EPA Method 525.3 OCP Calibration Standard (with 525.3 Internal and Surrogate Standards) on Rxi®-5Sil MS
Applications | | RestekInstrumentation
The accurate quantification of organochlorine pesticides and related compounds in environmental and water samples is critical for regulatory compliance and public health protection. EPA Method 525.3 provides a standardized protocol for analysis of a range of chlorinated analytes. Using a robust GC-MS platform with the Rxi-5Sil MS column enhances sensitivity, resolution, and reproducibility, enabling laboratories to meet strict detection limits and quality assurance criteria.
This work describes the chromatographic separation and calibration of 36 target analytes, including surrogate and internal standards, in a mixed calibration solution following EPA Method 525.3. The objectives are to demonstrate retention time reproducibility, peak resolution, and method suitability for routine monitoring of organochlorine compounds.
The calibration solution contains 2 µg/mL of each pesticide standard (8 µg/mL for pentachlorophenol) in ethyl acetate. A 1 µL pulsed splitless injection is performed on an Agilent 7890B GC coupled to a 5977A MSD. Separation uses an Rxi-5Sil MS column (30 m × 0.25 mm × 0.25 µm) with helium carrier gas at 1.2 mL/min. The oven program ramps from 70 °C to 200 °C at 10 °C/min, then to 320 °C at 7 °C/min. The MSD operates in scan mode (45–550 amu) with a DFTPP tune and electron ionization at 70 eV.
Retention times for analytes spanned 7 to 26 minutes, achieving baseline resolution for critical isomers such as alpha, beta, gamma, and delta hexachlorocyclohexane, as well as chlordane and endosulfan congeners. Surrogate standards 1,3-dimethyl-2-nitrobenzene, triphenyl phosphate, and benzo[a]pyrene-D12 demonstrated consistent retention within ±0.05 min. Peak shapes remained symmetric with minimal tailing under the chosen conditions. The scan rate provided sufficient data points across narrow peaks for reliable quantitation.
Advances in high-resolution mass spectrometry and multidimensional GC techniques may further improve detection limits and selectivity for emerging chlorinated contaminants. Automated sample preparation and data processing workflows will streamline method implementation, while miniaturized systems could enable field-deployable monitoring.
This study validates the EPA Method 525.3 calibration protocol on an Rxi-5Sil MS column, confirming reliable retention, resolution, and detection of a broad suite of organochlorine analytes. The optimized GC-MS conditions support routine environmental monitoring and regulatory compliance.
EPA Method 525.3 Organics Analysis in Drinking Water by Liquid–Solid Extraction and Capillary Column Gas Chromatography/Mass Spectrometry 2008
GC/MSD, GC/SQ, GC columns, Consumables
IndustriesEnvironmental
ManufacturerAgilent Technologies, Restek
Summary
EPA Method 525.3 Calibration Standard Analysis on Rxi-5Sil MS
Significance of the Topic
The accurate quantification of organochlorine pesticides and related compounds in environmental and water samples is critical for regulatory compliance and public health protection. EPA Method 525.3 provides a standardized protocol for analysis of a range of chlorinated analytes. Using a robust GC-MS platform with the Rxi-5Sil MS column enhances sensitivity, resolution, and reproducibility, enabling laboratories to meet strict detection limits and quality assurance criteria.
Aims and Overview of the Study
This work describes the chromatographic separation and calibration of 36 target analytes, including surrogate and internal standards, in a mixed calibration solution following EPA Method 525.3. The objectives are to demonstrate retention time reproducibility, peak resolution, and method suitability for routine monitoring of organochlorine compounds.
Methodology and Used Instrumentation
The calibration solution contains 2 µg/mL of each pesticide standard (8 µg/mL for pentachlorophenol) in ethyl acetate. A 1 µL pulsed splitless injection is performed on an Agilent 7890B GC coupled to a 5977A MSD. Separation uses an Rxi-5Sil MS column (30 m × 0.25 mm × 0.25 µm) with helium carrier gas at 1.2 mL/min. The oven program ramps from 70 °C to 200 °C at 10 °C/min, then to 320 °C at 7 °C/min. The MSD operates in scan mode (45–550 amu) with a DFTPP tune and electron ionization at 70 eV.
