Automated Determination of Formaldehyde Emissions from Materials by On-Sorbent Derivatization and Thermal Desorption GC/MS

Applications | 2018 | GERSTELInstrumentation
GC/MSD, HeadSpace, Thermal desorption, GC/SQ
Industries
Energy & Chemicals , Materials Testing
Manufacturer
Agilent Technologies, GERSTEL

Summary

Importance of the Topic


The accurate and automated measurement of formaldehyde and other carbonyls from building materials and consumer products is critical for indoor air quality, health risk assessment, and regulatory compliance. Traditional DNPH/HPLC methods involve manual sampling, consumable cartridges, and solvent elution, leading to high costs, limited analyte scope, and elevated detection limits. A fully automated thermal desorption GC/MS approach using on-sorbent derivatization offers a solvent-free, multi-target solution with improved sensitivity and workflow efficiency.

Study Objectives and Overview


This work aimed to develop and validate a fully automated method combining in-situ pentafluorophenylhydrazine (PFPH) derivatization, dynamic headspace (DHS) loading onto Tenax TA tubes, thermal desorption, and GC/MS analysis. Key goals included minimizing background contamination, optimizing reagent loading and sampling parameters, assessing tube reusability, and demonstrating method performance through calibration and real-world emission testing on plywood and scented candles.

Methodology and Used Instrumentation


The system comprised a GERSTEL MultiPurpose Sampler (MPS) with DHS for automated PFPH loading, a TDU 2 thermal desorption unit, a CIS 6 cold trap inlet, and an Agilent GC/MS single quadrupole detector. Tenax TA tubes were loaded with 100 µL of a 1.44 mM PFPH stock via DHS (30 °C, 50 mL/min, 700 mL purge) to yield 144 nmol PFPH per tube. Formaldehyde sampling (10 µL aqueous aliquots) employed DHS at 50 °C, 40 mL/min, 500 mL purge, followed by desorption (30 °C to 280 °C) and GC separation on an HP-5 column with split 1/40. MS detection covered 30–450 amu.

Main Results and Discussion


• Automated DHS loading reduced PFPH-CH2O background to 0.6–2.5 % and enabled tube reuse after thermal desorption without carryover.
• Optimal formaldehyde collection occurred at 40 mL/min and ≥350 mL purge volume; 500 mL chosen for robustness.
• Calibration linear from 8 to 81 nmol (0.2–2.0 µg) with R2 = 0.9989 and average RSD 1.4 % (n = 5).
• LOD and LOQ for formaldehyde were 3.0 nmol (0.09 µg) and 8.3 nmol (0.25 µg), respectively.
• Plywood emissions measured at 0.51 mg/m3; scented candle at 0.09 mg/m3; additional carbonyls (pentanal, hexanal, nonanal, Triplals) were efficiently captured and identified via EI mass spectra.

Benefits and Practical Applications


  • One-step automation for VOCs, SVOCs, and carbonyls eliminates DNPH tubes and liquid elution.
  • Reusability of Tenax TA tubes lowers consumable costs and sample-to-sample variability.
  • GC/MS provides high separation power and structural confirmation, even for isomeric derivatives.
  • Solvent-free workflow enhances laboratory safety and reduces environmental impact.

Future Trends and Potential Applications


Expansion to larger-scale chambers and higher sample volumes through advanced flow controllers; adaptation to diverse matrices such as e-liquids, food packaging, and indoor air monitoring; integration with high-resolution MS for ultra-trace and ultra-polar carbonyls; and development of on-line sampling for real-time emission profiling.

Conclusion


The fully automated PFPH/DHS-TD-GC/MS method delivers sensitive, reproducible, and broad-scope analysis of formaldehyde and related carbonyl emissions, while simplifying sample preparation and reducing blanks. It meets international guideline criteria and offers a versatile platform for material emission testing and indoor air quality assessment.

References


  • Ho SS, Yu JZ. Environ. Sci. Technol. 2004;38:862–870.
  • Nie Y, Lerch O, Kleine-Benne E. GERSTEL AppNote No. 193, 2017.
  • Salthammer T, Mentese S. Chemosphere. 2008;72:1351–1356.
  • HJ/T 400-2007: Determination of VOCs and Carbonyls in Vehicle Cabins.
  • Cecinato A. Chromatographia. 2001;54:263–269.
  • Perkins M. Anatune AS147, 2014.
  • Salthammer T, Mentese S, Marutzky R. Chem. Rev. 2010;110:2536–2572.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Automated Determination of Formaldehyde Emissions from E-Liquids by On-Sorbent Derivatization and Thermal Desorption GC/MS
GERSTEL Application Note No. 201, 2018 Automated Determination of Formaldehyde Emissions from E-Liquids by On-Sorbent Derivatization and Thermal Desorption GC/MS Yunyun Nie, Lokman Coban, Shuai Li, and Eike Kleine-Benne GERSTEL GmbH & Co. KG, Eberhard-Gerstel-Platz 1, 45473 Mülheim an der…
Key words
pfph, pfphformaldehyde, formaldehydetenax, tenaxcarbonyls, carbonylsgerstel, gerstelloaded, loadeddetermination, determinationcigarette, cigarettehydrozone, hydrozonederivatization, derivatizationdesorption, desorptionbrand, brandderivatives, derivativesvocs, vocsagent
Determination of Formaldehyde and VOCs in Wood-based Products using an Automated Micro-Scale Chamber
GERSTEL Application Note No. 193, 2017 Determination of Formaldehyde and VOCs in Wood-based Products using an Automated Micro-Scale Chamber Yunyun Nie, Oliver Lerch, Eike Kleine-Benne GERSTEL GmbH & Co. KG, Eberhard-Gerstel-Platz 1, 45473 Mülheim an der Ruhr, Germany ABSTRACT INTRODUCTION…
Key words
dnph, dnphformaldehyde, formaldehydeldd, lddchamber, chamberemission, emissionair, airdad, dadscale, scalelod, lodmicro, microbuilding, buildingplywood, plywoodacrolein, acroleinloq, loqhexanal
Direct Thermal Extraction Analysis of Solid and Liquid Samples using the GERSTEL MPS Robotic Sampler, Thermal Desorption Unit (TDU 2) and Cooled Inlet System (CIS)
GERSTEL Application Note No. 205, 2019 Direct Thermal Extraction Analysis of Solid and Liquid Samples using the GERSTEL MPS Robotic Sampler, Thermal Desorption Unit (TDU 2) and Cooled Inlet System (CIS) Laurel A. Vernarelli, Jacqueline A. Whitecavage, and John R.…
Key words
scented, scentedpumpkin, pumpkinbrand, brandcandle, candlewax, waxspice, spicecinnamaldehyde, cinnamaldehydegerstel, gerstelair, aircandles, candleslinalool, linalooligniting, ignitingtdu, tduthermal, thermaldte
The Development of Standard Methods Relating to Vehicle Interior Air Quality (VIAQ) and How to Comply With Them
Application Note The Development of Standard Methods Relating to Vehicle Interior Air Quality (VIAQ) and How to Comply With Them Abstract This Application Note discusses the introduction of regulations limiting the quantities of volatile and semivolatile organic compounds (VOCs and…
Key words
sampling, samplingvehicle, vehicleviaq, viaqemissions, emissionsvapors, vaporsvocs, vocscar, carinterior, interiorchamber, chambertrim, trimdesorbed, desorbedtrap, trapvolatile, volatilesvocs, svocstube
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike