Essential Oils—Area Percent Calculations Using Deconvoluted Total Ion Chromatogram
Applications | 2008 | LECOInstrumentation
Accurate quantification of essential oil components is critical in flavor, fragrance and quality control laboratories. Complex mixtures and coeluting compounds often compromise total ion chromatogram (TIC)–based area percent calculations, leading to biased results and protracted run times. Spectral deconvolution in time-of-flight mass spectrometry (TOFMS) offers a robust solution by exploiting spectral continuity to resolve overlapping peaks.
This application note demonstrates the use of LECO ChromaTOF software with Pegasus II GC-TOFMS to calculate precise area percentages for closely coeluting components in a geranium Bourbon essential oil (Bourdonol). The aim is to compare conventional TIC integration with deconvoluted TIC (DTIC) and highlight improvements in accuracy and throughput.
Experimental conditions included an Agilent 6890 GC with DB-5 column (10 m × 0.18 mm, 0.18 µm film), split 100:1, oven ramp from 40 °C to 320 °C at 40 °C/min, constant flow at 1.5 mL/min, 200 °C injector. The Pegasus II GC-TOFMS acquired 45–450 amu at 50 spectra/s with 200 °C ion source.
Conventional TIC integration failed to resolve overlapping peaks, misassigning areas where no clear valleys existed. Deconvolution using DTIC identified additional components hidden in the TIC and set correct integration limits even for non-Gaussian shapes. For example, five coeluting peaks between 214–217 s were resolved versus four in the TIC. Closely coeluting peaks at 160–162 s with large concentration differences were correctly quantified only by DTIC.
Advances in high-speed acquisition and machine-learning deconvolution promise even faster and more accurate profiling of complex matrices. Integration with cloud-based spectral libraries and real-time feedback could further accelerate method development and industrial monitoring.
GC-TOFMS with spectral deconvolution enables precise area percent measurements for essential oil analysis despite severe coelutions. The approach reduces chromatographic run times, improves data reliability and boosts laboratory productivity.
GC/MSD, GC/TOF
IndustriesFood & Agriculture
ManufacturerAgilent Technologies, LECO
Summary
Significance of Topic
Accurate quantification of essential oil components is critical in flavor, fragrance and quality control laboratories. Complex mixtures and coeluting compounds often compromise total ion chromatogram (TIC)–based area percent calculations, leading to biased results and protracted run times. Spectral deconvolution in time-of-flight mass spectrometry (TOFMS) offers a robust solution by exploiting spectral continuity to resolve overlapping peaks.
Study Objectives and Overview
This application note demonstrates the use of LECO ChromaTOF software with Pegasus II GC-TOFMS to calculate precise area percentages for closely coeluting components in a geranium Bourbon essential oil (Bourdonol). The aim is to compare conventional TIC integration with deconvoluted TIC (DTIC) and highlight improvements in accuracy and throughput.
Methodology and Instrumentation
Experimental conditions included an Agilent 6890 GC with DB-5 column (10 m × 0.18 mm, 0.18 µm film), split 100:1, oven ramp from 40 °C to 320 °C at 40 °C/min, constant flow at 1.5 mL/min, 200 °C injector. The Pegasus II GC-TOFMS acquired 45–450 amu at 50 spectra/s with 200 °C ion source.
Results and Discussion
Conventional TIC integration failed to resolve overlapping peaks, misassigning areas where no clear valleys existed. Deconvolution using DTIC identified additional components hidden in the TIC and set correct integration limits even for non-Gaussian shapes. For example, five coeluting peaks between 214–217 s were resolved versus four in the TIC. Closely coeluting peaks at 160–162 s with large concentration differences were correctly quantified only by DTIC.
Benefits and Practical Applications
- Reliable area percent calculations for coeluted compounds without extended run times
- Automated peak finding, spectral deconvolution and library search streamline workflows
- Enhanced sample throughput and data quality in flavor, fragrance and QA/QC settings
Future Trends and Opportunities
Advances in high-speed acquisition and machine-learning deconvolution promise even faster and more accurate profiling of complex matrices. Integration with cloud-based spectral libraries and real-time feedback could further accelerate method development and industrial monitoring.
Conclusion
GC-TOFMS with spectral deconvolution enables precise area percent measurements for essential oil analysis despite severe coelutions. The approach reduces chromatographic run times, improves data reliability and boosts laboratory productivity.
Instrumentation Used
- Gas Chromatograph: Agilent 6890 (EPC mode)
- Column: DB-5, 10 m × 0.18 mm × 0.18 µm
- Detector: Pegasus II GC-TOFMS (EI mode), mass range 45–450 amu, 50 spectra/s, 200 °C ion source
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Rapid Qualitative GC-TOFMS Analysis of Unleaded Gasoline
2010|Agilent Technologies|Applications
® Rapid Qualitative GC-TOFMS Analysis of Unleaded Gasoline LECO Corporation; Saint Joseph, Michigan USA Key Words: GC-TOFMS, Petrochemical, Deconvolution 1. Introduction Analyses of petroleum fuels are complicated by the relatively large number of volatile and semivolatile components contained in these…
Key words
similarity, similarityreverse, reversesec, secname, namepeak, peakpegasus, pegasusscience, sciencelife, lifedelivering, deliveringunleaded, unleadedchemical, chemicalright, rightsolutions, solutionsspectra, spectraaccurately
Qualitative Analysis of Essential Oils Using GC/MS with Hydrogen Carrier Gas and the Agilent HydroInert Source
2024|Agilent Technologies|Applications
Application Note Food & Beverage Testing Qualitative Analysis of Essential Oils Using GC/MS with Hydrogen Carrier Gas and the Agilent HydroInert Source Author Abstract Bruce D. Quimby Agilent Technologies, Inc. Due to ongoing concerns with the price and availability of…
Key words
counts, countshydrogen, hydrogenhydroinert, hydroinerthelium, heliumlms, lmssource, sourceoxide, oxidecarvone, carvonecharge, chargeextractor, extractordefaults, defaultsmethod, methoddelta, deltaoil, oilneroli
Analysis of Grape Volatiles by Solid Phase Microextraction and Comprehensive Two-Dimensional Gas Chromatography with Time-of-Flight Mass Spectrometry (GCxGC-TOFMS)
2008|Agilent Technologies|Applications
® Analysis of Grape Volatiles by Solid Phase Microextraction and Comprehensive Two-Dimensional Gas Chromatography with Time-of-Flight Mass Spectrometry (GCxGC-TOFMS) LECO Corporation; Saint Joseph, Michigan USA Key Words: GCxGC-TOFMS, SPME, Food and Flavors The aim of this work was to study…
Key words
isomer, isomerester, estergcxgc, gcxgcgrapes, grapesvarietal, varietalindene, indeneacid, acidethyl, ethylbutanoic, butanoicionone, iononehexanoic, hexanoicdodecane, dodecanegrape, grapechromatof, chromatofscience
Analysis of Perfumes With SPME and High-Speed GC-TOFMS
2008|Agilent Technologies|Applications
® Analysis of Perfumes With SPME and High-Speed GC-TOFMS LECO Corporation; Saint Joseph, Michigan USA Key Words: GC-TOFMS, SPME, Low Thermal Mass (LTM), Headspace, VOC's In this study, several perfumes are analyzed using Solid Phase Micro Extraction (SPME) coupled to…
Key words
tolerance, toleranceuniquemass, uniquemasspinene, pinenemyrcene, myrcenearea, areaunknown, unknownout, outbuilding, buildingindoor, indoorspme, spmecompare, comparecamphene, camphenetofms, tofmsname, namesick