GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Automated MRM Method Development for US EPA Method 8270 with the Agilent MassHunter Optimizer for GC/TQ

Applications | 2020 | Agilent TechnologiesInstrumentation
GC/MSD, GC/MS/MS, GC/QQQ, Software
Industries
Environmental
Manufacturer
Agilent Technologies

Summary

Importance of the topic


Automating development of multiple reaction monitoring methods for US EPA 8270 compounds accelerates environmental contaminant analysis, improves reproducibility, and reduces manual workload in analytical laboratories. High-throughput MRM workflows enable consistent, robust semivolatile organic compound quantitation in complex matrices, supporting regulatory compliance and quality assurance.

Objectives and overview of the study


This work demonstrates Agilent MassHunter Optimizer for GC/TQ as an end-to-end solution for generating optimized MRM acquisition methods for 83 semivolatiles defined in EPA Method 8270. Four distinct workflows—starting from full-scan data, SIM ions, existing MRM transitions, and retention-time updates—were evaluated to assess automation efficiency, coelution handling, and method transfer from single-quad GC/MSD to triple-quad GC/TQ.

Methodology and instrumentation


The study used an Agilent 7890 GC coupled to a 7000D triple quadrupole mass spectrometer with MassHunter GC/MS Data Acquisition v10.0 and Unknowns Analysis v10.0. A 22-minute chromatographic method for 83 semivolatile targets (full mix) was challenged by coelutions. NIST17 spectral library guided compound identification. Key optimization steps included:
  • Library search of deconvoluted full-scan spectra to identify precursor ions
  • Product ion scans at multiple collision energies (5, 15, 25, 35 eV)
  • Selection of top product ions based on abundance and cluster uniqueness
  • Collision energy ramp optimization around selected energies
  • Retention-time updating via acquisition of SIM or MRM data under new chromatographic conditions

Main results and discussion


All four workflows achieved fully developed MRM methods with significant time savings:
  • From full-scan data: 12 injections in 6 h
  • From SIM ions: 11 injections in 5.5 h
  • From existing MRMs: 4 injections in 2 h
  • Retention-time updates: 1 injection for minor shifts (0.5 h) and 6 injections for major shifts (3 h)
Coelution effects required additional injections and manual review in complex mixtures. Spectral deconvolution reliably distinguished overlapping peaks, and user-guided overwriting of automatically selected ions ensured method specificity.

Benefits and practical applications


Automated MRM method development yields:
  • Reduced hands-on time and method development cycle
  • Improved transfer of legacy GC/MSD SIM methods to GC/TQ MRM
  • Enhanced reproducibility and recordkeeping through software-driven workflows
  • Built-in review tools for rapid validation of transitions
This approach supports environmental testing, regulatory laboratories, and QA/QC operations handling large sample volumes.

Future trends and applications


Emerging directions include integration of machine learning for predictive collision energy settings, real-time method adaptation using dynamic MRM, cloud-based library sharing, and expansion to multi-residue screening in diverse matrices. Enhanced instrument-software interoperability and LIMS connectivity will further streamline high-throughput environmental analyses.

Conclusion


Agilent MassHunter Optimizer for GC/TQ accelerates and standardizes the development of MRM methods for EPA Method 8270 semivolatiles. Automated ion identification, energy optimization, and retention-time updating deliver robust, reproducible acquisition methods with minimal manual intervention, improving laboratory efficiency and data quality.

Reference


  1. Churley M, Szelewski M, Quimby B. EPA 8270 Re-Optimized for Widest Calibration Range on the 5977 Inert Plus GC/MSD. Agilent Technologies Application Note 5994-0350EN, 2018.
  2. Churley M, Quimby B, Andrianova A. A Fast Method for EPA 8270 in MRM Mode Using the 7000 Series Triple Quadrupole GC/MS. Agilent Technologies Application Note 5994-0691EN, 2019.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Automated MRM Method Development for Pesticides in Cannabis Using the Agilent MassHunter Optimizer for GC/TQ
Application Note Cannabis & Hemp Testing Automated MRM Method Development for Pesticides in Cannabis Using the Agilent MassHunter Optimizer for GC/TQ Authors Anastasia Andrianova, He Liu, and Alex Graettinger Agilent Technologies, Inc. Abstract This application note demonstrates the use of…
Key words
collision, collisionmrm, mrmcannabis, cannabisprecursor, precursorenergies, energiesions, ionstransitions, transitionsoptimization, optimizationion, ionproduct, productenergy, energyidentification, identificationoptimizer, optimizerscan, scandeveloped
An Optimization Tool for MS Signal Acquisition in GC Triple Quadrupole Mass Spectrometry
Poster Reprint ASMS 2019 TP305 An Optimization Tool for MS Signal Acquisition in GC Triple Quadrupole Mass Spectrometry Anastasia Andrianova1, Melissa Churley2 1Agilent Technologies Inc., Wilmington, DE 19808 USA; 2Agilent Technologies Inc., Santa Clara, CA 95051 USA Introduction Development of…
Key words
collision, collisionoptimization, optimizationmrm, mrmenergy, energyions, ionsoptimizer, optimizerscan, scanproduct, productselection, selectionmethod, methodtool, toolenergies, energiesdeconvoluted, deconvolutedprecursor, precursortriple
A Forensic Triple Quadrupole GC/MS MRM Database for Forensic and Toxicological Workflows
Application Note Forensics A Forensic Triple Quadrupole GC/MS MRM Database for Forensic and Toxicological Workflows Authors Celine Gys1, Anna Klimowska1,2, and Adrian Covaci1 Toxicological Center, University of Antwerp, Universiteitsplein 1, Wilrijk, 2610, Belgium 1 Department of Toxicology, Medical University of…
Key words
database, databasetms, tmsmrm, mrmforensic, forensiccas, castransitions, transitionsacquisition, acquisitiondata, datatoxicological, toxicologicaltoxicants, toxicantsmasshunter, masshuntercompound, compoundketamine, ketaminecsv, csvcompounds
A Triple Quadrupole GC/MS MRM Database for Forensic and Toxicological Workflows
Poster Reprint ASMS 2024 Poster number WP 810 A Triple Quadrupole GC/MS MRM Database for Forensic and Toxicological Workflows Celine Gys1, Anna Klimowska1,2, Adrian Covaci1, Remko van Loon3, Joel Ferrer4, Anastasia Andrianova4 1Toxicological Center, University of Antwerp, Wilrijk, Belgium; 2Department…
Key words
database, databaseforensic, forensicmrm, mrmtoxicological, toxicologicaltoxicants, toxicantsacquisition, acquisitiondetected, detectedtransitions, transitionsoptimizer, optimizermasshunter, masshunterauthentic, authenticdmrm, dmrmtemperature, temperatureretention, retentionexercising
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike