cooling GC oven and optimized headspace sampling, the work evaluates whether sub-six-minute measurement per sample is achievable without compromising analytical performance.
Methodology and Instrumentation
Samples were heated and equilibrated using an automated headspace sampler operated in constant mode. VOCs were separated on a narrow-bore Rtx-624 capillary column and detected by mass spectrometry in selected-ion monitoring (SIM) mode.
- Headspace sampler: TurboMatrix HS with overlapped heating for up to 12 vials
- Gas chromatograph–mass spectrometer: Shimadzu GCMS-QP2010 Ultra
- Column: Rtx-624, 20 m length × 0.18 mm I.D. × 1.0 µm film
- Oven program: 50 °C hold for 0.5 min, ramp at 35 °C/min to 200 °C, hold 1 min
- Carrier gas: Helium
- Transfer line: 150 °C; Interface: 230 °C; Ion source: 200 °C
- SIM event time: 0.2 s
Main Results and Discussion
All 28 VOCs, including three internal standards, were detected within a GC-MS analysis window of under 6 minutes. Column cooling to starting temperature required 3.8 minutes, including a 1 minute equilibration. Typical limits of quantitation were 0.1 µg/L (5 µg/L for 1,4-dioxane) with seven-fold replicate measurements yielding relative standard deviations below 7 % for all target compounds. Chromatograms showed sharp peaks and baseline stability, demonstrating that accelerated temperature ramps and fast headspace cycles do not degrade chromatographic resolution or sensitivity.
Benefits and Practical Applications
The fast-GC/MS approach significantly increases sample throughput to approximately six analyses per hour. This efficiency benefits environmental testing laboratories, drinking water quality monitoring, and industrial process control by reducing turnaround time and instrument idle periods.
Future Trends and Potential Applications
Continued development of rapid-cooling oven technologies and multiplexed headspace sampling could further boost throughput. Integration with automated data processing and advanced mass spectrometry detectors promises enhanced sensitivity for trace-level monitoring. Expansion to other volatile analyte classes and coupling with field-deployable GC-MS systems may broaden applications in on-site environmental diagnostics.
Conclusion
The combination of a double-shot cooling GC oven and optimized headspace sampling enables sub-six-minute analysis cycles for a broad range of VOCs in water without sacrificing sensitivity or repeatability. This method offers a robust solution for high-throughput environmental and industrial analyses.
References
Shimadzu Corporation. Shortening Cycle Times for Analyzing Volatile Compounds by Headspace GC/MS, LAAN-J-MS-E034, First Edition, November 2011.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Fast Analysis of Volatile Organic Compounds (VOCs) in Water Using Headspace-GC/MS
2015|Shimadzu|Applications
LAAN-J-MS-E108 GC-MS Gas Chromatograph Mass Spectrometer 108 Fast Analysis of Volatile Organic Compounds (VOCs) in Water Using Headspace-GC/MS The headspace-GC/MS method is utilized to analyze volatile organic compounds (VOCs) in drinking water and environmental water. VOCs are easy to evaporate…
Key words
vocs, vocscoefficient, coefficientcorrelation, correlationheadspace, headspacebromofluorobenzene, bromofluorobenzenevolatile, volatilefluorobenzene, fluorobenzenename, nametetrachloroethylene, tetrachloroethylenetrichloroethylene, trichloroethyleneorganic, organicrsd, rsddibromochloromethane, dibromochloromethanebromodichloromethane, bromodichloromethanebromoform
Analysis of VOCs in Water Using Headspace-GC/MS
2013|Shimadzu|Applications
LAAN-J-MS-E076 GC-MS Gas Chromatograph Mass Spectrometer Analysis of VOCs in Water Using Headspace-GC/MS 76 Introduction Volatile Organic Compounds (VOCs) in water are regulated by environmental water quality laws or standards because of health hazard factors. Purge and trap methods and…
Key words
equilibrating, equilibratingtime, timevocs, vocsload, loadloop, loopunit, unitwater, waterlaws, lawspressurizing, pressurizingvial, vialmode, modecompounds, compoundsshaker, shakername, nametetrachloroethylene
Volatile organic compounds
|GL Sciences|Applications
InertSearch for GC TM InertCap® Applications Volatile organic compounds Data No. GA239-0591 19+20 17 14 8+9 21 10 23 6 7 1 3 2 4 5 11 12 13 Time (min) 16 15 18 22 InertSearch for GC TM InertCap®…
Key words
キシレン, キシレンテトラクロロエチレン, テトラクロロエチレンジブロモクロロメタン, ジブロモクロロメタンブロモジクロロメタン, ブロモジクロロメタンブロモホルム, ブロモホルムトリクロロエチレン, トリクロロエチレンinertcap, inertcapジクロロメタン, ジクロロメタンクロロホルム, クロロホルムtrichloroethylene, trichloroethyleneトルエン, トルエンbromodichloromethane, bromodichloromethanedichloride, dichloridedichloromethane, dichloromethanechloroform
Introduction of a New, High-quality, Cost-effcient Headspace GC Autosampler, the HS-10
2016|Shimadzu|Posters
PO-CON1615E Introduction of a New, High-quality, Cost-efficient Headspace GC Autosampler, the HS-10 Pittcon 2016 2270-1 Clifford M. Taylor, Marketing Manager, Gas Chromatography Products; Marty Smith, GC Project Engineer; Zhuangzhi “Max” Wang, Ph.D. GC Senior Product Specialist; Shimadzu Scientific Instruments, Columbia,…
Key words
analysis, analysisline, linevial, vialisoamyl, isoamyltemp, tempmin, minahead, aheadintake, intaketransfer, transferbeer, beeroverlap, overlapvocs, vocscycles, cyclesmixing, mixingblood