Analysis of Gas-Phase Formaldehyde by GC-FID Method Utilizing DNPH Derivatization
Applications | 2020 | ShimadzuInstrumentation
Formaldehyde is one of the simplest yet most hazardous airborne aldehydes. Its widespread presence in indoor and outdoor environments poses health risks, making reliable trace measurement crucial for air quality monitoring, regulatory compliance, and occupational safety.
This study presents a sensitive gas-phase formaldehyde quantitation method. By derivatizing formaldehyde with 2,4-dinitrophenylhydrazine (DNPH) and analyzing the resulting hydrazone derivatives via gas chromatography with flame ionization detection (GC-FID), the work aims to achieve linear response, low detection limits, and practical sampling workflows.
Cartridges packed with silica gel coated in DNPH trap formaldehyde from gas samples drawn by a pump. Collected hydrazones are eluted with acetonitrile, passed through an ion exchange resin to remove excess DNPH, and injected into a Shimadzu Nexis GC-2030 AF system with an AOC-20i Plus autosampler. Key chromatographic conditions include split injection (1:30), a SH-Rtx-1 capillary column (30 m × 0.32 mm I.D., 1 μm), temperature programming from 240 °C to 260 °C, and an FID operated at 280 °C with hydrogen, air, and helium flows.
A calibration curve prepared using DNPH-formaldehyde standards over 1 to 100 ng/μL exhibited excellent linearity (R2 0.9999) and repeatability at each level (RSD ≤ 1%). Signal-to-noise measurements at 1 ng/μL indicated a detection limit near 0.1 ng/μL. Field samples with approximately 1 and 10 ppm formaldehyde produced consistent mass-based quantitation across dual cartridges, confirming no breakthrough. Moisture control via an ozone scrubber proved essential in preserving derivative stability.
Advances may include integration of online sampling with automated derivatization, miniaturized GC-FID systems for portable monitoring, and application to other reactive carbonyls. Coupling the DNPH cartridge approach with sensor arrays or digital data platforms can enhance real-time air quality management and risk assessment.
The presented DNPH derivatization coupled with GC-FID delivers a reliable, sensitive analysis of gas-phase formaldehyde. Excellent calibration linearity, low detection limits, and cartridge-based sampling make this method highly practical for diverse air monitoring needs. Instrument choice should balance required sensitivity with operational simplicity.
GC
IndustriesEnergy & Chemicals
ManufacturerShimadzu
Summary
Importance of Gas-Phase Formaldehyde Analysis
Formaldehyde is one of the simplest yet most hazardous airborne aldehydes. Its widespread presence in indoor and outdoor environments poses health risks, making reliable trace measurement crucial for air quality monitoring, regulatory compliance, and occupational safety.
Objectives and Study Overview
This study presents a sensitive gas-phase formaldehyde quantitation method. By derivatizing formaldehyde with 2,4-dinitrophenylhydrazine (DNPH) and analyzing the resulting hydrazone derivatives via gas chromatography with flame ionization detection (GC-FID), the work aims to achieve linear response, low detection limits, and practical sampling workflows.
Methodology and Instrumentation
Cartridges packed with silica gel coated in DNPH trap formaldehyde from gas samples drawn by a pump. Collected hydrazones are eluted with acetonitrile, passed through an ion exchange resin to remove excess DNPH, and injected into a Shimadzu Nexis GC-2030 AF system with an AOC-20i Plus autosampler. Key chromatographic conditions include split injection (1:30), a SH-Rtx-1 capillary column (30 m × 0.32 mm I.D., 1 μm), temperature programming from 240 °C to 260 °C, and an FID operated at 280 °C with hydrogen, air, and helium flows.
Main Results and Discussion
A calibration curve prepared using DNPH-formaldehyde standards over 1 to 100 ng/μL exhibited excellent linearity (R2 0.9999) and repeatability at each level (RSD ≤ 1%). Signal-to-noise measurements at 1 ng/μL indicated a detection limit near 0.1 ng/μL. Field samples with approximately 1 and 10 ppm formaldehyde produced consistent mass-based quantitation across dual cartridges, confirming no breakthrough. Moisture control via an ozone scrubber proved essential in preserving derivative stability.
