Comparison of Fritted and Wool Liners for Analysis of Semivolatile Organic Compounds by Gas Chromatography/Mass Spectrometry
Applications | 2020 | Agilent TechnologiesInstrumentation
Gas chromatography coupled with mass spectrometry (GC/MS) is a cornerstone technique for detecting semivolatile organic compounds (SVOCs) in environmental samples. The choice of inlet liner critically influences system uptime, analytical reliability, and the frequency of maintenance interventions. By comparing fritted and wool-packed liners, laboratories can optimize performance, reduce downtime, and ensure consistent quantitation of trace contaminants.
This study evaluated four splitless single-taper liners—one fritted and three glass-wool types—to assess their initial inertness, calibration compatibility, and longevity under a heavy soil extract matrix challenge. Key targets included SVOCs specified in EPA Methods 8270D/E, using a 97-compound standard mix enriched with deuterated internal standards.
Soil extracts prepared according to EPA 8270 procedures were spiked with a working standard (0.1–100 µg/mL) and an internal standard blend (40 µg/mL). Analyses were conducted on an Agilent 7890B GC with a J&W DB-8270D Ultra Inert column (30 m×0.25 mm×0.25 µm) and an Agilent 5977A inert-EI MSD. Injection parameters included 1 µL splitless pulses and a stepped oven program from 40 °C to 320 °C. System suitability checks employed a DFTPP and DDT tuning standard to monitor ion ratios, tailing factors, and 4,4′‐DDT breakdown.
Initial inertness was comparable across liners, with 4,4′‐DDT breakdown below 2 % and tailing factors under 1.2. Calibration curves generated on the fritted liner remained valid when the column was paired with wool liners. Under repetitive soil injections, the fritted liner sustained acceptable DDT breakdown (<20 %) for an average of 24 matrix injections, more than double the ~10 injections achieved by wool liners. Critical calibration verification (CCV) failures remained below 10 % compounds for the frit liner, whereas wool liners approached the study’s failure threshold sooner. Liner or column maintenance restored performance metrics to acceptance criteria.
Advances may include novel inert coatings or alternative frit geometries to further resist matrix carryover. Automated maintenance tracking and predictive modeling of liner life could optimize GC/MS uptime. Integration with high-throughput sampling systems and expansion to more complex matrices (e.g., wastewaters, biological tissues) will enhance method applicability.
The Agilent Ultra Inert splitless low fritted liner demonstrated significantly greater resistance to matrix-induced activity than conventional glass wool liners, achieving more than twice the average lifetime under soil extract challenge. Its compatibility with established EPA 8270 calibration routines and easy interchangeability enhances robustness and cost efficiency in SVOC analysis.
GC/MSD, GC/SQ, Consumables
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Significance of the Topic
Gas chromatography coupled with mass spectrometry (GC/MS) is a cornerstone technique for detecting semivolatile organic compounds (SVOCs) in environmental samples. The choice of inlet liner critically influences system uptime, analytical reliability, and the frequency of maintenance interventions. By comparing fritted and wool-packed liners, laboratories can optimize performance, reduce downtime, and ensure consistent quantitation of trace contaminants.
Objectives and Study Overview
This study evaluated four splitless single-taper liners—one fritted and three glass-wool types—to assess their initial inertness, calibration compatibility, and longevity under a heavy soil extract matrix challenge. Key targets included SVOCs specified in EPA Methods 8270D/E, using a 97-compound standard mix enriched with deuterated internal standards.
Methodology and Instrumentation
Soil extracts prepared according to EPA 8270 procedures were spiked with a working standard (0.1–100 µg/mL) and an internal standard blend (40 µg/mL). Analyses were conducted on an Agilent 7890B GC with a J&W DB-8270D Ultra Inert column (30 m×0.25 mm×0.25 µm) and an Agilent 5977A inert-EI MSD. Injection parameters included 1 µL splitless pulses and a stepped oven program from 40 °C to 320 °C. System suitability checks employed a DFTPP and DDT tuning standard to monitor ion ratios, tailing factors, and 4,4′‐DDT breakdown.
Main Results and Discussion
Initial inertness was comparable across liners, with 4,4′‐DDT breakdown below 2 % and tailing factors under 1.2. Calibration curves generated on the fritted liner remained valid when the column was paired with wool liners. Under repetitive soil injections, the fritted liner sustained acceptable DDT breakdown (<20 %) for an average of 24 matrix injections, more than double the ~10 injections achieved by wool liners. Critical calibration verification (CCV) failures remained below 10 % compounds for the frit liner, whereas wool liners approached the study’s failure threshold sooner. Liner or column maintenance restored performance metrics to acceptance criteria.