Main Results and Discussion
Retention times for analytes spanned 7 to 26 minutes, achieving baseline resolution for critical isomers such as alpha, beta, gamma, and delta hexachlorocyclohexane, as well as chlordane and endosulfan congeners. Surrogate standards 1,3-dimethyl-2-nitrobenzene, triphenyl phosphate, and benzo[a]pyrene-D12 demonstrated consistent retention within ±0.05 min. Peak shapes remained symmetric with minimal tailing under the chosen conditions. The scan rate provided sufficient data points across narrow peaks for reliable quantitation.
Benefits and Practical Application of the Method
- Comprehensive coverage of 36 organochlorine targets with robust surrogate and internal standard strategy
- High chromatographic resolution on Rxi-5Sil MS enables separation of closely eluting congeners
- Pulsed splitless injection enhances sensitivity for trace-level detection
- Standardized conditions ensure reproducibility across laboratories
Future Trends and Opportunities
Advances in high-resolution mass spectrometry and multidimensional GC techniques may further improve detection limits and selectivity for emerging chlorinated contaminants. Automated sample preparation and data processing workflows will streamline method implementation, while miniaturized systems could enable field-deployable monitoring.
Conclusion
This study validates the EPA Method 525.3 calibration protocol on an Rxi-5Sil MS column, confirming reliable retention, resolution, and detection of a broad suite of organochlorine analytes. The optimized GC-MS conditions support routine environmental monitoring and regulatory compliance.
Reference
EPA Method 525.3 Organics Analysis in Drinking Water by Liquid–Solid Extraction and Capillary Column Gas Chromatography/Mass Spectrometry 2008
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
EPA Method 525.3 Full Analyte List on Rxi®-5Sil MS
|Agilent Technologies|Applications
EPA Method 525.3 Full Analyte List on Rxi®-5Sil MS Peaks 1. Diisopropyl methylphosphonate (DIMP) 2. Isophorone 3. 1,3-Dimethyl-2-nitrobenzene (SS) 4. Dichlorvos 5. Hexachlorocyclopentadiene (HCCPD) 6. S-Ethyl dipropylthiocarbamate (EPTC) 7. Mevinphos I 8. Mevinphos II 9. Butylate 10. Vernolate 11. Dimethylphthalate…
Key words
min, mingoup, goupscan, scanpeaks, peaksdftpp, dftppamu, amuextractor, extractorscans, scansdelay, delayethyl, ethylsec, secacetate, acetatestart, starttime, timeoven
EPA Method 525.3 PCB Calibration Standard (with 525.3 Internal and Surrogate Standards) on Rxi®-5Sil MS
|Agilent Technologies|Applications
EPA Method 525.3 PCB Calibration Standard (with 525.3 Internal and Surrogate Standards) on Rxi®-5Sil MS 8 6 19 2 18 1 3 6.00 8.00 10.00 4 12.00 21 5 7 14.00 9 10 11 12 16.00 18.00 Time (min) 13…
Key words
goup, goupscan, scandftpp, dftppmin, minamu, amuextractor, extractorscans, scansdelay, delayethyl, ethylsec, secacetate, acetatestart, starttime, timeoven, ovenquadrupole
EPA Method 525.3 OPP Calibration Standard (with 525.3 Internal and Surrogate Standards) on Rxi®-5Sil MS
|Agilent Technologies|Applications
EPA Method 525.3 OPP Calibration Standard (with 525.3 Internal and Surrogate Standards) on Rxi®-5Sil MS 5 m/z = 127.00 m/z = 109.00 m/z = 192.00 4 10 11 6 23 22 24 2 12 7 3 1 17 9 6.00…
Key words
goup, goupscan, scandftpp, dftppmin, minamu, amuextractor, extractorscans, scansdelay, delaysec, secethyl, ethylacetate, acetatestart, starttime, timeoven, ovenquadrupole
EPA Method 525.3 ONP Calibration Standard (with 525.3 Internal and Surrogate Standards) on the Rxi®-5Sil MS
|Agilent Technologies|Applications
EPA Method 525.3 ONP Calibration Standard (with 525.3 Internal and Surrogate Standards) on the Rxi®-5Sil MS 21 23 m/z 200.00 m/z 215.00 m/z 214.00 m/z 229.00 Cyanazine Trailing Factor = 1.34 19 m/z 201.00 m/z 173.00 m/z 225.00 m/z 68.00…
Key words
goup, goupscan, scanmin, mindftpp, dftpppeaks, peaksamu, amuextractor, extractorscans, scansdelay, delaysec, secethyl, ethylacetate, acetatestart, starttime, timeoven