GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Integration and Optimization of Hardware and Software for a Differential Flow Modulated GCxGC

Presentations | 2007 | Agilent TechnologiesInstrumentation
GCxGC
Industries
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Comprehensive two-dimensional gas chromatography (GC×GC) with differential flow modulation enhances separation power and sensitivity for complex mixtures. Integrating optimized hardware and software components streamlines operation, eliminates cryogens, and extends temperature limits, enabling robust high-throughput analysis in environmental, petrochemical, and food testing.

Objectives and Study Overview


This study presents the design, implementation, and evaluation of a flow‐modulated GC×GC system using Agilent’s capillary flow technology. Key goals include seamless synchronization of modulation events, reliable performance without cryogenic cooling, and maintenance of chromatographic resolution and sensitivity comparable to thermal modulation.

Methodology and Instrumentation


  • Instrument Configuration
    • Split/splitless inlet with hydrogen carrier gas at 0.8 mL/min
    • 7890A valve driver and timing board
    • Capillary Flow Technology (CFT) modulator with three‐way differential flow valve
    • Two columns: first nonpolar 30 m×0.25 mm, second polar 5 m×0.25 mm
    • Flame ionization detector (FID) operating at 200 Hz
    • 7683 autosampler for automated injections
  • Data Processing Software
    • GC Image (2D visualization and data analysis)
    • Zoex Corporation software for modulation control and timing

Used Instrumentation


  • Photolithographically machined CFT modulator with diffusion‐bonded flow plate for low dead volume
  • Projection‐welded fittings and deactivated surfaces for inertness
  • Micro modulation valve enabling precise flush and collect steps
  • ChemStation interface for setting modulation and injection timings

Main Results and Discussion


  • Modulation Performance
    • Load (collect) and inject phases achieved differential flows of ~21 mL/min and 0.8 mL/min respectively
    • Peak area agreement between modulated and unmodulated runs within 4%
    • Second‐dimension peak widths of 60–100 ms at half‐height without cryo‐focusing
  • Repeatability
    • Overlays of nine C10 injections and single C14 modulations illustrate sub‐1% retention time precision
  • Application Examples
    • Hydrocarbon characterization: heavy gasoline and kerosene analyses under temperature ramps
    • Biodiesel profiling: C10–C18 fatty acid methyl esters in B20 blends
    • GC×GC–FID/MS coupling: detailed two‐dimensional maps of complex gasoline samples
  • System Pressures
    • Column head pressures maintained at ~23.5 psi for first column and ~19.7 psi for second column at 50 °C

Practical Benefits and Applications


  • Cryogen‐Free Operation
    Cryogen elimination reduces cost and complexity, enabling high‐temperature separations up to 350 °C.
  • Shortened Run Times
    Typical GC×GC runs require less than half the time of thermal modulation systems.
  • Enhanced Throughput
    Automated modulation ensures reproducible injections and minimal intervention.
  • Broad Applicability
    Suitable for petrochemical, environmental, and biofuel analyses where high resolution and sensitivity are critical.

Future Trends and Potential Applications


Advances may include integration with high‐resolution mass spectrometry, real‐time process monitoring, and machine‐learning–driven data interpretation. Development of miniaturized modulators and universal software interfaces will expand GC×GC accessibility across laboratories and on‐site analytics.

Conclusion


The optimized integration of Agilent’s capillary flow modulator with the 7890A GC platform delivers reliable, high‐resolution GC×GC separations without cryogens. This configuration achieves rapid run times, broad temperature capability, and sensitivity on par with thermal systems, fulfilling the demands of modern analytical laboratories.

References


No formal reference list was provided in the original document.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Multi-detector Configurations and Applications of Differential Flow Modulation GCXGC
Multi-detector Configurations and Applications of Differential Flow Modulation GCXGC (K.29) Roger L Firor, Agilent Technologies, Inc., Wilmington, DE, USA Multi-Detector Configurations using CFT Technology Introduction ECD-FID-MSD System Performance Pesticide Mix 10ppm: ECD MSD-FID-ECD System Pneumatics The enhanced separation power of…
Key words
mod, modecd, ecdmsd, msdfid, fidmodulated, modulatedrestrictor, restrictorbhc, bhcflow, flowscans, scanspneumatics, pneumaticssecond, secondmodulator, modulatormmi, mmisplitter, splittermodulation
Comprehensive GC System Based on Flow Modulation for the 7890A GC
Comprehensive GC System Based on Flow Modulation for the 7890A GC Application Brief Introduction A hardware solution is available on the 7890A for the practice of comprehensive GC. The system uses a capillary flow modulator controlled by the 7890A GC.…
Key words
modulator, modulatormodulation, modulationflow, flowsecond, secondpcm, pcmvalve, valvetechnology, technologychannel, channelcapillary, capillaryaccessory, accessorycollection, collectionband, bandcryogen, cryogencollect, collectsolenoid
Capillary Flow Technology: GCxGC Flow Modulator
Capillary Flow Technology: GCxGC Flow Modulator
2013|Agilent Technologies|Brochures and specifications
Capillary Flow Technology: GCxGC Flow Modulator GET A SECOND DIMENSION OF INFORMATION ON COMPLEX MIXTURES Maximize information while collecting and analyzing GC data Comprehensive two-dimensional GC, or GCxGC, is a powerful technique that can be used to separate very complex…
Key words
modulator, modulatorflow, flowgcxgc, gcxgcsecond, secondtechnology, technologycapillary, capillarychannel, channelaromatics, aromaticsmodulation, modulationagilent, agilentcollection, collectionfeedstock, feedstockpcm, pcmmixtures, mixturesinlet
Comprehensive Flow Modulated Two-Dimensional Gas Chromatography System
Comprehensive Flow Modulated Two-Dimensional Gas Chromatography System Application Brief Roger L. Firor Two-dimensional GC (GC x GC), a relatively new gas chromatographic technique, is being recognized for its powerful separation capabilities for the analysis of complex mixtures. The methodology involves…
Key words
aromatics, aromaticsmodulation, modulationflow, flowvalve, valvealkanes, alkanesmodulator, modulatormono, monoagilent, agilentdirection, directionburdens, burdenscpm, cpmchannel, channelnaphthalenes, naphthalenessuccessively, successivelyseparation
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike