An Alternative to Direct Probe: Pyroprobe® Coupled to High Resolution Time-of-Flight Mass Spectrometry
Posters | 2015 | LECOInstrumentation
The rapid analysis of nonvolatile or thermally labile samples is essential in many fields of analytical chemistry, especially when traditional gas chromatography methods are not viable.
Pyroprobe coupling with high‐resolution time‐of‐flight mass spectrometry (Py‐HRT) offers enhanced flexibility for direct thermal desorption, pyrolysis, and preliminary chromatographic separations in a single workflow.
This study evaluates the performance and versatility of a CDS Analytical Pyroprobe® 5200 coupled to a LECO Pegasus® GC‐HRT versus a Scientific Instrument Services (SIS) Direct Exposure Probe (DEP) with time‐of‐flight MS.
Applications include characterization of non‐GC amenable liquids and solids such as surfactants, saccharides, drugs of abuse, organometallic compounds, and polymeric flame retardants.
Pyroprobe-HRT is not a direct replacement for traditional direct probes but adds valuable functionality through combined thermal desorption, pyrolysis, and optional GC separation.
The system’s high mass accuracy, automated deconvolution, and robust design make it a versatile tool for non-GC amenable sample screening across diverse applications.
GC/MSD, GC/HRMS, Pyrolysis, GC/TOF
IndustriesFood & Agriculture
ManufacturerCDS Analytical, LECO
Summary
Importance of the Topic
The rapid analysis of nonvolatile or thermally labile samples is essential in many fields of analytical chemistry, especially when traditional gas chromatography methods are not viable.
Pyroprobe coupling with high‐resolution time‐of‐flight mass spectrometry (Py‐HRT) offers enhanced flexibility for direct thermal desorption, pyrolysis, and preliminary chromatographic separations in a single workflow.
Objectives and Study Overview
This study evaluates the performance and versatility of a CDS Analytical Pyroprobe® 5200 coupled to a LECO Pegasus® GC‐HRT versus a Scientific Instrument Services (SIS) Direct Exposure Probe (DEP) with time‐of‐flight MS.
Applications include characterization of non‐GC amenable liquids and solids such as surfactants, saccharides, drugs of abuse, organometallic compounds, and polymeric flame retardants.
Methodology and Instrumentation
- Direct Exposure Probe/TOFMS (DEP/TOFMS): SIS DEP PC‐3, 0–1.5 A filament ramp (1 A/min), ion source at 250 °C, mass range 35–1000 m/z, acquisition 3 spectra/s.
- Pyroprobe‐HRT (Py‐HRT): CDS Pyroprobe 5200 interfaced to LECO Pegasus GC‐HRT, interface 100–300 °C, pyrolysis 300–800 °C at 100 °C/min, inlet and transfer lines at 300 °C, mass range 30–1000 m/z, EI and CI modes, acquisition 3 spectra/s.
- Analytical workflow allowed direct thermal desorption, pyrolysis, and optional chromatographic separation via the GC inlet.
Main Results and Discussion
- Surfactants (Triton X-114): DEP/TOFMS provided clear oligomer distributions via extracted ion chromatograms, analogous to NIST reference spectra.
- Food and Flavor Compounds: Fructose showed good library match (864/1000) by DEP but was not detectable by Py‐HRT; caffeine yielded excellent scores by both methods, with Py‐HRT achieving sub-1 ppm mass accuracy and a similarity of 926/1000.
- Drugs of Abuse (AM-2201, ZX-1): DEP/TOFMS deconvolution delivered reliable structural ions; Py‐HRT CI data provided precise adduct ions (0.16 ppm accuracy) even when EI failed to show molecular ions.
- Polymeric Flame Retardants in Foam: A Cl-H mass defect plot from Py‐HRT revealed brominated compounds (TBB, TBPH) and organophosphates, indicating use of commercial flame retardant mixtures.
Benefits and Practical Applications
- Pyroprobe coupling extends direct analysis to include pyrolysis and fractionation without sample transfer steps.
- High-resolution time-of-flight MS minimizes spectral skew and enhances mass accuracy for confident identification.
- Automated deconvolution streamlines peak finding in complex mixtures.
- Low-maintenance ion source design suits routine QA/QC and industrial laboratories.
Future Trends and Opportunities
- Integration of real-time data processing and machine learning for advanced deconvolution and compound identification.
- Expansion into ambient pyrolysis sources and in-situ analysis for field applications.
- Development of standardized thermal profiles for targeted classes of compounds.
- Enhanced coupling with tandem MS and hybrid analyzers for structural elucidation.
Conclusion
Pyroprobe-HRT is not a direct replacement for traditional direct probes but adds valuable functionality through combined thermal desorption, pyrolysis, and optional GC separation.
The system’s high mass accuracy, automated deconvolution, and robust design make it a versatile tool for non-GC amenable sample screening across diverse applications.
Instrumentation Used
- Scientific Instrument Services Direct Exposure Probe PC-3 with TOFMS
- CDS Analytical Pyroprobe® 5200
- LECO Pegasus® HT GC-HRT Mass Spectrometer
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Direct Exposure Probe Analysis on the Pegasus HT: An Application Compendium
2015|LECO|Applications
Direct Exposure Probe Analysis on the ® Pegasus HT: An Application Compendium LECO Corporation; Saint Joseph, Michigan USA Key Words: Direct Exposure Probe (DEP), Pegasus HT, Surfactant, Drugs of Abuse, Organometallic, Food and Flavor 4 3 5 n = 7.5…
Key words
filament, filamentpegasus, pegasusleco, lecodelivering, deliveringmass, massdep, depexposure, exposuresynthetic, syntheticprobe, probetic, ticscience, sciencelife, lifesource, sourceright, rightdirect
Py-GC HR-TOF of Circuit Board Polymer
|CDS Analytical|Applications
172 olutions APPLICATIONS INFORMATION USING ADVANCED SAMPLE HANDLING TECHNOLOGY Py-GC HR-TOF of Circuit Board Polymer Considering pyrolysis creates fragments which are typically not seen in traditional GC/MS(solvent) injections, interpretation may be tricky; some compounds are not found in traditional mass…
Key words
hrt, hrtcircuit, circuitboard, boardpolymer, polymertetrabromobisphenol, tetrabromobisphenolmass, masstof, toftricky, trickytraditional, traditionalpyrogram, pyrogramretardant, retardantcontacting, contactingolutions, olutionsmeasurements, measurementspolycarbonate
Characterization of a Cannabinomimetic Compound in Synthetic Cannabis by Gas Chromatography-High Resolution Time of Flight Mass Spectrometry (GC-HRT)
2011|Agilent Technologies|Applications
Characterization of a Cannabinomimetic Compound in Synthetic Cannabis by Gas Chromatography-High Resolution Time of Flight Mass Spectrometry (GC-HRT) LECO Corporation; Saint Joseph, Michigan USA Keywords: Gas Chromatography; High Resolution Mass Spectrometry; Forensics; Synthetic Cannabis 1. Introduction Synthetic marijuana-type products have…
Key words
hrt, hrtleco, lecopegasus, pegasusunknown, unknownmass, masstofms, tofmsincense, incensesearch, searchtrutof, trutofcommercially, commerciallyspectral, spectralpresent, presentffp, ffpacetate, acetatemethane
Evaluation of HR-TOFMS with the Multi-Mode Ion Source and the Automated Direct Inlet Probe
2021|LECO|Posters
Evaluation of HR-TOFMS with the Multi-Mode Ion Source and the Automated Direct Inlet Probe George Tikhonov, Viatcheslav Artaev Overview • Direct Inlet Probe (DIP) is an efficient sample introduction tool for the express analysis of pure samples and for the…
Key words
dip, dipmms, mmsionization, ionizationmass, massmodes, modessource, sourceanalyzer, analyzerffp, ffpion, ionoperate, operatesim, simmode, modefilament, filamentinserted, insertedmulti