Contributions and Practical Applications
- The sintered glass frit serves as an effective barrier to nonvolatile residues, extending liner life and minimizing column fouling.
- Calibration integrity is maintained across liner exchanges, allowing flexibility in maintenance scheduling.
- Environmental testing laboratories benefit from reduced downtime, lower operating costs, and reliable quantitation of regulated SVOCs.
Future Trends and Opportunities
Advances may include novel inert coatings or alternative frit geometries to further resist matrix carryover. Automated maintenance tracking and predictive modeling of liner life could optimize GC/MS uptime. Integration with high-throughput sampling systems and expansion to more complex matrices (e.g., wastewaters, biological tissues) will enhance method applicability.
Conclusion
The Agilent Ultra Inert splitless low fritted liner demonstrated significantly greater resistance to matrix-induced activity than conventional glass wool liners, achieving more than twice the average lifetime under soil extract challenge. Its compatibility with established EPA 8270 calibration routines and easy interchangeability enhances robustness and cost efficiency in SVOC analysis.
References
- Padilla-Sánchez J.A., Plaza-Bolaños P., Frenich A.G. Applications and Strategies Based on Gas Chromatograph–Low-Resolution Mass Spectrometry (GC-LRMS) for the Determination of Residues and Organic Contaminants in Environmental Samples. In Comprehensive Analytical Chemistry; Cappiello A., Palma P., Eds.; Elsevier, 2013; Vol. 61, pp 181–199.
- United States Environmental Protection Agency. Semivolatile Organic Compounds by Gas Chromatography/Mass Spectrometry (GC/MS); Method 8270D, Revision 4, February 2007.
- United States Environmental Protection Agency. Semivolatile Organic Compounds by Gas Chromatography/Mass Spectrometry (GC/MS); Method 8270E, Revision 4, June 2018.
- Smith Henry A. Analysis of Semivolatile Organic Compounds with Agilent Sintered Frit Liner by Gas Chromatography/Mass Spectrometry. Agilent Technologies, Publication 5994-0953EN, 2019.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Selecting a GC Liner to Decrease Column Trim Frequency
2022|Agilent Technologies|Technical notes
Application Brief Environmental Selecting a GC Liner to Decrease Column Trim Frequency Author Abstract Angela Smith Henry, PhD Agilent Technologies, Inc. GC inlet liner selection is an important part of reduced column maintenance. A liner with a frit or wool…
Key words
liner, linerfritted, frittedbreakdown, breakdownddt, ddtpercent, percenttaper, taperwool, woolcolumn, columnmatrix, matrixtrim, trimliners, linerstrims, trimsinitial, initialsingle, singlefrit
Analysis of Semivolatile Organic Compounds with Agilent Sintered Frit Liner by Gas Chromatography/Mass Spectrometry
2019|Agilent Technologies|Applications
Application Note Environmental Analysis Analysis of Semivolatile Organic Compounds with Agilent Sintered Frit Liner by Gas Chromatography/Mass Spectrometry Author Angela Smith Henry, PhD Agilent Technologies, Inc. Abstract Gas chromatography/mass spectrometry (GC/MS) is integral to the analysis of semivolatile organic compounds…
Key words
nitroaniline, nitroanilineliner, linerpentachlorophenol, pentachlorophenoltrim, trimbenzo, benzobenzidine, benzidinephthalate, phthalatefluoranthene, fluoranthenepass, passccv, ccvbenzoic, benzoicsintered, sinteredfrit, fritcolumn, columncompounds
Endrin/DDT Stability for Fritted and Wool Liners
2020|Agilent Technologies|Applications
Application Note Environmental Endrin/DDT Stability for Fritted and Wool Liners Authors Lukas Wieder and Angela Smith Henry, PhD Agilent Technologies, Inc. Abstract This application note demonstrates the lifetime of Agilent Ultra Inert fritted liners for environmental applications with complex, nonvolatile…
Key words
wool, woolfrit, fritmid, midliner, linerendrin, endrinlow, lowbreakdown, breakdownagilent, agilentddt, ddtliners, linersfritted, frittedmixture, mixtureddd, dddinert, inertdde
Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9000 Gas Chromatograph
2018|Agilent Technologies|Applications
Application Note Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9000 Gas Chromatograph Author Matthew Giardina, Ph.D., Applications Chemist, Gas Phase Separations Division, Agilent Technologies, Inc. Abstract Gas chromatography/mass spectrometry is a workhorse for the analysis of semivolatile organic…
Key words
intuvo, intuvochip, chipreplacement, replacementguard, guardpass, passccv, ccvbenzidine, benzidinebenzo, benzomatrix, matrixresponse, responsepentachlorophenol, pentachlorophenolcalibration, calibrationafter, afternumber, numberinjections