Benefits and Practical Applications
- The DNPH-GC-FID approach offers robust sensitivity and quantitative accuracy for trace formaldehyde in air.
- Cartridge-based sampling simplifies field workflows and minimizes sample preparation time compared with HPLC or GC-MS methods.
- The method supports occupational exposure assessment, environmental monitoring, indoor air quality studies, and compliance testing.
Future Trends and Potential Applications
Advances may include integration of online sampling with automated derivatization, miniaturized GC-FID systems for portable monitoring, and application to other reactive carbonyls. Coupling the DNPH cartridge approach with sensor arrays or digital data platforms can enhance real-time air quality management and risk assessment.
Conclusion
The presented DNPH derivatization coupled with GC-FID delivers a reliable, sensitive analysis of gas-phase formaldehyde. Excellent calibration linearity, low detection limits, and cartridge-based sampling make this method highly practical for diverse air monitoring needs. Instrument choice should balance required sensitivity with operational simplicity.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Alternative Analysis of Formaldehyde-DNPH and Other Carbonyl-DNPH Derivatives by Capillary GC
1998|Merck|Applications
Application Note 107 Alternative Analysis of Formaldehyde-DNPH and Other Carbonyl-DNPH Derivatives by Capillary GC GC analysis of 15 carbonyl-DNPH compounds has some advantages over the HPLC analyses described in US EPA and other environmental methods. GC provides good resolution and…
Key words
dnph, dnphformaldehyde, formaldehydetolualdehyde, tolualdehydecarbonyl, carbonylfid, fidexcess, excesshexaldehyde, hexaldehydeflame, flameacceptably, acceptablyextinguish, extinguishisovaleraldehyde, isovaleraldehydecrotonaldehyde, crotonaldehydevaleraldehyde, valeraldehydebutyraldehyde, butyraldehydealternative
Analysis of Lower Aliphatic Aldehydes using Nexis GC-2030
2017|Shimadzu|Applications
LAAN-A-GC-E057 Application News No. G292 Gas Chromatograph Analysis of Lower Aliphatic Aldehydes using Nexis GC-2030 Lower aliphatic aldehydes are known to be associated with sick building syndrome. Thus various regulations require strict control of their concentrations. A common method of…
Key words
aldehydes, aldehydesaliphatic, aliphaticgas, gasschematic, schematiccollection, collectionflowrate, flowratediagram, diagrammethod, methoddnphs, dnphscartridges, cartridgeslower, lowerflame, flamecartridge, cartridgethermionic, thermionicnews
Determination of Formaldehyde and VOCs in Wood-based Products using an Automated Micro-Scale Chamber
2017|Agilent Technologies|Applications
GERSTEL Application Note No. 193, 2017 Determination of Formaldehyde and VOCs in Wood-based Products using an Automated Micro-Scale Chamber Yunyun Nie, Oliver Lerch, Eike Kleine-Benne GERSTEL GmbH & Co. KG, Eberhard-Gerstel-Platz 1, 45473 Mülheim an der Ruhr, Germany ABSTRACT INTRODUCTION…
Key words
dnph, dnphformaldehyde, formaldehydeldd, lddchamber, chamberemission, emissionair, airdad, dadscale, scalelod, lodmicro, microbuilding, buildingplywood, plywoodacrolein, acroleinloq, loqhexanal
Automated Determination of Formaldehyde Emissions from Materials by On-Sorbent Derivatization and Thermal Desorption GC/MS
2018|Agilent Technologies|Applications
GERSTEL Application Note No. 197, 2018 Automated Determination of Formaldehyde Emissions from Materials by On-Sorbent Derivatization and Thermal Desorption GC/MS Yunyun Nie, Lokman Coban, Shuai Li, and Eike Kleine-Benne GERSTEL GmbH & Co. KG, Eberhard-Gerstel-Platz 1, 45473 Mülheim an der…
Key words
pfph, pfphformaldehyde, formaldehydetenax, tenaxdnph, dnphloaded, loadedcarbonyls, carbonylsderivatization, derivatizationgerstel, gerstelvocs, vocsairborne, airbornecandle, candlesampling, samplingsorbent, sorbenttubes, tubestube